# IMPLEX CONTINUUM KNEE- HEDROCEL REVISION FEMORAL SPACERS (K980781)

_Implex Corp. · JWH · May 22, 1998 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K980781

## Device Facts

- **Applicant:** Implex Corp.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** May 22, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Implex CKS Hedrocel Revision Femoral Spacers are indicated in the revision or salvage of failed previously reconstructed knee procedures and implants; particularly in situations where bone loss has occurred from removal of prior implantations, infection or trauma. This device is intended for cemented use only.

## Device Story

Implex CKS Hedrocel Revision Femoral Spacers are porous tantalum implants designed to address bone loss in revision knee arthroplasty. Available in distal (4, 8, 12, 16 mm) and posterior (4, 8, 12 mm) thicknesses, with full or chamfered options. Spacers are fastened to the Implex Continuum Knee System Revision Femoral Components using standard instrumentation and fastener bolts. Intended for cemented use only. Used by orthopedic surgeons in clinical settings to restore joint stability and provide a foundation for fixation in salvage procedures.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing and performance data of predicate devices and materials, referenced in Master File MAF #920 and K962468.

## Technological Characteristics

Constructed from Hedrocel porous tantalum. Available in multiple thicknesses (4-16 mm) and configurations (full/chamfered). Designed for cemented fixation to Implex Continuum Knee System components using mechanical fastener bolts. No software or electronic components.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Implex CKS Revision Femoral Spacers
- Implex Continuum Porous Patella
- Implex Hedrocel Acetabular Restrictor

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

ставительных вича

ﺳﺮ

ANNOSS COLLECTION COLLECTION

# MAY 22 1998

## 510(k) SUMMARY - IMPLEX CKS Hedrocel® Revision Femoral Spacers

| Submitter Name:          | Implex Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address:       | 80 Commerce Drive<br>Allendale, New Jersey 07401-1600                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Contact Person(s):       | Robert Poggie or Robert Cohen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Phone Number:            | (201) 818-1800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Fax Number:              | (201) 818-0567                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Date Prepared:           | May 14, 1998                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Trade Name:       | Implex Continuum Knee System Hedrocel Revision Femoral Spacers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Common Name:      | Revision Femoral Spacers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name:     | Prosthesis, Knee, Femoral Spacers, Cemented                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Substantial Equivalence: | The term "substantial equivalence" as used in this 510(k)<br>notification is limited to the definition of substantial<br>equivalence found in the Federal Food, Drug and Cosmetic<br>Act, as amended and as applied under 21 CFR 807, Subpart E<br>under which a device can be marketed without premarket<br>approval or reclassification. A determination of substantial<br>equivalency under this notification is not intended to have any<br>bearing whatsoever on the resolution of patent infringement<br>suits or any other patent matters. No statements related to, or<br>in support of substantial equivalence herein shall be construed<br>as an admission against interest under the US Patent Laws or<br>their application by the courts. |
| Predicate Devices:       | The Implex CKS Revision Femoral Spacers, the Implex Continuum<br>Porous Patella, and the Implex Hedrocel Acetabular Restrictor.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Description:      | The CKS Hedrocel Distal Femoral Spacers will have 4 thickness<br>options of 4, 8, 12, and 16 mm. The distal femoral spacers will be<br>available in both full and chamfered spacer options. The CKS<br>Hedrocel Posterior Femoral Spacers will be available in 3<br>thickness options of 4, 8, and 12 mm. The CKS Hedrocel Spacers<br>are comprised of Hedrocel porous tantalum and are to be fastened<br>to Implex Continuum Knee System Revision Femoral<br>Components using the same Instrumentation and fastener bolts.                                                                                                                                                                                                                           |
| Intended Use:            | The Implex CKS Hedrocel Revision Femoral Spacers are<br>indicated in the revision or salvage of failed previously<br>reconstructed knee procedures and implants; particularly in<br>situations where bone loss has occurred from removal of prior<br>implantations, infection or trauma. This device is intended for<br>cemented use only.                                                                                                                                                                                                                                                                                                                                                                                                            |

Allendale, New Jersey 07401-1600 :

Phone: (201) 818-1800 Fax: (201) 818-0567

{1}------------------------------------------------

1.41 ULS UNIT

OL . . UI 010 UDD

Image /page/1/Picture/3 description: The image shows the word "Implex" in a bold, sans-serif font. To the left of the word is a geometric shape that includes a triangle and a circle. The triangle is on the bottom left, and the circle is on the top left. The word "Implex" is written in all lowercase letters, and there is a small symbol to the upper right of the word.

Indications For Use:

The use of the Continuum Knee System Hedrocel Revision Femoral Spacers are indicated for:

- For cemented use only. a)
- Total Knee Replacement (TKR) in severely disabled joints b) as a result of degenerative arthritis;
- Secondary revision of a previously unsuccessful revision c) femoral component;
- Other knee problems where clinical experience has shown d) that alternative modes of treatment are less likely to achieve satisfactory results;
- Trauma, or other fractures in which adequate bony fixation ૯) cannot be obtained;
- Non-union of the proximal tibia; and f)
- Salvage of a failed primary or secondary total knee or unig) knee arthroplasty.

#### Performance Data:

Previous testing of the predicate devices and their respective materials are sufficient for substantial equivalence determination. The relevant data is found in Master File MAF #920 and K962468.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

A. A. C.

MAY 2 2 1998

Robert A. Poggie, Ph.D. Director of Applied Research Implex Corporation 80 Commerce Drive 07401-1600 Allendale, New Jersey

Re : K980781 Implex Continuum Knee System Hedrocel Trade Name: Revision Femoral Spacers Regulatory Class: II Product Code: JWH February 27, 1998 Dated: Received: March 2, 1998

Dear Dr. Poggie:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Robert A. Poggie, Ph.D.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

O Celia M. Witten, Ph.D., M

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Implex Hedrocci@ Femoral Spacers 510(k) Premarket Notification K980781

Submitted by: Implex Corp.

510(k) Number (if known):

Device Name:

Implex Continuum Knee – Hedrocel Revision Femoral Spacers

### Indications For Use:

The use of the Continuum Knee System Hedrocel Revision Femoral Spacers are indicated for:

- For cemented use only. a)
- Total Knee Replacement (TKR) in severely disabled joints as a result of b) degenerative arthritis;
- Secondary revision of a previously unsuccessful revision femoral c) component;
- Other knee problems where clinical experience has shown that alternative d) modes of treatment are less likely to achieve satisfactory results;
- Trauma, or other fractures in which adequate bony fixation cannot be e) obtained;
- f) Non-union of the proximal tibia; and
- Salvage of a failed primary or secondary total knee or uni-knee g) arthroplasty.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH; Office of Device Evaluation (ODE) |         |
|--------------------------------------------------------|---------|
| (Division Sign Off)                                    |         |
| Division of General Restorative Devices                |         |
| 510(k) Number                                          | K980781 |

| Prescription Use<br>(Per 21 CFR 801.109) | X |
|------------------------------------------|---|
|------------------------------------------|---|

OR...

| Over-The-Counter Use |  |
|----------------------|--|
|----------------------|--|

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K980781](https://fda-staging.innolitics.com/device/K980781)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).