COMPLETE® (and Private Label) brand Multi-Purpose Solution is indicated for use in the chemical (NOT HEAT) disinfection, cleaning, rinsing, protein removal and storing of soft (hydrophilic) contact lenses as recommended by your eye care practitioner
Device Story
COMPLETE® and Private Label Multi-Purpose Solutions are sterile, isotonic chemical solutions for soft contact lens care. The solution contains polyhexamethylene biguanide (TrisChem™) as a preservative, hydroxypropyl methylcellulose as a lubricant, tyloxapol as a surfactant, and edetate disodium as a chelating agent. The device is used by patients at home for daily lens maintenance, including cleaning, rinsing, disinfection, and storage. It functions by chemically disinfecting the lens surface and matrix and removing protein and non-protein deposits. The solution is intended to maintain lens hygiene and prevent ocular hazards associated with contaminated lenses. It is an over-the-counter product used according to practitioner recommendations.
Clinical Evidence
Clinical safety and efficacy were previously established under P910075. Additional bench testing demonstrated the ability of the solutions to remove protein deposits from the lens surface and matrix during disinfection, showing performance comparable to the predicate device (ReNu® Multi-Plus). Microbiological studies confirmed the solutions meet FDA criteria for disinfection against bacteria, yeast, and molds, and conform to USP sterility requirements.
Technological Characteristics
Sterile, isotonic aqueous solution. Ingredients: polyhexamethylene biguanide (0.0001%), hydroxypropyl methylcellulose, tromethamine buffer, tyloxapol surfactant, edetate disodium chelating agent, sodium chloride, and purified water. Chemical disinfection (non-heat). No chlorhexidine or thimerosal.
Indications for Use
Indicated for chemical disinfection, cleaning, rinsing, protein removal, and storage of soft (hydrophilic) contact lenses as recommended by an eye care practitioner.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
K 980775
Image /page/0/Picture/1 description: The image shows the date "JUL 6 1998" in bold, black font. The month is abbreviated to "JUL". The day is the number "6". The year is "1998". The text is clear and easy to read.
# 510(k) SUMMARY
## Submitted by:
Paul J. Nowacki Manager, Regulatory Affairs Allergan, Inc. 2525 Dupont Drive P.O. Box 19534 Irvine, CA 92623-9534 (714) 246-6761 (Voice) (714) 246-5457 (Fax)
### Device Name:
| Common Name: | Contact Lens Care Multi-Purpose Solution |
|-------------------|------------------------------------------------------------------------------------------------------|
| Proprietary Name: | COMPLETE® brand Multi-Purpose Solution (8772X)<br>Private label brand Multi-Purpose Solution (8474X) |
Indications for Use:
COMPLETE® (and Private Label) brand Multi-Purpose Solution is indicated for use in the chemical (NOT HEAT) disinfection, cleaning, rinsing, protein removal and storing of soft (hydrophilic) contact lenses as recommended by your eye care practitioner
## Description:
COMPLETE® brand Multi-Purpose Solution is a sterile, isolonic solution containing hydroxypropyl methylcellulose as a lubricant, preserved with TrisChem™ (polyhexamethylene biguanide) 0.0001%, buffered with tromethamine, tyloxapol as a surfactant, ederate disodium as a chelating agent, sodium chloride, and purified water. COMPLETE® brand Multi-Purpose Disinfecting Solution contains no chlorhexidine or thimerosal.
Private label brand Multi-Purpose Solution is a sterile, isotonic solution containing purified water, sodium chloride, preserved with TrisChem™ (polyhexamethylene biguanide) 0.0001%, buffered with tromethamine, tyloxapol as a surfactant, and edetate disodium as a chefating agent. Private label brand Multi-Purpose Disinfecting Solution contains no chlorhexidine or thimerosal.
### Substantial Equivalence:
COMPLETE® brand Multi-Purpose Solution and private label brand Multi-Purpose Solution are identical to the products that were approved under P910075 and are substantially equivalent to Bausch & Lomb's ReNu@ Multi-Plus Multi-Purpose Solution (P860023/S12).
{1}------------------------------------------------
K980775
### Non-Clinical Data A.
# Microbiological Studies
COMPLETE® brand Multi-Purpose Solution and private label brand Multi-Purpose Solution were previous evaluated and approved under P910075.
- Both products meet the current FDA criteria for disinfection of contact lenses against . bacteria, yeast, and molds
- Both products are effectively preserved by FDA standards. .
- Both products conform to USP sterility requirements. .
# Preclinical
COMPLETE® brand Multi-Purpose Solution and private label brand Multi-Purpose Solution were previously evaluated under P910075. These products are safe for use in cleaning, rinsing, chemical disinfection and up to 30 day storage of all soft contact lenses.
COMPLETE® brand Multi-Purpose Solution and private label brand should not present an ocular hazard to the contact lens wearer when used as directed in the labeling for soft contact lenses.
# Compatibility/Cleaning Efficacy
COMPLETE® brand Multi-Purpose Solution and private label brand Multi-Purpose Solution were demonstrated previously under P910075 to be compatible with all soft contact lenses and able to remove non-protein deposits from the surface of the contact lens. An additional study was performed that demonstrated the ability of COMPLETE® brand Multi-Purpose Disinfecting Solution and private label brand Multi-Purpose Solution to remove protein deposits on the contract lens surface and in the lens matrix during disinfection compared to Bausch & Lomb's ReNu@ MultiPlus Multi-Purpose Solution ..
### B. Clinical
COMPLETE® brand Multi-Purpose Solution and private label brand Multi-purpose Solution were proven clinically safe and effective under P910075 and are substantially equivalent to Bausch & Lomb's ReNu@ Multi-Plus Multi-Purpose Solution (P860023/S12).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which consists of three stylized human profiles facing right, arranged in a stacked formation. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
6 1998 JUL
Mr. Paul J. Nowacki Manager, Regulatory Affairs Allergan 2525 Dupont Drive P.O. Box 19534 Irvine, CA 92623-9534
Re: K980775
Trade Name: Complete ® brand Multi-Purpose Solution and Private Label brand Multi-Purpose Solution Regulatory Class: II Product Code: 86 LPN Dated: May 20, 1998 Received: May 21, 1998
Dear Mr. Nowacki:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{3}------------------------------------------------
Page 2 - Mr. Paul J. Nowacki
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K980775
510(k) NUMBER: (IF KNOWN):
DEVICE NAME:
COMPLETE® brand Multi-Purpose Solution and
Private Label brand Multi-Purpose Solution
INDICATIONS FOR USE:
COMPLETE® (and Private Label) brand Multi-Purpose Solution is indicated for use in a COMPLE I Fivate Label prand Mulier droose ontalier an and and storing of
the chemical (NOT HEAT) disinfection, cleaning, protein removal and storing of
the chemical in the co the chemical (NOT HEAT) ulsimection, cleaning, Misling Provinces.
soft (hydrophilic) contact lenses as recommended by your eye care practitioner.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109 OR
Over-The-Counter-Use (Optional Format 1-2-96)
(Division Sign-Off)
Division of Ophthalmic Devices
\$10(k) Number K 980775
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.