The CDI 400 Extracorporeal Blod Gas Monitoring System is intended for use during cardiopulmonary bypass procedures when continuous monitoring of blood gases is desired.
Device Story
3M CDI QUIK-CELLS are sterile, single-use components for the CDI 400 Extracorporeal Blood Gas Monitoring System; used during cardiopulmonary bypass. Modification involves application of covalently bound heparin coating to fluid path and use of slightly thicker membrane. Device functions as part of an extracorporeal circuit to facilitate continuous blood gas monitoring. Operated by clinical staff in surgical settings. Heparin coating intended to improve hemocompatibility. Output provides real-time blood gas data to clinicians to assist in patient management during bypass.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Sterile, single-use extracorporeal blood gas monitoring cells. Available in 1/4", 1/8", and 1/2" diameters. Features covalently bound heparin coating on fluid path and modified membrane thickness. Designed for integration into CDI 400 monitoring system.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass procedures requiring continuous blood gas monitoring.
Regulatory Classification
Identification
A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.
Predicate Devices
CDI 400 Extracorporeal Hematocrit/Oxygen Saturation Monitoring System (K890113)
CDI 100 Extracorporeal Hematocrit/Oxygen Saturation Monitoring System (K902654)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in Firs summer with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned
510(k) number is
Submitter's Name: Submitter's Address: Contact Person: Phone Number: FAX Number: Summary Date:
3M Health Care 1311 Valencia Avenue, Tustin, CA 92780 Diana Thorson (714) 258-8001 X233 (714)258-0810
Device Trade Name: CDI 400 Extracorporeal Blood Gas Monitoring System
Device Classification Name:
Cardiopulmonary bypass on-line blood gas monitor and cardiopulmonary bypass in line blood gas sensor (21 CFR 870.4330)
Predicate Devices:
| Product | 510(k) Number | Clearance Date |
|-----------------------------------------------------------------------------|---------------|----------------|
| CDI 400 Extracorporeal<br>Hematocrit/Oxygen Saturation Monitoring<br>System | K890113 | 5/23/89 |
| CDI 100 Extracorporeal<br>Hematocrit/Oxygen Saturation Monitoring<br>System | K902654 | 11/5/90 |
### Device Description:
The 3M CDI H/S QUIK-CELLS with Heparin Treatment are sterile, single-use medical devices. They are available in 1/4", 1/8", and 1/2" diameter sizes. Additionally, the 3M CDI QUIK-CELLS with Heparin Treatment have a covalently bound heparin coating.
## Indications for Use:
The CDI 400 Extracorporeal Blod Gas Monitoring System is intended for use during cardiopulmonary bypass procedures when continuous monitoring of blood gases is desired.
{1}------------------------------------------------
#### Technological Characteristics:
This modification to the 3M CDI QUIK-CELLS includes the application of a covalently bound heparin coating to the fluid path of the device. In addition, the 3M CDI QUIK-CELLS with Heparin Treatment have a slightly thicker membrane. There are no other dimensional changes to the cells due to the addition of the Heparin Treatment.
#### Nonclinical Performance:
The performance characteristics of the 3M CDI OUIK-CELLS with Heparin Treatment were exhaustively tested and compared with the performance characteristics of the currently marketed 3M CDI OUIK-CELLS. All new and existing performance characteristics of the 3M CDI Heparin Coated QUIK-CELLS have been validated.
#### Clinical Performance:
Clinical testing was not performed on these devices.
#### Conclusions from Nonclinical Tests:
The 3M CDI OUIK-CELLS with Heparin Treatment perform as intended according to their performance specifications. The 3M CDI QUIK-CELLS with Heparin Treatment are substantially equivalent to their predicate devices.
Page 2-3
{2}------------------------------------------------
#### 2.2 Safety Literature Search
Although a safety literature search is not required for class II devices, one has been included below.
I certify that I have conducted a reasonable search of all information known or otherwise available to me about the types and causes of safety and/or effectiveness problems that have been reported with cardiopulmonary bypass on-line blood gas monitors and cardiopulmonary bypass in line blood gas sensors within the past five years. I further certify that I am aware of the types of problems to which cardiopulmonary bypass on-line blood gas monitors and cardiopulmonary bypass in line blood gas sensors are susceptible and that the following list of safety and/or effectiveness problems about these devices is complete and accurate:
- 1.) Occlusion of the cell
- Leakage of fluid from the cell, or ingress of air emboli into the 2.) cell
- Loss of sterility 3.)
Below is a bibliography of the materials upon which the above summary is based:
- 1.) Customer Complaint File, various dates, 3M Health Care, CDI Products.
- U.S. Food and Drug Administration Recall Number Z 1084/1085-1. 2.) Initiated by CDI, 3M Health Care on May 20, 1991.
- U.S. Food and Drug Administration Recall Number Z 2628-2 through 3.) Z2633-2. Initiated by CDI, 3M Health Care On December 18, 1991.
- U.S. Food and Drug Administration Recall Numbers Z-635-2 through 4.) Z-645-2. Initiated by CDI, 3M Health Care on February 2, 1992.
- Medical Device Report M174488, filed 1/30/92 by CDI, 3M Health 5.) Care.
- Medical Device Report M174493, filed 2/4/92 by CDI, 3M Health 6.) Care.
- Medical Device Report M268836, filed 2/17/92 by CDI, 3M Health 7.) Care.
- Medical Device Report M268956, filed 2/19/92 by CDI, 3M Health 8.) Care.
- 9.) Medical Device Report M268957, filed 2/19/92 by CDI, 3M Health Care.
- 10.) Medical Device Report M255427, filed 2/13/92 by CDI, 3M Health Care.
- Medical Device Report M255429, filed 2/13/92 by CDI, 3M Health 11.) Care.
{3}------------------------------------------------
- Medical Device Report M174556, filed 2/12/92 by CDI, 3M Health 12.) Care.
- Medical Device Report M174629, filed 2/2/92 by CDI, 3M Health 13.) Care.
- Medical Device Report M174966, filed 4/16/92 by CDI, 3M Health 14.) Care.
- Medical Device Report M174942, filed 4/16/92 by CDI, 3M Health 15.) Care.
- 16.) Medical Device Report M174941, filed 4/16/92 by CDI, 3M Health Care.
- 17.) Medical Device Report M174943, filed 4/16/92 by CDI, 3M Health Care.
- Medical Device Report M174964, filed 4/16/92 by CDI, 3M Health 18.) Care.
- 19.) Medical Device Report M320347, filed 3/14/93 by CDI, 3M Health Care.
- Medical Device Report M377386, filed 3/26/93 by CDI, 3M Health 20.) Care.
- Medical Device Report M377387, filed 3/26/93 by CDI, 3M Health 21.) Care.
- 22.) Medical Device Report M377388, filed 3/26/93 by CDI, 3M Health Care.
- 23.) Medical Device Report M544290, filed 8/5/94 by CDI, 3M Health Care.
- 24.) Medical Device Report M800711, filed 4/17/95 by 3M Health Care.
- Medical Device Report M820946, filed 7/31/95 by 3M Health Care. 25.)
- 26.) Medical Device Report 2023117-1997-00001, filed 3/14/97 bv 3M Health Care.
Printed name of person required to submit 510(k): Diana M. Thorson Signature of person required to submit 510(k): Machines M) theresto Title of person submitting 510(k): Advanced Regulatory Affairs Coordinator Name of Company: 3M Health Care Date: 2/24/98
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
## MAY 1 5 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Diana M. Thorson Advanced Requlatory Affairs Coordinator 3M Health Care, Cardiovascular Systems CDI Products 1311 Valencia Avenue Tustin, CA 92780
Re : K980732 3M CDI Quick-Cell with Heparin Treatment Regulatory Class: II (Two) Product Code: DRY Dated: February 24, 1998 Received: February 25, 1998
Dear Ms. Thorson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set-forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{5}------------------------------------------------
#### Page 2 - Ms. Diane Thorson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K980732-
3M CDI QUIK-CELLS With Heparin Treatment Device Name:__________________________________________________________________________________________________________________________________________________________________
These cells are to be used with the CDI System Indications For Use: Indications for 6000 Gas Monitoring System, which is intended for 400 Extractporcal brood our heartering by sound.
use during cardiopulmonary bypass procedures when continous monitoring of blood gases is desired.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sette a. Cooper le
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K-98073. 510(k) Number_
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.