PREMILENE NONABSORBABLE POLYPROPYLENE SURGICAL SUTURE

K980703 · Aesculap, Inc. · GAW · May 4, 1998 · General, Plastic Surgery

Device Facts

Record IDK980703
Device NamePREMILENE NONABSORBABLE POLYPROPYLENE SURGICAL SUTURE
ApplicantAesculap, Inc.
Product CodeGAW · General, Plastic Surgery
Decision DateMay 4, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Premilene Nonabsorbable Polypropylene Surgical Sutures are indicated for use in all types of general soft tissue approximation and ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue.

Device Story

Premilene is a sterile, nonabsorbable, flexible monofilament surgical suture composed of isotactic polypropylene polymer. It is supplied in various lengths and diameters, with or without attached needles. The device is used by surgeons in clinical settings (OR, clinic) for soft tissue approximation and ligation across cardiovascular, ophthalmic, microsurgical, and neural applications. It functions as a mechanical closure device to hold tissue edges together during healing. The device is provided sterile and is intended for prescription use only.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including physical parameter evaluation (diameter, tensile strength, flexibility, elongation), sterilization validation, shelf-life testing, chemical identity/purity assays, and one-year animal implant studies.

Technological Characteristics

Material: Isotactic polypropylene polymer. Colorant: [Phthalocyaninato(2-)] copper (≤0.5% by weight, per 21 CFR §74.3045). Form: Sterile, flexible monofilament thread. Conformance: U.S.P. XXIII (Monograph for Nonabsorbable Surgical Suture). Physical properties: Diameter, knot/straight pull tensile strength, flexibility, elongation, elasticity, needle attachment strength. Sterilization: Validated process.

Indications for Use

Indicated for general soft tissue approximation and ligation in patients requiring cardiovascular, ophthalmic, microsurgical, or neurological procedures. Prescription use only.

Regulatory Classification

Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### MAY 4 1998 ## VII. 510(k) Summary In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided: ## A. Submitted By R. Stephen Reitzler, RAC, Vice President, Regulatory Affairs Advanced Bioresearch Associates One America Plaza 600 West Broadway, Suite 900 San Diego, California 92101 Telephone: (619) 702-0700 Regulatory Agents for: Aesculap® 1000 Gateway Boulevard South San Francisco, California 94080-7030 Telephone: (650) 876-7000 Contact: Victoria MacKinnon, Manager, Regulatory Affairs Date Prepared: February 11, 1998 ## B. Device Name Premilene Polypropylene Surgical Suture Trade or Proprietary Name: Nonabsorbable Polypropylene Surgical Suture Common or Usual Name: Nonabsorbable Polypropylene Surgical Suture Classification Name: #### C. Predicate Devices - Prolene® Nonabsorbable Polypropylene Surgical Suture (Ethicon, Inc.) . - Surgilene® Nonabsorbable Polypropylene Surgical Suture (Davis & Geck) ● - Sharpoint Polypropylene Surgical Suture (Sharpoint) . The subject device is substantially equivalent to predicate devices listed above. {1}------------------------------------------------ # D. Device Description The subject device is a nonabsorbable, flexible monofilament suture thread which is supplied sterile. It is composed of isotactic polypropylene polymer, and is indicated for soft tissue approximation. It will be offered undyed, and dyed with the FDA approved colorant [Phthalocyaninati(2-)] copper in accordance with Title 21 CFR, §74.3045. It will be available with and without standard needles attached. ## E. Intended Use Premilene Nonabsorbable Polypropylene Surgical Sutures are indicated for use in all types of general soft tissue approximation and ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue. ## F. Comparison to Predicate Devices The subject Premilene Nonabsorbable Polypropylene Surgical Suture is composed of the same material as are the predicate devices, that being the isotactic form of the polyolefin polymer polypropylene. Further, the subject device is offered undyed, and dyed with the same colorant as are the predicate devices, that being {phthalocyaninato(2-)] copper at a concentration of ≤0.5% by suture weight in accordance with Title 21 CRR, §74.3045. The subject device has the same design as do the predicate devices, being a sterile, flexible monofilament thread which is offered in a variety of lengths and a range of diameters conforming with the requirements of U.S. Pharmacopeia (U.S.P.) XXIII, and which is offered with or without one of a selection of standard needles attached. Further, as is the case with the predicate devices, the subject device conforms in all respects to the requirements of the Official Monograph for Nonabsorbable Surgical Suture in U.S.P. XXIII, including <861> Sutures -- Diameter, <871> Sutures -- Needle Attachment, and <881> Tensile Strength. Physical properties of the subject device are substantially equivalent to those of the predicate devices, including fiber diameter, knot pull tensile strength, {2}------------------------------------------------ straight pull tensile strength, flexibility, elongation, elasticity, and needle attachment strength. The subject device is manufactured in the same manner as at least one of the predicate devices, being supplied to Aesculap® by the manufacturer of said predicate suture. As such, the suture fiber from which the subject device is made has the same chemical characteristics, biocompatibility profile, and in vivo performance properties as does that predicate device. The subject device is packaged and sterilized in the same or equivalent manner, and has the same labeling claims as do the predicate devices, including indications, contraindications, warnings, cautions and precautions. ## G. Summary of Non-Clinical Tests Non-Clinical testing conducted on the subject device to demonstrate its substantial equivalence to predicate devices included physical testing for all parameters identified above, sterilization validation and evaluation of sterilant residues, and shelf-life testing. Testing conducted by the manufacturer and supplier to Aesculap® of the finished suture fiber included chemical assays for identity and purity, testing of physical properties to prove conformance to the requirements of U.S.P., in vitro and in vivo biosafety studies, and one (1) year implant studies in animals to demonstrate retention of tensile strength. #### H. Summary of Clinical Tests (Not applicable) ### I. Conclusions of Non-Clinical and Clinical Tests The results of all testing demonstrated the substantial equivalence, if not superiority, of the subject device to one or more predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 #### 1 1098 MAY Aesculap c/o Mr. Steve Reitzler, RAC Vice President, Regulatory Affairs Advanced Bioresearch Associates One America Plaza 600 West Broadway, Suite 900 San Diego, California 92101-3302 #### K980703 Re: Premilene Nonabsorbable Polypropylene Surgical Suture Dated: February 19, 1998 Received: February 23, 1998 Dear Mr. Reitzler: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices that were regulated as transitional devices and that have been reclassified into class II. Notice of this reclassification was published in the Federal Register on Friday, May 31, 1991 (Vol. 56, No. 105, Pages 24684 and 24685). A copy of this Federal Register can be obtained by calling the Division of Small Manufacturers Assistance (DSMA) at (800) 638-2041 or (301) 443-6597. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations: - 1. The Premilene Nonabsorabable Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery, and neurological procedures. - 2. This device may not be manufactured from any material other than a long chain polyolefin polymer known as polypropylene. In addition, you must maintain documentation at your premises regarding vendor certification for raw or semiprocessed source material, all manufacturing and quality control release procedures, and validation of sterilization procedures used in the manufacture of the Polypropylene surgical suture. Any deviation of the source material or processing as described in this 510(k) notification requires submission of a new premarket notification and Food and Drug Administration (FDA) clearance prior to commercial distribution of the modified device. {4}------------------------------------------------ Page 2 - Mr. Reitzler The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibition against misbranding and adulteration. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, The Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control Provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4595. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597, or at its internet address http://www.fda.gov/cdrh/dsmamain.html. Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ Pagc / of l 510(k) Number (if known): K980703 Device Name: Premilene® Nonabsorbable Polypropylene Surgical Suture Indications For Use: 11:00 .. : : : : Premilene® sutures are indicated for use in all types of general soft tissue approximation and ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General Restorative Devices 510(k) Number 12980703 メ Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...