The Summit Krumeich-Barraquer Microkeratome is an AC-powered ophthalmic surgical instrument used to shave corneal tissue for lamellar transplants or harvesting. It consists of a service terminal, microcomputer, vacuum pump, motor-driven oscillating blade, and suction ring. The device is operated by qualified personnel in a surgical setting. The operator sets the blade oscillation frequency and forward advancement speed (0.1 to 3.0 mm/sec). A stepper motor controls the automated advancement of the blade across the cornea, while the suction ring positions the device. The system includes an electrical and vacuum reserve to ensure procedure completion during power failure. The surgeon monitors the procedure via an LCD display and manometer. The device provides precise, automated corneal sectioning, which can improve the quality of the cut surface compared to manual methods, potentially benefiting patients by facilitating more accurate corneal transplantation procedures.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
AC-powered keratome; materials include stainless steel, plastic, glass, and natural diamond. Features a stepper motor for automated blade advancement (0.1–3.0 mm/sec) and adjustable blade oscillation. Includes a vacuum-based suction ring for eye positioning. System components: service terminal, microcomputer, vacuum pump, and motor assembly. Sterilization via autoclave. Connectivity includes foot pedals for vacuum and cutting control, and an LCD display for parameter monitoring.
Indications for Use
Indicated for keratoplasty, corneal harvesting, and microlamellar keratoplasty in patients requiring partial thickness corneal transplantation or tissue removal.
Regulatory Classification
Identification
A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.
Predicate Devices
Eyetech Barraquer-Krumeich-Swinger Refractive Set (K954058)
Submission Summary (Full Text)
{0}------------------------------------------------
980675
December 15, 1998
Summary of Safety and Summit Krumeich-Barraquer Microkeratome
Effectiveness Information
et Notification, Section 510(k) Regulatory Authority: Safe Medical Devices Act of 1990,
4|21 CFR 807.92
- 1. Device Name: Summit Krumeich-Barraquer Microkeratome Trade Name: Electronic Microkeratome Common Name: Keratome Classification Name: KERATOME, AC-POWERED
- 2. Establishment Name & Registration Number: Summit Technology, Inc. Name: Number: 1287364
- 3. Classification: & 886.4370 Keratome.
\$ 886.4370 Keratome. (a) Identification. A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant. {b} Classification. Class I. [55 FR 48443, Nov. 20, 1990]
Device Class: Class I Classification Panel: Ophthalmic devices panel Product Code: 86HNO
- Special Controls: 4. Not applicable to this device.
- 5. Labeling:
Federal (United States) Law restricts this device to sale by or on the order of a physician.
The warnings, cautions for the device are presented here. The balance of the product labeling is found in Appendix I.
Image /page/0/Picture/13 description: The image shows four warning signs with accompanying text. The first sign is a triangle with an exclamation point inside, labeled "Danger", and the text warns of the danger of cutting oneself during installation or removal of the cutting blade. The second sign is a triangle with a lightning bolt inside, and the text warns not to remove the cover of the electronic part of the instrument. The third sign is an "i" inside a square, and the text states that the instrument complies with safety regulations. The fourth sign is a triangle with an exclamation point inside, labeled "Caution", and the text describes the microkeratome as a precision instrument and warns of damage if the instrument falls.
{1}------------------------------------------------
K980675
#### Fauivalent / Predicate Device: 6.
- Evetech Barraquer-Krumeich-Swinger Refractive Set, K954058 .
#### 7. Device Description:
The Summit Krumeich-Barraquer Microkeratome is an automated version of the Eyetech-Barraquer-Krumeich Swinger Refractive Set found substantially equivalent under K954058. Other than the automated cutting action option of the electronic model, both microkeratomes share the following basic description.
The Microkeratome is a medical "instrument" exclusively designed for use in ophthalmic surgery. This instrument should only be installed, used and serviced by suitably qualified personnel. The speed of forward movement during cutting and the frequency of oscillation of the blade can be individually set and controlled by the operator. The instrument will remain programmed until the data is overwritten by new data. For safety reasons the instrument is equipped with an electrical energy and vacuum reserve which is maintenance free. There is enough reserve energy and vacuum stored that in the event of an electrical failure or loss of main voltage at an inconvenient point in the procedure, the procedure can still be completed.
The primary function of this instrument is to make a thin anterior corneal lamellar section leaving the cornea attached nasally by means of a hinge or to remove or harvest the cornea completely. The suction ring is responsible for positioning the microkeratome over the eye. The microkeratome has an oscillating blade, which, with a defined advancing speed (0.1 to 3.0mm/sec.) is guided over the cornea of the patient. The frequency of oscillation and the advancing speed of the blade greatly influence the quality of the cut surface.
## Indications.
- . Keratoplasty & corneal harvesting
- . Microlamellar keratoplasty
## The instrument consists of 4 components:
- Service Terminal, Microcomputer and Vacuum-Pump .
- Motor for oscillating blade and Step Motor for Forward Movement .
- . Microkeratome
- Suction Ring with Vacuum Connection .
#### Elements of the Instrument:
- 1. LCD Display
- Service keys and LED 2.
- 3. Vacuum tubing to suction ring
- 4. Manometer, pressure indicator of vacuum container
- 5. Water separator
- Button for forward movement of microkeratome drive steel leaf 6.
- Stepper motor for microkeratome advancement 7.
- 8. Foot pedal for vacuum
- 9. Foot pedal for cutting, start, stop function
- 10. Button to position microkeratome drive steel leaf
- 11. Microkeratome drive steel leaf of stepper motor for forward movement
- 12. Pin guide for forward movement of microkeratome drive steel leaf
- 13. Coupling clip for microkeratome drive steel leaf
- 14. Suction ring
{2}------------------------------------------------
December 15, 1998
K980675
- 15. Suction ring handle
- 16. Microkeratome
- 17. Motor for blade oscillation
# 8. Modified Device Data:
The Summit Krumeich-Barraquer Microkeratome is an improved model of a previously reviewed substantially equivalent device, the Eyetech-Barraquer-Krumeich-Swinger Refractive Set, covered by K954058.
- 9. Applicant Name & Address: Summit Technology, Inc. 21 Hickory Drive Waltham, MA 02451
# 10. Submission Correspondent:
Eric P. Ankerud Vice President, Quality, Regulatory and Clinical Affairs Summit Technology, Inc. 21 Hickory Drive Waltham, MA 02451 Ph: 781-672-0570 Fax: 781-890-6316
# 11. Manufacturing Facility:
At the present time, the Summit Krumeich-Barraquer Microkeratome is manufactured by GEBAUER GMBH., MONBACHSTR. 7/1, 75242 NEUHAUSEN - ENZKREIS. GEBAUER GMBH, is a registered medical device manufacturing facility located in Germany.
# 12. Feature Comparison Table:
| FEATURE | Summit K-B<br>Microkeratome | Eyetech Refractive<br>Set K954058 | SE? |
|-------------------------|---------------------------------------------------------------------|------------------------------------------|-----|
| Indications for Use(s): | Keratoplasty & corneal<br>harvesting; Microlamellar<br>keratoplasty | Same | Yes |
| Method of cut | Automatic - Step motor<br>controlled 0.1 to 3.0mm/sec. | Manual advance -<br>micrometer-type dial | Yes |
| Sterilization Method: | Autoclave | Same | Yes |
| Blade Oscillation | Adjustable, with preset<br>default | Fixed | Yes |
| Materials: | Stainless Steel, Plastic, Glass,<br>Natural Diamond | Same | Yes |
| Manufacturer: | Gebauer GMBH | Same | Yes |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The text is black against a white background, providing high contrast and readability.
Image /page/3/Picture/2 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 2 1998
Eric Ankerud, J.D. Vice President, Quality, Regulatory and Clinical Affairs Summit Technology, Inc. 21 Hickory Drive Waltham, Massachusetts 02451
Re: K980675 Trade Name: Summit Krumeich - Barraquer Microkeratome Regulatory Class: I Product Code: 86 HNO Dated: December 15, 1998 Received: December 16, 1998
Dear Mr. Ankerud:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number 1980 (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Robert Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure ..
{5}------------------------------------------------
# 510(k) Number (if known): K980675
Device Name: Summit Krumeich-Barraquer Microkeratome
## Indications For Use:
- Keratoplasty & corneal harvesting ●
- Microlamellar keratoplasty .
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NECESSSARY
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Ementt Beens
of Onhthalmic Devices 510(k) Number
ل Prescription Use
OR
Over-The-Counter Use _
(Per 21 CFR 801.109)
(Optional format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.