The Preference® standard, CooperClear™, CV 43™, Vantage® and Vantage® Accents (tetrafilcon A) Hydrophilic Contact Lenses are intended for use as a daily wear lens for the correction of visual acuity by not-aphakic persons with non-diseased eyes that are myopic or hyperopic, who exhibit astigmatism of 2.50 diopters or less that does not interfere with visual acuity. Preference Toric™ and Cooper Toric™ Hydrophilic Contact Lenses are indicated for daily wear use by not-aphakic persons with non-diseased eyes that are myopic or hyperopic, who exhibit astigmatism up to 9.00 Diopters and can obtain satisfactory visual acuity, in a spherical range of -20.00 to +12.00.
Device Story
Tetrafilcon A hydrophilic contact lenses; hemispherical shells; placed on cornea to focus light on retina. Material: hydrophilic random terpolymer of 2-hydroxyethyl methacrylate, N-vinylpyrrolidone, and methylmethacrylate; cross-linked with divinylbenzene. Available as spherical or astigmatic (toric) lenses. Optional handling or enhancement tints using C.I. Reactive Blue 163, Yellow 86, and Red 22. Used by patients for daily wear to correct visual acuity. Prescribed by eye care professionals.
Clinical Evidence
Bench testing only. Testing included water content, light transmittance, refractive index, and mechanical properties. Biocompatibility testing included cytotoxicity, ocular irritation, and acute systemic toxicity studies.
Technological Characteristics
Hydrophilic random terpolymer of 2-hydroxyethyl methacrylate, N-vinylpyrrolidone, and methylmethacrylate; cross-linked with divinylbenzene. Hemispherical shell form factor. Available with handling or enhancement tints (C.I. Reactive Blue 163, Yellow 86, Red 22).
Indications for Use
Indicated for daily wear correction of myopia or hyperopia in non-aphakic persons with non-diseased eyes. Preference standard/CooperClear/CV 43/Vantage models indicated for astigmatism ≤ 2.50D. Preference Toric/Cooper Toric models indicated for astigmatism up to 9.00D and spherical range -20.00 to +12.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Submission Summary (Full Text)
{0}------------------------------------------------
MAY | 3 1998
# G. 510(K) Summary (page 1 of 2)
2.
1. Submitter's Name:
Contact Person:
CooperVision, Inc. 711 North Road Scottsville, NY 14546 Phone: (716) 385-6810 FAX:
Bonnie Tsymbal Phone: (716) 264-3210 FAX: (716) 889-5688
3. Date Summary Prepared: February 18, 1998
- 4. Name of Device:
- Trade Name: .
Preference® standard Preference Toric™ Cooper Toric™ CooperClear™ CooperHT™ CV 43TM Vantage® Vantage® Accents
- Common Name: .
- . Classification Name:
Soft Contact Lens
Soft Hydrophilic Contact Lens (Per 21 CFR §886.5925)
5. Legally Marketed Device:
Same as Trade Name
6. Description of Device:
All tetrafilcon A Soft (hydrophilic) Contact Lenses are hemispherical shells and are available as a spherical or astigmatic lens. When placed on the cornea, the hydrated lens acts as a refracting medium to focus light rays on the retina.
The lens material is a hydrophilic random terpolymer of 2-hydroxyethyl methacrylate, N-vinylpyrrolidone and methylmethacrylate joined in a three dimensional network of terpolymer chains by divinylbenzene cross links.
{1}------------------------------------------------
- K980649
- 6. Description of Device (continued)
When produced with a handling or enhancement tint, the lens is coupled with one or more of three dye moieties through the use of varying combinations and concentrations of C.I. Reactive Blue 163, Yellow 86, and Red 22. Lenses with a handling tint are tinted blue on the surface of the lens. The handling tint increases the visibility of the lens when not worn on the eye. Lenses tinted wit an enhancement tint are tinted on the anterior surface in the shape of an iris. The enhancement tint colors available are Sky Blue, Turquoise, Violet Blue, Spring Green, Auburn and Misty Brown.
## Intended Use:
The Preference® standard CooperClear™, CV 43™, Vantage® and Vantage® Accents
tetrafilcon A) Hydrophilic Contact Lenses are intended for use as a daily wear lens for the correction of visual aculty by not-aphakic persons with non-diseased eyes that are myopic or borrouber of visual adulty by The uphane por who exhibit astigmatism of 2.50 diopters or les that does not interfere with visual acuity.
Preference Toric™ and Cooper Toric™ Hydrophilic Contact Lenses are indicated for daily se by not-aphakic persons with non-diseased eyes that are myopic or hyperopic, who astigmatism up to 9.00 Diopters and can obtain satisfactory visual acuity, in 20100 to 103000
# 8. Technological Characteristics:
The technological characteristics of tetrafilcon A lenses manufactured at Aspect Vision, Ltd. are the same as the predicate device manufactured at Scottsville.
- 9. Summary of Non-Clinical Tests:
Physical and chemical properties testing was performed on lenses made by Aspect using the alternate manufacturing method as called for in the May 1994 Permarket Notification (510(k) Guidance Document for Daily Wear Contact Lenses, Water content, light transmittance, refractive index and mechanical properties were tested. Toxicity was determined by Cytotoxicity, Ocular Irritation Study and Acute Systemic Toxicity Study.
- 10. Conclusion:
The determination of substantial equivalence is based on the results of non-clinical testing. Review of all test data demonstrates tetrafilcon A lenses manufactured at Aspect Vision, Ltd. are equivalent to the currently marketed tetrafilcon A lenses manufactured by CooperVision. Inc. The lenses manufactured at Apsect have the same design, indication and directions for use. CooperVision, Inc. concludes, therefore, that the tetrafilcon A lenses manufactured by Aspect are substantially equivalent to the tetrafilcon A lenses manufactured at Scottsville.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, depicted with flowing lines to suggest movement or flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 1998
Mrs. Bonnie Tsymbal Regulatory Associate Cooper Vision 711 North Road Scottsville, NY 14546
Re: K980649
Trade Name: Preference ® standard, Preference Toric ™, Cooper Toric, ™CooperClear ™ CooperHT ™,CV 43 ™, Vantage ® and Vantage ®Accents (tetrafilcon A) Hydrophilic Contact Lenses for Daily Wear Regulatory Class: II Product Code: 86 LPL Dated: February 18, 1998 Received: February 19, 1998
Dear Mrs. Tsymbal:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good ---------Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal
{3}------------------------------------------------
### Page 2 - Mrs. Bonnie Tsymbal
Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- _ Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
David R. Wipple
for A. Ralph Rosenthal, M.D.
A. Ralph Rosenthal, Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the CooperVision logo. The logo consists of the word "CooperVision" in a stylized font. The "C" in Cooper is made up of two circles, one inside the other. To the right of the word "CooperVision" is a black square with a white eye-like symbol inside.
Scottsville, Ne v York 14546 (716) 385-6810 Fax (716) 889-5688
# Indications for Use Statement
K980649 510(k) Number:
Device Name:
Preference® standard Preference Toric™ Cooper Toric™ CooperClear™ CooperHT™ CV 43TM Vantage® Vantage® Accents
#### Indications for Use:
The Preference® standard, CooperHTM, CooperHTM, CV 43™, Vantage® and Vantage® Accents (tetrafilcon A) Hydrophilic Contact Lenses are indicated for the correction of visual accity by nowabnake persons with non-diseased eyes that are myopic or hyperopic. The lenses can be worn by persons who exhibit astigmatism of 2.50 diopters or less that does not interfere with visual acuity.
The Preference Toric ™ and Cooper Toric™ hydrophilic contact lenses are indicated for daily wear use by not anhake persons with non-diseased eyes that are myopic or hyperopic, who exhibit astigmatism up to 9.00 diopters and can obtain satisfactory visual acuity, in a spherical range of -20.00 to +12.00.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ V
Over-the-Counter Use_
smletts
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K980649
-2-
Or
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.