DENTATUS MTI MODULAR TRANSITIONAL IMPLANTS AND PROSTHETIC SYSTEM
K980620 · Dentatus USA , Ltd. · DZE · May 15, 1998 · Dental
Device Facts
Record ID
K980620
Device Name
DENTATUS MTI MODULAR TRANSITIONAL IMPLANTS AND PROSTHETIC SYSTEM
Applicant
Dentatus USA , Ltd.
Product Code
DZE · Dental
Decision Date
May 15, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These devices are intended to be placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. Examples: Support anchors for temporary restorations during the healing and osseointegration process of permanent implants. Immediate loading temporary abutments for repairing failing tooth and implant supported restorations. Transitional supports for immediate replacement of missing teeth. Provide assistance in case planning, implant position alignments, and interim anchor foundations.
Device Story
Dentatus MTI Modular Transitional Implants are titanium or titanium-alloy threaded posts (1.8-2.5 mm diameter; 14-22 mm length) designed for placement in jaw bone. Device features self-tapping threads at the base and varying head shapes. Used by dental professionals in clinical settings to provide temporary support for prosthetic devices, stabilize full dentures, or serve as interim anchors during osseointegration of permanent implants. Device is manually placed by the clinician. Output is a stable mechanical foundation for temporary dental restorations, facilitating patient chewing function and oral comfort during transitional treatment phases.
Clinical Evidence
No clinical data provided. Safety and effectiveness are supported by literature review of endosseous implants, including the Toronto 10-year study on Branemark implants and consensus conference proceedings, demonstrating predictable outcomes for dental implant techniques. Bench testing confirms material compliance with ASTM F-136 and ASTM F-67-95.
Technological Characteristics
Materials: High-purity titanium (ASTM F67-95) or titanium-6 aluminum-4 vanadium alloy (ASTM F136). Form: Threaded endosseous posts, 1.8-2.5 mm diameter, 14-22 mm length. Principle: Mechanical osseointegration/stabilization. Sterilization: Standard methods. No software or electronic components.
Indications for Use
Indicated for patients requiring temporary support for prosthetic devices in the upper or lower jaw arches, including temporary restorations during permanent implant healing, repair of failing tooth/implant restorations, immediate replacement of missing teeth, and interim anchor foundations for case planning.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a handwritten string of characters, "KA80620". The characters are written in a dark ink, and the handwriting appears somewhat cursive. The string seems to be a combination of letters and numbers.
#### XIII. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA. (Separate Pages)
A. Submitter: Stephen B. Rubinsky, Dentatus USA, Ltd. 192 Lexington Ave., New York, N.Y., 10016. Registration No. 2435118.
I. Classification Name and Number: Endosseous Implant (76DZE),
II. Common/Usual Name: Dental Implant, Endosseous, Post (or screw)-type, titanium or titanium allov.
III. Proprietary Name: Dentatus Transitional Implants, MTI-MP™
IV. Classification: These devices were classified by the Dental Devices Panel (Title 21 CFR 872.3640.
V. Performance Standards: None applicable. All metals meet ASTM specifications including ASTM F-136 for the titanium (ELI) alloy and ASTM F 67-95 for pure titanium.
VI. Description: The Dentatus Mini Transitional Implants (MTTM) are titanium or titanium-alloy threaded posts, 1.8 mm to 2.5 mm in diameter and from 14 to 22 mm. in length. There are variations; they are threaded at the bottom-self-tapping, and the "head" is of slightly varying shapes.
VII. Labels and Instructions for use are provided as are labels for competitive products.
VIII. Intended use: These devices are intended to be placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices and to restore the patient's chewing function. IX. Substantial Equivalence: This device is equivalent to devices manufactured and sold before 1976, having a U. S. classification number of 76 DZE and described under CFR 872.3640. It is also equivalent to several devices currently on the market that have been determined by the FDA to be substantially equivalent to the above preamendment devices. This device is equivalent (and nearly identical) to the Sendax MDI implant cleared by K-972351. Its shape, form, composition, and usage are nearly identical to those cleared in K-972351. These devices are manufactured of high-purity (99+%) titanium, or titanium alloy (ASTM F 136-84 titanium - 6 aluminum - 4 vanadium). The excellent biocompatibility and wear characteristics of titanium and this alloy have been demonstrated by the long history of their use in surgical and dental prothesis. In its transitional mode, this device is substantially equivalent (nearly identical) to the device of Sendax MDIC Management, New York, N.Y., "Sendax MDI, K-972351.In its extended use this device is substantially equivalent to the many endosseous implant devices of the screw or cylinder type that have been cleared under a variety of 510(k)s. Several of these are listed below.
1. Bofors Nobelpharma, Goteborg, Sweden: Titanium Implant Device, K-820013.
2. Sherwood, St. Louis, Mo: Monoject Endosseous Dental Implant, K-781060A.
- 3. Core-Vent Corp., Encino, CA: Screw-Vent Endosseous Implants, K-861426, A, B,
4. Impla-med, Inc., Sunrise, FL: P.M.T. Surgical Implant Components, K892124A, (Titanium Grade 3, ASTM F67-83); K-921854 and others.
- 5. Steri-Oss (Subs. Denar, Anaheim, CA.), Denar Dental Implant, K-852802; Steri-Oss Implant, K-884845.
6. Artech, Inc., Chantilly, VA: Artech Submerged Screw Implant, K-891346, K-895267.
8. Odontit, SA, Buenos Aires, Argentina: Osseointegrated Implants System, K-915375.
{1}------------------------------------------------
X. Brief Literature Review: Detailed literature surveys were made of the typical endosseous implant in its many forms and of the transitional use of endosseous implants.
The possible adverse effects summarized in this 510(k) cover those listed by the United States classification panel [Federal Register, vol. 45, No. 251, pp 86025-6, Dec. 30, 1980], as well as to those revealed in a recent literature search. Matukas, "Medical Risks Associated with Dental Implants," states, "Little or no hard data could be found on the medical risks associated with [dental] implants." Because of the wide-spread usage of dental implants, Smith and Zarb made a careful review of the literature and proposed criteria for implant success.
The Journal of Dental Education published a special issue "Proceedings of the Consensus Development Conference on Dental Implants [National Institutes of Health, Bethesda, MD, June 13-15, 1988], Vol. 52, No. 12, pp. 677-831, Dec. 1988. This added to the literature search above, with sqme especially pertinent reprints from the scientific literature, provide a comprehensive summary of available scientific data.
Zarb completed his report of the detailed Toronto 10-year study by concluding that "the tried and tested Branemark implant technique has revolutionized the treatment options open to the prosthodontist. For the edentulous patient ... the prospect for a lifetime of restored oral comfort, function, and appearance have now become predictable and reliable." These results are ample evidence of the safety and effectiveness of these endosseous implants.
A literature search of the "transitional" use of endosseous implants shows the rapidly developing interest of implantologists in making the application of endosseous implants less traumatic. Gottherer and Singer report on effective means for preliminary stabilization of full denture patients.
These devices are manufactured from the titanium materials proved effective by the above summarized years of clinical usage, are sterilized by the standard methods, and are meant for use according to methods in the cited substantially equivalent products. Thus, they involve no new types of technology and no new technological questions.
### END OF SUMMARY
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, with flowing lines beneath them, possibly representing water or movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 5 1998
Mr. Stephen B. Rubinsky ·President Dentatus USA, Ltd. 192 Lexington Avenue New York, New York 10016
Re : K980620 Dentatus MTI Modular Transitional Implants Trade Name: and Prosthetic Requlatory Class: III Product Code: DZE Dated: February 14, 1998 Received: February 18, 1998
Dear Mr. Rubinsky:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{3}------------------------------------------------
Page 2 - Mr. Rubinsky
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). " Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# VII.2 Indications for Use: [Separate Page]
## 510(k) Number: K980620
# Device Name: DENTATUS MTI MODULAR TRANSITIONAL IMPLANTS
Intended to be placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
Examples:
Support andhors for temporary restorations during the healing and osseointegration process of permanent implants.
Immediate loading temporary abutments for repairing failing tooth and implant suported restorations.
Transitional supports for immediate replacement of missing teeth.
Provide assistance in case planning, implant position alignments, and interim anchor foundations.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
| Concurrence of Office of Device Evaluation (ODE) | |
|--------------------------------------------------|---------------------------------------------------------------------|
| (Division Sign-Off) | |
| Division of Dental, Infection Control, | |
| and General Hospital Devices | |
| 510(k) Number | K950620 |
| Prescription Use | <div style="display:inline-block;">V</div> or Over-The-Counter Use_ |
| (Per 21 CFR 801.109) | |
| | (Optional Format 1-2-96) |
(repl. p. 5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.