K980591 · Bausch & Lomb, Inc. · LPN · Jun 24, 1998 · Ophthalmic
Device Facts
Record ID
K980591
Device Name
RENU MULTIPLUS LUBRICATING & REWETTING DROPS
Applicant
Bausch & Lomb, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Jun 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
Bausch & Lomb ReNu MultiPlus™ Lubricating & Rewetting Drops are indicated for use to lubricate and rewet soft (hydrophilic) contact lenses during wear. The product may be used with daily wear or extended wear lenses and with disposable lenses or lenses prescribed for frequent replacement.
Device Story
ReNu MultiPlus Lubricating & Rewetting Drops is a sterile, buffered, isotonic solution containing povidone as a demulcent. It is used by contact lens wearers to lubricate and rewet soft (hydrophilic) lenses during wear. The solution is supplied in a low-density polyethylene bottle. It functions by providing moisture to the lens surface, enhancing comfort during wear. The device is intended for over-the-counter use by patients.
Clinical Evidence
A one-month clinical study compared the test device to the predicate device with 87 subjects (174 eyes). Results showed no adverse events and no significant differences between test and control groups regarding slit lamp evaluations, patient symptoms, physiological response, visual acuity, or average daily wear times.
Indicated for users of soft (hydrophilic) contact lenses, including daily wear, extended wear, disposable, and frequent replacement lenses, to lubricate and rewet lenses during wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
## JUN 2 4 1998
510 (k) PREMARKET NOTIFICATION BAUSCH & LOMB ReNu MultiPlus™ Lubricating & Rewetting Drops
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# BAUSCH & LOMB ReNu MultiPlus™ Lubricating & Rewetting Drops
#### 1. Submitter Information:
Bausch & Lomb Incorporated Global Vision Care 1400 North Goodman Street Rochester, New York 14692-0450
Kim S. DeVitto Contact Person: Manager, Regulatory Affairs
(716) 338-6401 Telephone Number:
#### 2. Device Name:
In-Eye Soft (hydrophilic) Contact Lens Solution (Lubricating Classification Name: and/or Rewetting Drops
ReNu MultiPlus™ Lubricating & Rewetting Drops Proprietary Name:
#### 3. Predicate Device:
The Bausch & Lomb products, ReNu Lubricating & Rewetting Drops and Sterile Lens Lubricant were selected as the predicate devices for ReNu MultiPlus™ Lubricating & Rewetting Drops.
#### 4. Description of the Device
Bausch & Lomb ReNu MultiPlus Lubricating & Rewetting Drops is a sterile, buffered, isotonic solution containing the demulcent povidone. The solution is preserved with edetate disodium and sorbic acid, both at a concentration of 0.1%. Boric acid, potassium chloride, sodium borate and sodium chloride act as buffering and tonicity agents. The solution is supplied sterile in a low density polyethylene bottle labeled with a lot number and expiration date.
{1}------------------------------------------------
#### Indications for Use: પં
Bausch & Lomb ReNu MultiPlus™ Lubricating & Rewetting Drops are indicated for use to lubricate and rewet soft (hydrophilic) contact lenses during wear. The product may be used with daily wear or extended wear lenses and with disposable lenses or lenses prescribed for frequent replacement.
### Description of Safety and Substantial Equivalence 6.
A series of preclinical tests and clinical testing was performed to demonstrate the safety and effectiveness of ReNu MultiPlus Lubricating & Rewetting Drops. A summary of the test results is provided below.
### Preclinical Testing
A series of in-vitro and in-vivo preclinical chemical, toxicological and microbiological studies were performed to assess the safety and effectiveness of the product in accordance with the guidelines set forth in FDA's May 1, 1997 Guidance for Industry - Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.
The results of these studies indicate that the physical, chemical and microbiological properties of Bausch & Lomb ReNu MultiPlus Lubricating & Rewetting Drops are substantially equivalent to the predicate devices, Bausch & Lomb ReNu Lubricating & Rewetting Drops and Sterile Lens Lubricant.
The solution is non-toxic to the ocular tissue as demonstrated by in-vivo preclinical testing in laboratory animals. In-vitro lens compatibility testing was conducted to establish product compatibility with soft (hydrophilic) contact lenses.
### Clinical Testing
A one (1) month study was conducted to clinically evaluate the safety and efficacy of Bausch & Lomb ReNu MultiPlus Lubricating & Rewetting Drops for use in the lubricating and rewetting of soft (hydrophilic) contact lenses during wear. The clinical study was designed in accordance with the Premarket Notification (510 (k)) Guidance for Contact Lens Care Products of May 1, 1997. The test cell included Bausch & Lomb ReNu MultiPlus™ Lubricating & Rewetting Drops. The control cell consisted of the predicate device, Bausch & Lomb ReNut Lubricating & Rewetting Drops.
A total of 87 subjects (174 eyes) successfully completed the study, 59 subjects (93,7%) from the Test Group and 28 subjects (90.3%) from the Control Group. During this study, no adverse events were reported; there were not significant differences between Test and Control regimens for slit lamp evaluations or patient symptoms. Significant differences were not seen between the Test and Control Groups with respect to physiological response, visual acuity and average daily wear times.
0 : 1
{2}------------------------------------------------
Results of the clinical study demonstrate the safety, effectiveness and substantial equivalence of Bausch & Lomb ReNu MultiPlus™ Lubricating & Rewetting Drops to the predicate device for lubricating and rewetting soft (hydrophilic) contact lenses.
#### 7. Substantial Equivalence
Bausch & Lomb ReNu MultiPlus Lubricating & Rewetting Drops are substantially equivalent in terms of indications for use, safety and effectiveness to the predicate devices: Bausch & Lomb ReNu Lubricating & Rewetting Drops (approved for marketing under PMA P820031) and Bausch & Lomb Sterile Lens Lubricant (approved for marketing under NDA N17-945). Any differences between the new device and its predicates do not effect the use of this product.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a partial view of a logo or emblem. The logo features a stylized depiction of an eagle or bird-like figure, rendered in a minimalist, abstract style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - (USA)" is arranged around the upper portion of the emblem, suggesting it represents a governmental organization related to health and human services in the United States. The image is in black and white, with the logo appearing to be a stamp or printed graphic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### JUN 24 1998
Kim S. DeVitto Manager, Regulators Affairs Baush & Lomb Inc. 1400 North Goodman St. Rochester, NY 14692-0450
Re: K980591 Trade Name: ReNu Multiplus ™ Lubricating & Rewetting Drops Regulatory Class: II Product Code: 86 LPN Dated: May 21, 1998 Received: May 22, 1998
Dear Ms. DeVitto:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{4}------------------------------------------------
Page 2 - Ms. Kim S. Devitto
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, . please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Roerl lorentbal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
ausch & Lamb 0(k) Premarket Notificution File K980591
> Bausch & Lomb Incorporated 1400 North Goodman Street Rochester, NY 14692-0450
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
K980591
Device Name:
ReNu MultiPlus™ Lubricating & Rewetting Drops
Indications for Use:
Bausch & Lomb ReNu MultiPlus™ Lubricating & Rewetting Drops are indicated for use to lubricate and rewet soft (hydrophilic) contact lenses during wear. The product may be used with daily wear or extended wear lenses and with disposable lenses or lenses prescribed for frequent replacement.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Prescription Use<br>Use X<br>(Per 21 CFR 801.109) | <div> <span style="text-decoration:underline;">Daniel W. Brown Ph.D.</span> <br/> (Division Sign-Off) <br/> Division of Ophthalmic Devices <br/> 510(k) Number K980591 </div> | Over-The-Counter X |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.