K980565 · General Scanning, Inc. · LMA · Mar 19, 1998 · Radiology
Device Facts
Record ID
K980565
Device Name
LD2000 SERIES
Applicant
General Scanning, Inc.
Product Code
LMA · Radiology
Decision Date
Mar 19, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2030
Device Class
Class 2
Indications for Use
The General Scanning model LD2000 SERIES laser film digitizer is intended to convert medical images stored on film to digital data for copying, storage, and transmission. It is intended for use to digitize copies of conventional x-rays and films produced by imagers that fall in the ranges for size and optical density range detailed in the specifications. The digitizer utilizes laser light to scan the film, extracts the image data by measuring the transmitted light, converts it into digital form, and transmits the data over a digital link. The digitized information may then be transmitted to a Laser Imager for printed hardcopy or to a Work Station for use in Teleradiology (transmission of images) or PACS (Picture Archiving Communication Systems). The LD2000 SERIES contains the scanning laser and detection module, a film transport, and electronics to convert and transmit the image data and provide control and calibration.
Device Story
LD2000 Series is a laser film digitizer; converts analog medical x-ray/film images into digital data. Input: physical film; process: focused laser beam scans film while motor-operated rollers transport film; transmitted light measured by nine solid-state photodiodes; A/D conversion creates digital pixel values. Output: digital image data transmitted via SCSI interface to workstations, PACS, or laser imagers. Used in clinical settings; operated by trained personnel. Benefits: enables teleradiology, digital storage, and electronic transmission of conventional film-based medical images.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through technical specifications including spatial resolution (up to 4096 x 4974 pixels), MTF (0.4 at 6 lp/mm), and optical density resolution.
Technological Characteristics
Laser film digitizer; visible semiconductor laser diode source; polygon scanner; nine solid-state photodiode detectors; 12-bit grey scale; SCSI interface. Dimensions: 11"D x 22.25"W x 40"H (with pedestal). Power: 87-132/200-264 VAC, 47-63 Hz. Firmware-controlled via PROM.
Indications for Use
Indicated for the digitization of conventional x-rays and medical films for copying, storage, and transmission in clinical environments. Applicable to films within specified size (8" to 14" wide) and optical density (0.03 to 3.5 OD) ranges. Prescription use only.
Regulatory Classification
Identification
A medical image digitizer is a device intended to convert an analog medical image into a digital format. Examples include systems employing video frame grabbers, and scanners which use lasers or charge-coupled devices.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std.). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K980565
## MAR 1 9 1998
### 510(k) SUMMARY 2.
| 2.1 | Company Name | General Scanning Inc. (GSI)<br>500 Arsenal Street<br>Watertown, MA 02172 USA |
|-----|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 2.2 | Contact Person | Steven P. Zis<br>Phone # (617) 924-1010 x181<br>Fax # (617) 923-1877 |
| 2.3 | Date | February 12, 1998 |
| 2.4 | Device Name | |
| | Classification Name | Medical Image Digitizer |
| | Classification Number | 90LMA - Radiology Device |
| | Common Name | Laser Film Digitizer |
| | Trade/Proprietary Name | LD2000 SERIES (Marketing Name) |
| 2.5 | Substantial Equivalence - Predicate Devices | |
| | General Scanning LD2800 | K973411 |
| | Lumiscan 85LF | |
| | Lumiscan 100 | K901423 |
| | Konica | K9333830 |
| | Nishimoto Sangyo ED-2000 | |
| | Nishimoto Sangyo ED-3500 | |
| 2.6 | Device Description and "Statement of Indications For Use"<br>The General Scanning model LD2000 SERIES laser film digitizer is<br>intended to convert medical images stored on film to digital data for<br>copying, storage, and transmission. It is intended for use to digitize copies<br>of conventional x-rays and films produced by imagers that fall in the<br>ranges for size and optical density range detailed in the specifications.<br><br>The digitizer utilizes laser light to scan the film, extracts the image data by<br>measuring the transmitted light, converts it into digital form, and transmits<br>the data over a digital link. The digitized information may then be | |
transmitted to a Laser Imager for printed hardcopy or to a Work Station for use in Teleradiology (transmission of images) or PACS (Picture Archiving Communication Systems).
The LD2000 SERIES contains the scanning laser and detection module, a film transport, and electronics to convert and transmit the image data and provide control and calibration.
{1}------------------------------------------------
#### Principle of Operation - Overview 2.6.1
The digitizer is a system for converting images on x-ray and films into digital data. It accomplishes this by scanning a focused laser beam across the film as the film itself is moved by means of motor-operated rollers. Solid state photo diodes are used to detect the amount of laser light which passes through the film at each pixel. This provides a measure of optical density at each pixel, and this measurement becomes the digital value for that pixel. The aggregate of the digital values for all the pixels on the film being digitized is thus a digital representation of the film's image. The digitizer sends this digital version of the image via an electronic interface.
| MODULE | DESCRIPTION | MAIN COMPONENTS |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Laser | Supplies light beam of<br>appropriate size,<br>wavelength, and power to<br>the Scan. | Laser diode<br>Diode modulator board |
| Scan | Takes light beam from<br>Laser and scans it across<br>film in x-direction. | Spinning polygon w/built-<br>in controller<br>Lenses, mirrors |
| Transport | Detects and aligns film;<br>moves film through system. | Film sensor<br>Film rollers<br>Stepper motor |
| Light<br>Collection | Detects transmitted light;<br>performs A/D conversion. | Photodiodes<br>Diffusion screen<br>ADC board |
| Controller | Provides image data path;<br>generates pixel clock for<br>Light Collection; drives<br>stepper motor for Transport;<br>monitors sensors; controls<br>interface I/O. | Controller board |
| Software | Code for all Controller<br>functions. | PROM-based firmware |
| Power Supply | Provides electrical power<br>for all modules. +5 VDC for<br>digital circuits. ±5 and +<br>24VDC for analog circuits. | 5/24V linear supply<br>(2) 5V linear supplies |
| Enclosure | Provides light-tight, EMI-<br>ontrolled, clean<br>nvironment with reasonable<br>ser and service access. | Skins<br>Film input slot<br>Film exit bin<br>User interface<br>AC and data cables |
The table below describes and identifies the main components of the modules that make up the LD2000 SERIES digitizer.
{2}------------------------------------------------
2.7
# LD2000 SERIES Specifications
| Light source | Visible semiconductor diode |
|-----------------------|-------------------------------------------------------------------------------------------------------------|
| Spot size | 85 - 170 µm |
| Image Detectors | |
| Type | Solid State |
| Quantity | Nine |
| Image Size | |
| Film Size: | 8" to 14" wide |
| Scan Size: | 14.0" x 17.0" min |
| Pixels Not On Film: | Set to Max OD value |
| Scan time (14" x 17") | 12 seconds typical for 2K resolution |
| Spatial Resolution | |
| Number of pixels: | 2048 x 2480<br>1024 x 1240<br>other pixel resolutions configurable up to<br>4096 x 4974 on a 14" x 17" film |
| Pixel Size: | ~170 µm @ 2048 pixels on 14"<br>~85 µm @ 4096 pixels on 14" |
| MTF @ 6 lp/mm: | 0.4 (85 µm pixel at 4096 pixels on 14") |
| Optical Density | |
| OD Range: | 0.03 to 3.5 |
| OD Resolution | OD .03 - 2.5: .01 OD<br>OD 2.5 - 3.0: .02 OD<br>OD 3.0 - 3.5: .05 OD |
| Interface | SCSI<br>(Custom Configurations available per Customer Specifications) |
| Grey scale resolution | 12 bits (4096) |
| Power requirements | |
| Voltage: | 87 - 132 / 200 - 264; 4 position user selected |
| Frequency: | 47 - 63 Hz Single Phase |
| Current Rating: | 650 mA |
{3}------------------------------------------------
Temperature
operating non-operating 15C to 40C -40C to 60C
Relative Humidity operating non-operating
25-70% non-condensing 10-90%
Dimensions with pedestal wall mount
11"D x 22.25"W x 40" H 60 - 77 lbs. (with pedestal)
11"D x 22.25"W x 48.5" H
Weight
2.7
{4}------------------------------------------------
| 2.8 | Substantial Equivalence Comparison and Chart<br>The General Scanning LD2000 SERIES is substantially equivalent to all<br>of the Lumisys, Nishimoto Sangyo and Konica digitizers listed.<br>Specifically, |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Size: | The General Scanning model LD2000 SERIES is slightly larger<br>in size than the other products listed. This is primarily due to a<br>stand which is offered with the product. |
| Weight: | General Scanning's product weight is 60 - 77 lbs. similar to the<br>Lumisys 85LF at 75 lbs. |
| Power: | All units offer dual voltage (120/240 VAC). |
| Scan size: | Maximum scan size for all listed digitizers is 14" x 17". |
| Spot size: | General Scanning's product has an 85um spot which falls<br>within the 50 - 420um spot offered by the Lumisys 100. |
| Density range: | Our density range of 0.03 - 3.5 OD is within the range offered<br>by all products listed. |
| Grey Scale: | All products listed except Konica offer 12 bit grey scale. |
| Laser source: | Konica and both Nishimoto Sangyo products use semi-<br>conductor laser sources. |
| Beam scan: | Konica and both Nishimoto Sangyo products use a polygon<br>scanner for scanning the x-axis of the image. |
| Maximum resolution: | General Scanning's product is exactly equal to the Nishimoto<br>Sangyo ED-3500, and is within the range offered in the<br>Lumisys 85LF. |
| Interface: | General Scanning is offering a standard SCSI or custom<br>interface. This is equivalent to the products listed. |
The following page show a chart highlighting the substantial equivalence of the LD2000 SERIES to the Lumiscan 85LF, Lumiscan100, Konica, Nishimoto Sangyo ED-2000 and ED-3500 Film Digitizer models.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, serif font. The text is centered on the image and is the only element present.
MAR 19 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Steven Zis Quality Systems Manager General Scanning, Inc. (GSI) Optical Scanning Products Division 500 Arsenal Street Watertown, MA 02172
Re:
K980565 LD2000 Series Dated: February 12, 1998 Received: February 13, 1998 Regulatory class: Unclassified Procode: 90 LMA
Dear Mr. Zis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we bare determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and problibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Felecal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Proctice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Statement of Indications for Use 1.1
The General Scanning model LD2000 SERIES laser film digitizer is intended to convert medical images stored on film to digital data for copying, storage, and transmission. It is intended for use to digitize copies of conventional x-rays and films produced by imagers that fall in the ranges for size and optical density range detailed in the specifications.
The digitizer utilizes laser light to scan the film, extracts the image data by measuring the transmitted light, converts it into digital form, and transmits the data over a digital link. The digitized information may then be transmitted to a Laser Imager for printed hardcopy or to a Work Station for use in Teleradiology (transmission of images) or PACS (Picture Archiving Communication Systems).
The LD2000 SERIES contains the scanning laser and detection module, a film transport, and electronics to convert and transmit the image data and provide control and calibration.
David A. Seymour
(Division Sign-Off)
Division of Reproductive, Abdominal, EN and Radiological De 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109) ✓
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.