K980547 · Nidek, Inc. · GEX · May 13, 1998 · General, Plastic Surgery
Device Facts
Record ID
K980547
Device Name
NIDEK PRIMA MODEL GYC-1500/2000
Applicant
Nidek, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 13, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nidek Frequency Doubled Nd: YAG Laser is intended to be used in all clinical applications for which an Argon Laser would be used in ophthalmic surgery, including but not limited to Retinal and Macular Photocoagulation; Iridotomy and Trabeculoplasty.
Device Story
Nidek GYC-1500/2000 are solid-state frequency-doubled Nd:YAG lasers producing 532nm green light via harmonic generation of a 1064nm infrared beam through a non-linear optical crystal. A 633nm HeNe laser provides an aiming beam. Operated by physicians in clinical settings, the device delivers treatment via slit lamp microscope, binocular indirect microscope, or fiber optic end-ocular probe. Clinicians select power, exposure duration, and spot size; accessory contact or aspheric lenses may be used. The system facilitates targeted tissue photocoagulation, allowing for precise ocular procedures. By providing a controlled laser source for retinal/macular treatment and glaucoma-related procedures (iridotomy/trabeculoplasty), the device enables therapeutic intervention for various ophthalmic conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use similarities to predicate devices.
Technological Characteristics
Solid-state frequency-doubled Nd:YAG laser; 532nm treatment wavelength; 633nm HeNe aiming beam; non-linear optical crystal for harmonic generation. Max power: 1.5W (GYC-1500) and 2.0W (GYC-2000). Delivery via slit lamp, binocular indirect microscope, or fiber optic end-ocular probe. Compatible with accessory contact/aspheric lenses.
Indications for Use
Indicated for ophthalmic surgery applications requiring laser photocoagulation, including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the NIDEK Incorporated logo. The logo consists of a stylized eye on the left and the text "NIDEK Incorporated" on the right. The eye is a simple, black and white design with a large pupil. The text is in a simple, sans-serif font.
47651 Westinghouse Drive Fremont, California 94539
X 1510) 226-5750
# MAY | 3 1998
K980Σ47
ובסוגוסטונים
는공
510(k) SUMMARY
This is a 510(k) Summary in accordance with CFR 807.92.
- Submitter: 1.
Nidek Inc. for Nidek Co., Ltd., 34-14 Maehama Hiroishicho Gamagori, 443
Correspondent: Ken Kato, VP (510)226-5700 Phone: Fax: (510)226-5750
- 2. Device Name: Models GYC-1500 and GYC-2000 Solid state ophthalmic laser photocoagulator Ophthalmic Laser
- 3. Predicate Devices:
Model Ophthalas 532 by Alcon (K962592)
Model Cerlas G by CeramOptec (K962948)
Model Oculight GL by Iris (K960971)
Models GYC-1500/2000 (K951079) - The purpose of this application is to add two indications of: iridotomy and trabeculoplasty, to the current 510(k). GYC-1500 has max. power of 1.5W and 2000 of 2.0W.
- Intended Use: র্বা
The Nidek Frequency Doubled Nd: YAG Laser is intended to be used in all clinical applications for which an Argon Laser would be used in ophthalmic surgery, including but not limited to Retinal and Macular Photocoagulation; Iridotomy
{1}------------------------------------------------
and Trabeculoplasty.
## 5. Device Description:
The Nidek GYC-1500/2000 Solid State Photocoagulators are frequency doubled Nd: YAG lasers producing a characteristic 532 nanometer (nm) wavelength light as A 633nm helium-neon (HeNe) laser is used for the aiming beam treatment beam. source.
The doubling process of the 1064nm wavelength results when the infrared beam goes through a special crystal, which is an optical dielectric that exhibits a non-The 532nm - Green - wavelength is produced by linear optical response. harmonic generation of the 1064nm laser beam.
The laser treatment beam is delivered through the cornea to the target tissue by an optical delivery system which is mounted either on a slit lamp microscope or binocular indirect microscope. Tissue may also be treated inter-operatively by direct radiation from a fiber optic end-ocular probe or a optical deliver system mounted on an operating microscope. The beam may be further manipulated by accessory contact lenses or near contact "aspheric" lenses. An aiming beam is provided by the system to indicate the target tissue prior to the treatment exposure.
The desired treatment parameters of power exposure duration, and spot size are selected as required. The location and operation of all controls and indicators is described in the Nidek Prima Frequency Doubled Nd:YAG Operator Manual. Treatments are always started with the lowest applicable exposure and the dosage is increased until the desired effect is obtained.
## ર્જ Significant Changes/Modifications from Predicate Device:
There are no significant changes or modifications from the predicate products that affects safety, effectiveness, or the intended use of the product.
- 7. Device Labels:
Copies of advertising brochure, operator's manual with product label information are attached.
### 8. Comparative Information:
Comparison table is attached to show the similarities and differences of the GYC
{2}------------------------------------------------
to the predicate devices.
### 9. Software Validation & Verification:
As the part of its quality system controls, Nidek has implemented the software development process which is described and defined in the Software Development Procedure. Changes on software are made in accordance with the Design Change Standard Procedure. These procedures address the requirements for software specification, design change control, and design verification and validation. At each phase, design reviews are conducted to ensure that policies and procedures are being followed and that requirements are being met.
During the development phase, overall design requirements (specifications) and software specification are developed. Based on the system requirements and software specifications, a hazard (risk) analysis is conducted and, where necessary, methods of mitigation defined. At the design review, reviewers verify that specification effectively address design requirements.
During the coding phase, the software specifications are translated, with the aid of appropriate software tools, into source, object and executable code. At the design review, reviewers, using procedures such as code walk through, verify that code address elements defined in the specifications.
During the module and integration testing phase, software emulation and prototyping tools are utilized to test module and larger sections of code. After module testing, the software is integrated with the target system and validation This testing ensures that specifications and system testing conducted. requirements have been met.
approval/release review, Engineering and Quality Assurance During the management verifies that required documentation is present and approved. The verification and validation reports are reviewed to assure that system specifications and requirements have been met.
These established policies and procedures ensure that current and future software development projects, including changes, will be verified and validated against appropriate software and system requirements and specifications.
{3}------------------------------------------------
Signed: Ken Kato, VP
(
Date: FEB 11 1998
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 1998
Mr. Ken Kato ·Vice President Nidek Incorporated 47651 Westinghouse Drive Fremont, California 94539
Re: K980547 Nidek Prima Model GYC-1500/2000 Solid State Trade Name: Photocoagulator Requlatory Class: II Product Code: GEX February 11, 1998 Dated: February 12, 1998 Received:
Dear Mr. Kato:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸੈਂ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{5}------------------------------------------------
Page 2 - Mr. Kato
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
2
of _ I l Page
510(k) Number (if known): _ K980547
Device Name:
GYC-1500/2000
Indications For Use:
(Pcr 21 CFR 801.109)
The Nidek Prima Frequency doubled Nd: YAG Laser is intended to be used in all clinical applications for which an Argon Laser would be used in ophthalmic surgery, including Retinal and Macular Photocoagulation; Iridotomy and Trabeculoplasty.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of General Restorative Devices | K980847 |
| 510(k) Number | |
| Prescription Use | X |
|------------------|----------------------|
| OR | Over-The-Counter Use |
(Optional Format 1-2-96)
..
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.