THE NUFFIELD TOTAL KNEE SYSTEM

K980545 · Corin USA · JWH · Apr 29, 1998 · Orthopedic

Device Facts

Record IDK980545
Device NameTHE NUFFIELD TOTAL KNEE SYSTEM
ApplicantCorin USA
Product CodeJWH · Orthopedic
Decision DateApr 29, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Relief of pain and restoration of knee function following the effects of osteo-, rheumatoid and inflammatory arthritis, post-trauma disease effects, avascular necrosis, and total knee revision with or without varus, valgus or flexion deformity. The tibial and femoral components are intended for use only with bone cement.

Device Story

The Nuffield Total Knee System is a line extension of a previously cleared knee replacement system. It consists of cobalt-chrome alloy femoral, tibial, and patellar components. The devices feature a plasma-sprayed coating to increase surface roughness, facilitating mechanical interlock with PMMA bone cement. The system is designed for use in primary or revision knee arthroplasty to restore function and relieve pain. It is intended for use only with bone cement; non-cemented fixation is not permitted. The device is implanted by orthopedic surgeons in a hospital setting. By providing a stable, cemented interface, the device aims to restore knee joint mechanics and improve patient mobility.

Clinical Evidence

No clinical data provided. The submission relies on extensive mechanical test data and eight years of clinical use outside the United States with no reported post-operative problems.

Technological Characteristics

Materials: Cobalt-chrome alloy. Surface treatment: Plasma-sprayed coating for increased roughness. Fixation: Cemented (PMMA) only. Components: Femoral (stemmed, stem-less, pegged), tibial (stemmed), and patellar components. Design: Four-part knee system compatible with UHMWPE tibial inserts.

Indications for Use

Indicated for patients requiring total knee arthroplasty or revision surgery due to degenerative, rheumatoid, or inflammatory arthritis, post-traumatic disease, or avascular necrosis. Contraindicated for non-cemented use.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K9805215 ## APR 2 9 1998 # 510(k) SUMMARY ### Name of Company: Corin Medical Ltd The Corinium Centre Cirencester Gloucestershire GL7 1 YJ United Kingdom The Nuffield Total Knee System Name of Device: The devices which are the subject of this submission constitute a line Device Description: extension to the previously cleared Nuffield Total Knee System which has been on the market since 1989. This submission addresses those Nuffield Total Knee System devices which are plasma sprayed. The devices restore knee function in patients with degenerative rheumatoid or osteoarthritis, and are suitable for primary arthroplasty or revision surgery. The plasma coated devices which are the subject of this submission are: - i. NTK Femurs (stemmed, stem-less and pegged) - NTK Tibias (stemmed) ii. - iii. Patellas These devices are used in conjunction with previously cleared NTK devices, ie UHMPE Tibial Inserts, all-UHMWPE Tibias and Patellas. The devices which are the subject of this submission are manufactured from cobalt-chrome alloy and are plasma sprayed; the plasma coating provides an increased surface roughness to act as a key for mechanical interlock with PMMA Bone Cement. Although the Nuffield Total Knee is designed to spare the posterior cruciate ligament, the stability of the replacement relies on neither the anterior cruciate ligament and the devices can be used when both ligaments are absent. The Plasma Coated Nuffield Total Knee System is intended for use ONL Y with bone cement. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a handwritten string of characters, which appears to be a combination of letters and numbers. The string reads 'K9805215'. The characters are written in a slightly rough, uneven style, suggesting they were written quickly or with a writing instrument that doesn't produce a perfectly smooth line. The image is in black and white. #### SAFETY AND SUBSTANTIAL EQUIVALENCE AND SUMMARY OF EFFECTIVENESS The Plasma Coated Nuffield Total Knee System is substantially equivalent to both the Duracon Knee System, promoted by Howmedica in the US since 1993, and the Nuffield Total Knee System, promoted in the US by Corin Medical since 1993. The reasons for this are summarised below: - All are four-part knees utilising cobalt chrome alloy femoral and tibial components, ત. ultra high molecular weight polyethylene tibial inserts and metal-backed and allpolyethylene patella components. - b. All components have a range of sizes and differing thicknesses of polyethylene. - All are designed for cemented use ONLY. C. - d. All contain within the range an all-polyethylene monoblock tibial component for low demand patients. This submission is supported by extensive mechanical test data confirming the devices' safety and effectiveness. The Plasma Coated Nuffield Total Knee System has been in clinical use outside the United States for eight years and no post-operative problems have been reported. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract image of a person. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Craig Corrance ·President Corin U.S.A. 10500 University Center Drive Suite 190 Tampa, Florida 33612 Re : K980545 The Nuffield Total Knee System (Ti Plasma Spray) Requlatory Class: II Product Code: JWH Dated: February 2, 1998 February 12, 1998 Received: Dear Mr. Corrance: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal This decision is based on Food, Drug, and Cosmetic Act (Act). this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone cement." You may, therefore, market your device subject to the general controls provisions of the Act and the following limitations: - This device may not be labeled or promoted for non-1. cemented use. - 2. All labeling for this device, including package label and labeling included within the package, must prominently state that the device is intended for cemented use only. - Any non-cemented fixation of this device is considered 3. investigational and may only be investigated as a siqnificant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the device for noncemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Druq Administration (FDA) to conduct the investigation. {3}------------------------------------------------ Page 2 - Mr. Craig Corrance The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory -------In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807:97);" Other general information on your responsibilities under the Act may be {4}------------------------------------------------ Page 3 - Mr. Craig Corrance obtained from the Division of Small Manufacturers Assistance obtained from the Divibion or 6mm-2041 or (301) 443-6597 or at at Ics Coll Free nambor (6:5)/www.fda.gov/cdrh/dsmamain.html". Sincerely yours, focalefa Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K980545 Device Name: The Nuffield Total Knee System ### INDICATIONS FOR USE Relief of pain and restoration of knee function following the effects of osteo-, rheumatoid and inflammatory arthritis, post-trauma disease effects, avascular necrosis, and total knee revision with or without varus, valgus or flexion deformity. The tibial and femoral components are intended for use only with bone cement. ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation Prescription Use (per 21 CFR 801.109) ri OR Over-The-Counter Use p colpo (Division Sign-Off) Division of General Restorative Devices 510(k) Number_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...