CARPENTER-EDWARDS CLASSIC RING WITH DURAFLO TREATMENT MODELS 4425 (MITRAL) AND 4525 (TRICUSPID), CARPENTER-EDWARDS PHYSM
Applicant
Baxter Edwards
Product Code
KRH · Cardiovascular
Decision Date
Jul 7, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Annuloplasty rings are intended for use in patients to correct annular dilatation, increase leaflet coaptation, reinforce annular suture lines, and prevent further dilatation of the annulus.
Device Story
Implantable annuloplasty rings (mitral and tricuspid) used to treat annular dilatation; surgically implanted by cardiac surgeons. Devices provide structural support to heart valve annulus to improve leaflet coaptation and prevent further dilation. Models include Carpentier-Edwards Classic, Physio, and Cosgrove-Edwards systems. All models feature Duraflo treatment (heparin-based) to enhance biocompatibility. Devices are constructed from titanium alloy, siloxane polymer rubber, polyester knit/velour fabric, or silicone rubber with barium sulfate. Implantation occurs during open-heart surgery; device remains in situ to provide permanent mechanical reinforcement of the valve annulus.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing including material components evaluation, biocompatibility testing, packaging integrity testing, validation of 100% ethylene oxide sterilization, and heparin leachout studies, supplemented by animal data.
Technological Characteristics
Materials: titanium alloy, siloxane polymer rubber, polyester knit/velour fabric, silicone rubber with barium sulfate. Duraflo treatment applied to surfaces. Sterilization: 100% ethylene oxide. Form factor: ring configurations (mitral/tricuspid) matching anatomical orifices.
Indications for Use
Indicated for patients requiring correction of annular dilatation, increased leaflet coaptation, reinforcement of annular suture lines, and prevention of further annular dilatation.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
Predicate Devices
Carpentier-Edwards Classic Annuloplasty Ring, Models 4400 and 4500
Carpentier-Edwards Physio Annuloplasty Ring, Model 4450
Cosgrove-Edwards Annuloplasty System, Model 4600
Submission Summary (Full Text)
{0}------------------------------------------------
980487
Page 244
# Summary of Safety and Efficacy APPENDIX X
- JUL 7 1998
510(k) Summary
# NAME & ADDRESS OF CONTACT PERSON A.
Larry Wood Baxter Healthcare Corporation CardioVascular Group Edwards CVS Division 17221 Red Hill Avenue Irvine, CA 92614-5686
#### DEVICE NAME B.
- 1. Carpentier-Edwards Classic™ Ring with Duraflo® Treatment, Models 4425 and 4525
- 2. Carpentier-Edwards Physio® Annuloplasty Ring with Duraflo® Treatment, Model 4475
- 3. Cosgrove-Edwards® Annuloplasty System with Duraflo® Treatment, Model 4625
#### C. PREDICATE DEVICE NAME
- Carpentier-Edwards® Classic™ Annuloplasty Ring, Models 4400 and 4500. 1.
- 2. Carpentier-Edwards Physio® Annuloplasty Ring, Model 4450
- 3. Cosgrove-Edwards® Annuloplasty System, Model 4600
# DEVICE DESCRIPTION D.
- 1. The Carpentier-Edwards Classic™ Ring with Duraflo® Treatment (Models 4425 and 4525) is constructed of titanium alloy and has a sewing ring margin that consists of a layer of siloxane polymer rubber covered with a polyester knit fabric.
The mitral ring. Model 4425, conforms to the configuration of a normal mitral orifice. It is kidney-shaped with one long curved segment corresponding to the posterior leaflet. The ring is open in the rectilinear portion corresponding to the anterior leaflet and has transverse colored threads indicating the anterior and posterior commissures.
The oval tricuspid ring, Model 4525, conforms to the configuration of a normal tricuspid orifice. The ring has one rectilinear segment corresponding to the septal leaflet and one long curved segment corresponding to the anterior and posterior leaflets. The ring is open at the anteroseptal commissure. Transverse colored threads indicate the site of the anteroposterior and posteroseptal commissures.
{1}------------------------------------------------
- 2. The Carpentier-Edwards Physio® Annuloplasty Ring with Duraflo® Treatment, Model 4450, is constructed of Elgiloy® bands separated by polyester film strips and has a sewing ring margin that consists of a layer of siloxane polymer tubing covered with a woven polyester cloth.
- The Cosgrove-Edwards® Annuloplasty System with Duraflo® Treatment, Model 3. 4600, is composed of a silicone rubber strip compounded with barium sulfate to improve radiographic visualization. This silicone rubber strip is covered with woven polyester velour cloth wrapped around the silicone and sewn together with a single seam using polyester thread.
#### ய் INTENDED USE
The intended use of Baxter's annuloplasty ring products with Duraflo® Treatment has not changed from the predicate devices. Annuloplasty rings are intended for use in patients to correct annular dilatation, increase leaflet coaptation, reinforce annular suture lines, and prevent further dilatation of the annulus.
### TESTING SUMMARY F.
The following studies were conducted to qualify Baxter's annuloplasty ring product lines with Duraflo® Treatment and 100% EO Sterilization.
- Material Components Evaluation .
- Biocompatability Testing .
- Packaging Integrity Testing .
- Validation of 100% Ethylene Oxide Sterilization .
- . Heparin Leachout Studies
- Animal Data .
All testing conducted showed no adverse effects of the Duraflo®-treated annuloplasty rings undergoing 100% EO sterilization, resulting in a safe product for use in an implantable application.
# G. SUBSTANTIAL EQUIVALENCE SUMMARY
The claim of substantial equivalence between the predicate and proposed devices is based on the following:
- 1. The predicate and proposed devices are intended for the same use.
- 2. The design of each predicate and proposed device are the same.
- 3. The component materials used to manufacture each of the rings have not changed except for the addition of the Duraflo® Treatment in the proposed rings.
{2}------------------------------------------------
- 4. The packaging materials are the same as those used in the currently marketed annuloplasty models (predicate devices).
- Note: The term "substantial equivalence" as outlined in this pre-market notification and the supporting information pertaining to equivalence are intended only to demonstrate equivalence to the predicate products for purposes of obtaining clearance of the device pursuant to the Federal Food, Drug and Cosmetic Act. Reference to equivalence as outlined above is in no way related to the term "equivalent" or similar terminology as outlined under the patent laws.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines extending from their faces, resembling a bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 7 1998
Mr. Larry Wood Manager, Regulatory Affairs Baxter Healthcare Corporation CardioVascular Group 17221 Red Hill Avenue (Irvine) P.O. Box 11150 Santa Ana, CA 92711-1150
Re: K980487 Heparin-Treated Annuloplasty Ring Model Numbers 4425 (mitral), 4525 (Tricuspid), 4475 (mitral), and 4625 (mitral and tricuspid) Requlatory Class: III Product Code: 74 KRH Dated: February 6, 1998 February 9, 1998 Received:
Dear Mr. Wood:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Be advised that any promotional or advertisement materials must be consistent with the labeling in the submission (i.e., bear the bolded statement regarding the lack of long-term clinical effectiveness data).
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance
{4}------------------------------------------------
# Page 2 - Mr. Larry Wood
with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html."
Sincerely yours,
Thomas J. Collelon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
4
:3分:
510(k) Number (if known): K980487
Device Name: Baxter Annuloplasty Ring Product Lines (Models 4425, 4525, 4475, and 4625)
Indications For Use: The intended uses of Baxter's Annuloplasty ring products with Duraflo Treatment are intended for use in patients to correct annular dilatation, increase leaflet coaptation, reinforce annular suture lines and prevent further dilatation of the annulus.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bete R. Bumperle
(Division Sign-Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K990487
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.