CRYSTAL AND CRYSTAL-SEAL IMPLANTS SYSTEM

K980442 · Crystal Medical Technology · DZE · Mar 16, 1998 · Dental

Device Facts

Record IDK980442
Device NameCRYSTAL AND CRYSTAL-SEAL IMPLANTS SYSTEM
ApplicantCrystal Medical Technology
Product CodeDZE · Dental
Decision DateMar 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

These devices are intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices and to restore the patient's chewing function.

Device Story

Post-type endosseous dental implants; surgically placed in upper or lower jaw bone; provide support for prosthetic devices; restore chewing function. Manufactured from high-purity titanium or titanium alloy; available in uncoated, hydroxyapatite-coated, or titanium plasma spray-coated forms. Used by dental professionals in clinical settings. Safety and effectiveness supported by historical clinical use of similar titanium/titanium alloy screw-type implants and literature reviews (e.g., Branemark technique).

Clinical Evidence

No clinical data generated for this specific device. Safety and effectiveness established via literature review of endosseous titanium implants, including the Toronto 10-year study and NIH Consensus Development Conference proceedings, confirming predictable outcomes for edentulous patients.

Technological Characteristics

Post-type endosseous dental implants. Materials: High-purity titanium (99+%) or titanium alloy (ASTM F 136-84). Surface options: Uncoated, hydroxyapatite-coated, or titanium plasma spray-coated. Mechanical design: Screw/post configuration for surgical placement in jaw bone.

Indications for Use

Indicated for patients requiring support for prosthetic devices (e.g., artificial teeth) in the upper or lower jaw arches to restore chewing function.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # XIV. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA. (Separate Page) A. Submitter: Clint Folsom, Crystal Medical Technology, Inc., 181 Cahaba Valley Pkwy, Pelham, AL 35124. Registration No. 51805. I. Classification Name and Number: Endosseous Implant (76DZE), II. Common/Usual Name: Dental Implant, Endosseous, Post (or screw)-type, titanium or titanium alloy. III. Proprietary Name: CRYSTAL and CRYSTAL-SEAL Implants System IV. Classification: This device is being finally classified by the Dental Devices Panel (Title 21 CFR 872.3640. V. Performance standards: None applicable. Materials meet ASTM voluntary standards VI. Description: The CRYSTAL Implants System implants are post-type endosseous dental implants with design and manufacturing concepts, materials, surgical procedures, and intended uses quite similar to the preamendment device and to others rated substantially equivalent to the preamendment device. This CRYSTAL-SEAL system is most similar to the Crystal and Crystal Plus Implants System which received concurrence of substantial equivalence from the Food and Drug Administration in premarket notification submission K-954432. Like many commercial implants, the CRYSTAL devices are manufactured of high-purity (99+%) titanium, or titanium alloy (ASTM F 136-84 titanium - 6 aluminum - 4 vanadium). See Appendix V for composition and characteristics of the titanium and/or titanium alloy used. CRYSTAL Implants also are available in coated form, either hydroxyapatite or titanium plasma spray. These devices are equivalent to various coated endosseous implants on the market, for example the Sabertech implants of Genentech. Inc., which were cleared under K-924112. VII. Labels and Instructions for Use are provided, as are labels for competitive products. VIII. Intended Use: These devices are intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices and to restore the patient's chewing function. VII. Substantial Equivalence: This device is equivalent to devices manufactured and sold before 1976, having a U. S. classification number of 76 DZE, and those described under 21 CFR 872.3640. It is also equivalent to several devices currently on the market that have been determined by the FDA to be substantially equivalent to the above preamendment devices, particularly to Crystal and Crystal Plus System cleared by K-954432. Some examples of other equivalent products are: {1}------------------------------------------------ Bofors Nobelpharma, Goteborg, Sweden: Titanium Implant Device, K-820013, Core-Vent Corp., Encino, CA: Screw-Vent Endosseous Implants, K-861426, A, B. Impla-med, Inc., Sunrise, FL: P.M.T. Surgical Implant Components, K-892124A, (Titanium Grade 3, ASTM F67-83); K-921854. Steri-Oss (Subs. Denar, Anaheim, CA.), Denar Dental Implant, K-852802; Steri-Oss Implant, K-884845. Artech, Inc., Chantilly, VA: Artech Submerged Screw Implant, K-891346, K-895267. Odontit, SA, Buenos Aires, Argentina, Osseo. Implants System, K-915375. Genentech, Seattle WA: K-924112 VIII. Clinical Discussion and Brief Literature Review: Endosseous implants, and especially those of titanium or titanium alloy, in the "post" or "screw" configuration, have been proved safe and effective through the years. The possible adverse effects summarized in this 510(k) cover those listed by the United States classification panel [Federal Register, vol. 45, No. 251, pp 86025-6, Dec. 30, 1980], as well as to those revealed in a recent literature search. Matukas, "Medical Risks Associated with Dental Implants," states, "Little or no hard data could be found on the medical risks associated with (dental] implants." Because of the wide-spread usage of dental implants, Smith and Zarb made a careful review of the literature and proposed criteria for implant success. A thorough computerized Medline literature search produced 585 entries. An update of this search produced 68 new review articles. The Journal of Dental Education published a special issue "Proceedings of the Consensus Development Conference on Dental Implants [National Institutes of Health, Bethesda, MD, June 13-15, 1988], Vol. 52, No. 12, pp. 677-831, Dec. 1988. This added to the literature search above, with some especially pertinent reprints from the scientific literature, provide a comprehensive summary of available scientific data. Zarb completed his report of the detailed Toronto 10-year study by concluding that "the tried and tested Branemark implant technique has revolutionized the treatment options open to the prosthodontist. For the edentulous patient ... the prospect for a lifetime of restored oral comfort, function, and appearance have now become predictable and reliable." These results are ample evidence of the safety and effectiveness of these endosseous implants. END OF 510(k) SUMMARY {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 6 1998 Mr. Clint Folsom ·President Crystal Medical Technology 181 Cahaba Valley Parkway Pelham, Alabama 35124 Re : K980442 Crystal and Crystal-Seal Implants System Trade Name: Requlatory Class: III Product Code: DZE January 30, 1998 Dated: February 4, 1998 Received: Dear Mr. Folsom: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance withthe Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Mr. Folsom This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97) . Other gener Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Patricia Cosenza Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### IX. Indications for Use: {Separate Page} 90444 510(k) Number: NA Device Name: Crystal Seal™ This device is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED #### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use or (Per 21 CFR 801/109) (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number ka80442. Over-The-Counter Use *No* (Optional Format 1-2-96) 7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...