FLEXIBLE FIBERSCOPES AND ACCESSORIES

K980401 · Richard Wolf Medical Instruments Corp. · FAJ · May 21, 1998 · Gastroenterology, Urology

Device Facts

Record IDK980401
Device NameFLEXIBLE FIBERSCOPES AND ACCESSORIES
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeFAJ · Gastroenterology, Urology
Decision DateMay 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Flexible fiberscopes are used for examination, diagnosis, and/or therapy in connection with endoscopic accessories and auxiliary instruments used through the working channel of the instrument for use in urology, surgery, and OB/GYN.

Device Story

Flexible fiberscopes consist of a flexible insertion tube, control handle, and eyepiece. Device transmits images via objective lens and fiber bundle for direct visualization or video camera attachment. Tip features active deflection (up to 360°) with optional locking brake. Working channel allows insertion of auxiliary instruments (e.g., grasping forceps, stone extractors, HF electrodes) and simultaneous irrigation. Used by physicians in urology, surgery, and OB/GYN for diagnostic and therapeutic procedures. Output is visual image; clinical decision-making relies on physician interpretation of anatomy and pathology observed during procedure. Benefits include minimally invasive access to internal structures for diagnosis and intervention.

Clinical Evidence

No clinical tests were performed. Bench testing was conducted to verify compliance with ANSI/AAMI HF18 and IEC601-1/IEC601-2-2 standards for high-frequency devices.

Technological Characteristics

Flexible endoscope with active 360° tip deflection and locking mechanism. Image transmission via objective lens and fiber bundle. Features wide working channel for irrigation and auxiliary instrument passage. Compatible with video camera systems. Sterilization via gas valve attachment. No software or electronic processing components.

Indications for Use

Indicated for patients requiring endoscopic examination, diagnosis, or therapy in urology, surgery, and OB/GYN settings.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 353 Corporatio Woods Parkway Vernon Hills, Illinois 60061 Phone: 847.913.1113 Fax: 847.913.1488 K980401 **RICHARD WOLF** MEDICAL INSTRUMENTS CORPORATION ![](https://i.imgur.com/9999999.png) 142 MAY 2 1 1999 | 510(k) Summary of Safety and Effectiveness | MAY 21 1998 | |--------------------------------------------|-------------| |--------------------------------------------|-------------| | Date of Preparation: | May 1, 1998 | |----------------------------------------|----------------------------------------| | Company / Institution name: | RICHARD WOLF MEDICAL INSTRUMENTS CORP. | | FDA establishment registration number: | 14 184 79 | | Division name (if applicable): | N.A. | | Phone number (include area code): | (847) 913-1113 | | Street address: | 353 Corporate Woods Parkway | | FAX number (include area code): | (847) 913-0924 | | City: | Vernon Hills | | State/Province: | Illinois | | Country: | USA | | ZIP / Postal Code: | 60061 | | Contact name: | Mr. Robert L. Casarsa | | Contact title: | Quality Assurance Manager | Product Information: | Trade name: | Flexible fiberscopes and accessories | |----------------------|--------------------------------------| | Model number: | 7305.xxx | | Common name: | Flexible fiberscopes | | Classification name: | Flexible endoscopes | Information on devices to which substantial equivalence is claimed: | | 510(k) Number | Trade or proprietary or model name | Manufacturer | |---|---------------|------------------------------------|--------------| | 1 | K962172 | Flexible Mini-Fiberscopes | Richard Wolf | | 2 | | Flexible CystoNephroscope ACN | Circon/ ACMI | | 3 | | Flexible Hysteroscope, AUR-FH | Circon/ACMI | | 4 | | Cysto-Urethro-Fiberscope 11272 AA | Karl Storz | | 5 | | Flexible Hysteroscope, 11261 BB | Karl Storz | | 6 | | Flexible Cystofiberscope CYF-3 | Olympus | | 7 | | Hysterofiberscope, HYF-1T | Olympus | ## 1.0 Description The flexible fiberscopes consist of a flexible insertion part, a control part and an eyepiece for direct view or connection to a video camera. {1}------------------------------------------------ MEDICAL INSTR K980401 2.12 ### 2.0 Intended Use Flexible fiberscopes are used for examination, diagnosis, and/or therapy in connection with endoscopic accessories and auxiliary instruments used through the working channel of the instrument for use in urology, surgery, and OB/GYN. #### 3.0 Technological Characteristics There are no significant technological changes or characteristics to the new devices compared to the existing devices. The tip of the sheath has an active deflection up to 360°, depending on the fiberscope. The deflecting position can be locked with a brake at the 7305.001. The image is transmitted via objective, fiber bundle and eyepiece for direct view or connection to a video camera. The total number of pixels and the fibers per mm2 are increased to get an image with higher resolution. Biopsy material can be taken by the wide working channel, which is simultaneously used for irrigation. Auxiliary instruments such as grasping forceps, stone extractors or HF button electrode are inserted via a proximal mounted insertion cock with supply and discharge. A leakage test unit or a gas sterilization valve can be connected to an attachment of the fiberscope. #### 4.0 Substantial Equivalence The submitted devices pose the same type of questions about safety or effectiveness as the compared devices. The new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing pre-enactment and 510(k) devices sold by Richard Wolf in the urology branch. In addition, the submitted devices are substantially equivalent to devices sold by Circon, Karl Storz and Olympus. #### 5.0 Performance Data No known FDA performance standard exists. The electrodes in combination with the fiberscopes were tested to meet the appropriate sections of the ANSI/ AAMI standard on high frequency devices HF18 and IEC601-1/ IEC601-2-2. #### 6.0 Clinical Tests Clinical tests performed were not performed. ### 7.0 Conclusions Drawn These devices are designed and tested to assure their safety and effectiveness when used according to the instructions manual. **By:** Robert L Casasa Robert L. Casarsa Quality Assurance Manager **Date:** May 1, 1998 9 - 2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN" is arranged in a circular fashion around the logo. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Robert L. Casarsa Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 MAY 2 | 1998 Re: K980401 Flexible Fiberscopes and Accessories Dated: May 1, 1998 Received: May 4, 1998 Regulatory Class: II 21 CFR §876.1500 and §884.1690 Product Code: 78 FAJ GCJ, and 84 HIH Dear Mr. Casarsa: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, narket the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the pronotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrb/dsmaldsmam.html". Sincerely yours, Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radial -------------------------------------------------------------------------------------------------------------------------------------------------------------------Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K980401 Device Name: __ Flexible Fiberscopes and Accessories (7305.001, 7305.011) # Intended Use: For examination, diagnosis and/or therapy in connection with endoscopic accessories and auxiliary instruments used through the working channel of the instrument for use in urology, surgery, and GYN. (PLEASE DO NOT WRITE BELOW TIJIS LINE - CONTINUE ON ANOTHER PAGE IF NEFFICED) Robert R. Satting/ (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K984401 Prescription Use Per 21 CFR 801.109 の 『『』『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『『』『 Over-The Counter_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...