← Product Code [GEX](/productcode/GEX) · K980398

# KAPLAN PENDULASER 115 SURGICAL LASER WITH OPTOSCAN II (K980398)

_Optomedic Medical Technologies , Ltd. · GEX · Sep 22, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K980398

## Device Facts

- **Applicant:** Optomedic Medical Technologies , Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Sep 22, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner accessory is indicated for surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: Dermatology, Plastic Surgery, Podiatry, Neurosurgery, Gynecology, Otorhinolaryngology, Arthroscopy (knee), Open & Endoscopic General Surgery. The Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner accessory is safe and effective when indicated for use in specific surgical applications including: - Ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of laser skin resurfacing, laser derm-abrasion, and laser burn debridement. - Laser skin resurfacing (ablation and/or vaporization) in dermatology and plastic surgery for the treatment of wrinkles, rhytids, and furrows. - Laser skin resurfacing (ablation and/or vaporization) of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts). - Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery for the performance of blepharoplasty.

## Device Story

CO2 surgical laser system; includes laser unit, scanner controller box, footswitch adapter, scanner, and handpiece. Device delivers laser energy to target soft tissue for ablation, vaporization, excision, incision, or coagulation. Operated by physicians in clinical settings (dermatology, plastic surgery, general surgery, etc.). System combines previously cleared laser and scanner components to provide precise tissue interaction. Output is controlled laser energy; enables surgical procedures like skin resurfacing and lesion removal. Benefits include controlled tissue destruction with minimal thermal damage to surrounding areas.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and predicate comparison.

## Technological Characteristics

CO2 surgical laser system. Includes laser source, scanner controller, footswitch, and handpiece. Operates via CO2 laser energy for soft tissue interaction. System is a combination of previously cleared laser and scanner modules.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Coherent UltraPulse 5000 system ([K963339](/device/K963339.md))
- Sharplan SilkTouch ([K860087](/device/K860087.md))
- Sharplan 1020 Laser ([K960521](/device/K960521.md), [K961935](/device/K961935.md))
- Laser component ([K950313](/device/K950313.md))
- Scanner component ([K964684](/device/K964684.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SEP 2 2 1998

K980398

## 510(k) Summary

# Kaplan PenduLaser 115 CO2 Surgical Laser with OptoScan II Scanner Accessory

Common/Classification Name: Laser Surgery Device, 21 CFR 878.4810 Optomedic Medical Technologies Yoni Netanyaho 3-B Or-Yehuda 60376 ISRAEL

Contact: Alex Harel, Prepared: January 29, 1998

#### LEGALLY MARKETED PREDICATE DEVICES A.

The Kaplan PenduLaser 115 CO2 Surgical Laser with OptoScan II Scanner Accessory is substantially equivalent to the Coherent UltraPulse 5000 system cleared under K963339 and to the Sharplan SilkTouch/Sharplan 1020 Laser cleared under K860087, K960521, and K961935. The new device is actually a combination of two products already cleared under K950313 (the laser) and K964684 (the scanner), and so the new device can also be considered to be substantially equivalent in parts to these two devices.

#### DEVICE DESCRIPTION B.

The Kaplan PenduLaser 115 CO, Surgical Laser with OptoScan II Scanner Accessory consists of the laser, the Scanner Controller Box, a Footswitch Adapter box, the Sahar SofTouch Scanner, and the handpiece. The Controller houses the electronics and power supply and interfaces to the Pendulaser for communication through the Footswitch Adaptor Box.

#### C. INTENDED USE

The Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner accessory is indicated for surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: Dermatology, Plastic Surgery, Podiatry, Neurosurgery, Gynecology, Otorhinolaryngology, Arthroscopy (knee), Open & Endoscopic General Surgery.

11:50 PM

{1}------------------------------------------------

The Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner accessory is safe and effective when indicated for use in specific surgical applications including:

- Ablation, vaporization, excision, incision, and coagulation of soft . tissue in dermatology and plastic surgery in the performance of laser skin resurfacing, laser derm-abrasion, and laser burn debridement.
- Laser skin resurfacing (ablation and/or vaporization) in dermatology . and plastic surgery for the treatment of wrinkles, rhytids, and furrows.
- . Laser skin resurfacing (ablation and/or vaporization) of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surqical scars, keloids, hemangiomas (including buccal hemangiomas}, tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts).
- Laser incision and/or excision of soft tissue in dermatology, plastic . and general surgery for the performance of blepharoplasty.

#### SUBSTANTIAL EQUIVALENCE SUMMARY D.

The KAPLAN PenduLaser 115 Surgical Laser with OptoScan II Scanner Accessory is a medical device, and it has the same indications for use and target population as the legally marketed predicate devices.

The KAPLAN PenduLaser 115 Surgical Laser with OptoScan II Scanner Accessory has the same technological characteristics as the predicate devices and these characteristics are sufficiently precise to ensure equivalence. This premarket notification has described the characteristics of the KAPLAN PenduLaser 115 Surgical Laser with OptoScan II Scanner Accessory in sufficient detail to assure substantial equivalence.

#### E. TECHNOLOGICAL CHARACTERISTICS

The technological characteristics of the combined new device are unchanged from those of the component devices and are the same as those of the predicate devices.

{2}------------------------------------------------

#### CONCLUSIONS F.

This pre-market submission has demonstrated Substantial Eguivalence, as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health, to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/11 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized symbol that resembles an abstract human figure or a bird in flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 2 1998

Optomedic Medical Technologies Limited c/o Whit Athey, Ph.D. C.L. McIntosh & Associates 12300 Twinbrook Parkway, Suite 625 Rockville, Maryland 20852

Re: K980398

Trade Name: Kaplan PenduLaser 115 CO2 Surgical Laser with OptoScan II Regulatory Class: II Product Code: GEX Dated: August 20, 1998 Received: August 20, 1998

Dear Dr. Athey:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions; or other Federal laws or -----------------------------------------------------------------------------------------regulations.

{4}------------------------------------------------

Page 2 - Whit Athey, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Jocelyn

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): k980398

#### Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner Device Name:

Indications For Use:

The Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner accessory is indicated for surgical applications requiring the ablation, vaporization, excision, and coagulation of soft tissue in medical specialties including: Dermatology, Plastic Surgery, Podiatry, Neurosurgery, Gynecology, Otorhinolaryngology, Arthroscopy (knee), Open & Endoscopic General Surgery.

The Kaplan PenduLaser 115 Surgical Laser with OptoScan II Scanner accessory is safe and effective when indicated for use in specific surgical applications including:

- · Ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of laser skin resurfacing, laser derm-abrasion, and laser burn debridement.
- · Laser skin resurfacing (ablation and/or vaporization) in dermatology and plastic surgery for the treatment of wrinkles, rhytids, and furrows.
- · Laser skin resurfacing (ablation and/or vaporization) of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis. solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts).
- · Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery for the performance of blepharoplasty.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                         |
|--------------------------------------------------------|-------------------------|
| (Division Sign-Off)                                    |                         |
| Division of General Restorative Devices                |                         |
| 510(k) Number                                          | K980398                 |
| Prescription Use (Per 21 CFR 801.109)                  | OR Over-The-Counter Use |

000032

---

**Source:** [https://fda-staging.innolitics.com/device/K980398](https://fda-staging.innolitics.com/device/K980398)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
