For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to dsDNA in human serum by enzyme immunoassay, as an aid in the diagnosis systemic lupus erythematosus (SLE). For manual use, or for use with the HyPrep System Plus. For In Vitro Diagnostic Use Only.
Device Story
SeraQuest Anti-dsDNA is a solid-phase enzyme immunoassay (EIA) for detecting IgG antibodies against dsDNA in human serum. The device uses microwells coated with dsDNA; patient serum is incubated, allowing anti-dsDNA antibodies to bind. After washing, an enzyme-labeled conjugate (goat anti-human IgG) is added, followed by an enzyme substrate. The substrate converts to a yellow end-product, measured photometrically at 405 nm. The test is performed manually or using the HyPrep System Plus semi-automated fluid handler in a clinical laboratory setting. Results are reported as IU/ml or Index values, aiding clinicians in diagnosing systemic lupus erythematosus (SLE).
Clinical Evidence
Clinical performance evaluated by testing 215 serum specimens using both SeraQuest and Shield DIASTAT assays. Excluding equivocal results, the study demonstrated a relative sensitivity of 92.8% (95% CI: 86.6-98.9%) and a relative specificity of 97.7% (95% CI: 95.1-100%). Overall agreement was 96.0% (95% CI: 93.2-98.7%).
Technological Characteristics
Solid-phase enzyme immunoassay (EIA) using plastic microwells coated with calf thymus DNA. Employs goat anti-human IgG conjugated with alkaline phosphatase. Detection via p-Nitrophenyl phosphate substrate, measured photometrically at 405 nm. Manual or semi-automated (HyPrep System Plus) operation. Room temperature incubation.
Indications for Use
Indicated for the qualitative, semi-quantitative, and quantitative detection of IgG antibodies to dsDNA in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE). For in vitro diagnostic use.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
Shield DIASTAT Anti-dsDNA kit
Submission Summary (Full Text)
{0}------------------------------------------------
APPENDIX 3. (revised 4/10/98)
APR 2 4 1998
510(k) SUMMARY
K980391
| Applicant: | Quest International, Inc.<br>1938 N.E. 148th Terrace<br>North Miami, FL 33181 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration No. | 1061839 |
| Contact Person: | Robert A. Cort, V.P. , Quality Assurance |
| Telephone: | (305) 948-8788 |
| Telefax: | (305) 948-4876 |
| Manufacturing Site: | Same as above |
| Device: | SeraQuest® Anti-dsDNA |
| Device Name: | Anti-nuclear antibody immunological test system (21CFR § 866.5100)<br>Systemic lupus erythematosus immunological test system<br>(21CFR § 866.5620) |
Tevice Classification: Class II (performance standards)
# Description:
The SeraQuest Anti-dsDNA test is a solid-phase enzyme immunoassay (EIA), which is performed in microwells, at room temperature, in three thirty minute incubations. It has been developed to detect IgG antibodies which are directed against dsDNA, in human serum.
The Calibrators in the SeraQuest Anti-dsDNA test set have been assigned IU/ml values which are traceable to the WHO First International Standard for Anti-Double-Stranded DNA, Wo-80, and Index values which based on an in-house standard anti-dsDNA serum. Test results are reported as IU/ml or as Index values.
# Principle:
Diluted samples are incubated in wells coated with dsDNA antibodies against ds DNA (if present) are immobilized in the wells. Residual sample is eliminated by washing and conjugate (enzyme-labeled antibodies to human IgG) is added and incubated. If IgG antibodies to DSDNA are present, the conjugate will be immobilized in the wells. Residual conjugate is eliminated by washing and draining, and the enzyme substrate is added and incubated. In the presence of the enzyme, the substrate is converted to a yellow end-product which is read photometrically at 405 nm.
{1}------------------------------------------------
#### ntended Use:
(
For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to dsDNA in human serum by enzyme immunoassay, as an aid in the diagnosis systemic lupus erythematosus (SLE). For manual use, or for use with the HyPrep System Plus. For In Vitro Diagnostic Use Only.
#### Predicate Device:
The SeraQuest Anti-dsDNA test is substantially equivalent in intended use and performance, to the Shield DIASTAT Anti-dsDNA kit, Shield Diagnostics, Dundee, UK.
#### Summary of technological characteristics:
| Characteristic | SeraQuest Anti-dsDNA | Shield Anti-dsDNA |
|--------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------|
| Description: | Enzyme Immunoassay | Enzyme Immunoassay |
| Intended Use: | The detection of IgG<br>antibodies against dsDNA<br>in human serum. | The detection of IgG & IgM<br>antibodies against dsDNA<br>in human serum. |
| Solid Phase: | Plastic Microwell | Plastic Microwell |
| Antigen: | Calf Thymus DNA | Calf Thymus DNA |
| Number of Incubation Periods: | Three | Three |
| Sample Dilution: | 1:50 | 1:100 |
| Sample Incubation<br>Duration: | 30 minutes | 60 minutes |
| Incubation Temperature: | Room temperature | Room temperature |
| Ezyme-labeled Conjugate: | | |
| Antibody | Goat anti-human IgG | Goat anti-human IgG & IgM |
| Enzyme | Alkaline phosphatase | Alkaline phosphatase |
| Conjugate Volume: | 100 µl | 100 µl |
| Conjugate Incubation Duration: | 30 minutes | 30 minutes |
| Substrate: | p-Nitrophenyl<br>phosphate | phenolphthalein<br>monophosphate |
{2}------------------------------------------------
| Subtrate Volume: | 100 µl | 100 µl |
|-----------------------------------|------------------------------|------------------------------|
| Substrate Incubation<br>Duration: | 30 minutes | 30 minutes |
| Stop Reagent: | 0.5 M Trisodium<br>phosphate | Sodium Hydroxide |
| Stop Reagent Volume: | 100 µl | 100 µl |
| Readout: | Spectrophotometric<br>405 nm | Spectrophotometric<br>550 nm |
### Summary of Clinical Testing:
Of 215 specimens tested, 64 were positive, 126 were negative and 1 was equivocal in both the SeraQuest and Shield tests (Table 1 below). Of the 25 remaining specimens were equivocal in the SeraQuest test and of these, 2 were positive and 3 were negative by the Shield test: 11 specimens were equivocal by the Shield test and of these, 4 were positive and 7 were negative by the SeraQuest test; 3 specimens which were positive by the SeraQuest test were negative by the Shield test, and 5 specimens which were positive by the Shield test were negative by the SeraQuest test.
### ГАВLE 1.
## RESULTS OF SeraQuest ANTI-dsDNA ASSAYS AND SHIELD DIASTAT ANTI-dsDNA ASSAYS OF 215 SERUM SPECIMENS, PERFORMED AT QUEST INTERNATIONAL, INC. MIAMI, FL.
| SHIELD<br>DIASTAT<br>ANTI-dsDNA | | SeraQuest ANTI-dsDNA | | | | % | 95% CIv |
|---------------------------------|----------|----------------------|----------|-----------------------|------|--------------|---------|
| | Positive | Equivocal | Negative | | | | |
| Positive | 64 {36} | 2 | 5 | Relative sensitivity* | 92.8 | 86.6 to 98.9 | |
| Equivocal | 4 {2} | 1 | 7 {1} | | | | |
| Negative | 3 {1} | 3 {2} | 126 {8} | Relative specificity* | 97.7 | 95.1 to 100 | |
| Overall agreement* | | | | | 96.0 | 93.2 to 98.7 | |
{ } Number of patients with clinically diagnosed SLE.
* Exeluding equivasel results
- Excluding equivocal results.
Calculated by the normal method.
{3}------------------------------------------------
APR 2 4 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Robert A. Cort Vice President, Quality Assurance Quest International, Inc. 1938 N.E. 148th Terrace North Miami, Florida 33181
Re : K980391/S1 Trade Name: SeraQuest® Anti-dsDNA Requlatory Class: II Product Code: LRM Dated: April 14, 1998 Received: April 17, 1998
Dear Mr. Cort:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in. the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, . "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. 137
#### APPENDIX 6 (revised 4/10/98)
Page _________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): K980391
Device Name: _ SeraQuest Anti-dsDNA
Indications For Use:
- For in vitro diagnostic use only. 1.
- For the qualitative, semi-quantitative and quantitative detection of IgG antibodies 2. to dsDNA in human serum by enzyme immunoassay.
- For use as an aid in the diagnosis of systemic lupus erythematosus (SLE). 3.
ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
- For manual use, or for use with the HyPrep System Plus semi-automated fluid 4. handler.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
[Prescription Use](https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/prescription-labeling)
(Per 21 CFR 801.109) OR [Over-The-Counter Use](https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/over-counter-otc-devices)
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.