K980375 · Applied Imaging Corp. · JOY · Jun 9, 1998 · Hematology
Device Facts
Record ID
K980375
Device Name
WINSCAN AUTOMATED IMAGING SYSTEM
Applicant
Applied Imaging Corp.
Product Code
JOY · Hematology
Decision Date
Jun 9, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5260
Device Class
Class 2
Indications for Use
WinSCAN is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC).
Device Story
WinSCAN is an automated imaging system for peripheral blood smear analysis. It utilizes a motorized microscope with brightfield optics (10x-50x objectives) and a camera to scan slides, locate nucleated cells, and capture color images. The system is operated by laboratory personnel via an Intel-based PC running Windows 3.1/DOS 6.0. The software coordinates motorized stage/focus/filter movements to acquire and archive cell images. The operator reviews these images on a monitor and performs the final classification of cells as nRBC or WBC based on visual characteristics (size, shape, light adsorption). The system aids in the enumeration of nRBCs per 100 WBCs. It includes a manual microscope mode for verification. The device benefits patients by automating the labor-intensive search for rare nucleated cells, improving efficiency and repeatability compared to manual microscopy.
Clinical Evidence
Bench testing only. Comparison to standard manual microscopy showed a correlation coefficient > 0.98 for nRBC/100 WBC counts. Precision testing across multiple samples, instruments, and operators yielded a mean coefficient of variation (CV) of 24%-26%, compared to 23% for manual microscopy.
Technological Characteristics
Automated imaging system consisting of an Intel-based PC, motorized microscope (10x, 20x, 40x, 50x objectives), camera, and image acquisition board. Features motorized stage, filter wheels, and focus drive. Connectivity via PC interface. Software-controlled image acquisition and archiving. Operates on Windows 3.1/DOS 6.0.
Indications for Use
Indicated for use in clinical laboratories to scan peripheral blood smears to assist operators in identifying and enumerating nucleated red blood cells (nRBC) per 100 white blood cells (WBC).
Regulatory Classification
Identification
An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)
{0}------------------------------------------------
K980375
## Summary of Safety and Effectiveness
#### Submitter
Applied Imaging 2380 Walsh Avenue, Building B Santa Clara, California 95051 Phone: 408-562-0250 Fax: 408-562-0264
#### Date summary was prepared
January 27, 1998
#### Name of the device
WinSCAN Automated Imaging System
#### Identification of predicate device
Micro 21 System (also known as IMS-200): K925670
#### Description of the device
The WinSCAN System is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC). The operator selects the recommended parameters for nucleated cells, starts the scan and can walk away. The system locates and stores color images of the nucleated cells for each slide, the operator reviews the data on the WinSCAN monitor, and classifies the nucleated cells as nRBC or WBC based on stain and nuclear characteristics. Identification parameters include light adsorption, size and shape.
{1}------------------------------------------------
Automatic relocation, capture and archiving of the cell images are performed by the instrument based upon operator selection. The instrument also can be used in the manual mode to systematically scan a slide and examine each field.
The WinSCAN System consists of the following components:
- Intel-based PC with Windows operating system .
- Monitor, kevboard, mouse .
- Switchstick control unit t
- Color printer for images and texts .
- Microscope with brightfield capability and 10x, 20x, 40x and 50x . objectives
- Set of transmission and excitation filters .
- Camera and image acquisition board t
- Motorized devices: stage, filter wheels, focus drive 0
- Motor controller unit and manual controls .
The Intel-based PC operates the instrument through the software program that coordinates control of the automated microscope. The base operating system of the PC consists of MicroSoft Windows 3.1 and DOS version 6.0. When the WinSCAN icon is selected from the Program Manager, WinSCAN software is downloaded and operation of the system can begin. The WinSCAN program includes all facilities to scan, relocate, and acquire images from a microscope slide. The WinSCAN software also performs image acquisition and scan functions, as well as archiving and relocation of images and facilitates operator review of cells. Classification of cells is performed by the operator.
The Monitor. Kevboard, Mouse, and switchstick control unit provide user interfaces to the system. The monitor displays all information to the user, including menus for navigating through the WinSCAN operation, image acquisition results and allows for operator visual review for operator classification of the nucleated cells found on a slide. The keyboard allows for user entry of scan parameters, slide numbers, user identification, comments, and annotation of results. The mouse/keyboard is used to interact with the instrument and software, including selecting menus and functions of the WinSCAN software and instrument.
A color printer is included with the system that allows the user to print results of scanning or acquired images of cells.
{2}------------------------------------------------
### Intended Use
The WinSCAN System is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC) and enumerate them as #nRBC/100 white blood cells (WBC).
### Comparison of device characteristics to predicate
The WinSCAN instrument is substantially equivalent to the Micro 21 system. Both instruments can automatically scan a slide, both instruments allow for operator review, and both instruments allow for viewing images on a computer monitor. Unlike the Micro 21 system, the WinSCAN system includes a manual microscope that allows the user to view slides manually for verification. Both systems can relocate a cell or object on a slide, even upon its removal and replacement.
### Testing
Results of studies demonstrate that WinSCAN is an effective tool that aids the operator in locating and classifying nucleated cells. The test results demonstrate that the WinSCAN is able to scan slides prepared from peripheral blood and aid the operator in locating white blood cells and nucleated red blood cells. When compared to the standard manual microscope method, the correlation coefficient is >/= 0.98. Additionally, precision testing using replicate counts of 30 samples demonstrates that WinSCAN performs in a repeatable manner, with a mean coefficient of variation of 24% - 26% over multiple samples, instruments and operators, as compared to 23% CV% precision demonstrated by manual microscopy.
### Conclusion
The WinSCAN instrument is substantially equivalent to the Micro 21 system. Both instruments can automatically scan a slide for nucleated cells, both instruments allow for operator review, and both instruments allow for viewing images on a computer monitor. Unlike the Micro 21, WinSCAN includes a manual microscope that the user can use to view nucleated cells on a slide manually for verification.
Both systems can relocate a cell or object on a slide, even upon its removal and replacement. Both systems require operator interaction for cell classification and for the enumeration of nRBCs/100 WBCs.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. The logo is black and white and has a simple, clean design. The eagle is facing to the right.
Public Health Service
JUN - 9 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Patricia A. Milbank, JD, RAC Vice President, Requlatory Affairs and Quality Assurance Applied Imaqinq 2380 Walsh Avenue, Bldg. B Santa Clara, California 95051
K980375/S1 Re : WinSCAN Automated Imaging System Trade Name: Requlatory Class: II Product Code: JOY Dated: May 6, 1998 Received: May 7, 1998
Dear Ms. Milbank:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure ............
{5}------------------------------------------------
Page of of ___
| 510(k) Number (if known): | K980375 |
|---------------------------|----------------------------------|
| Device Name: | WinSCAN Automated Imaging System |
# Indications For Use:
Intended Use
WinSCAN is an automated imaging system that scans peripheral blood smears for nucleated blood cells. WinSCAN software provides for operator interaction and decision making to identify nucleated red blood cells (nRBC), and enumerate them as #nRBC/100 white blood cells (WBC).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Deborah Morie
for Dr. Peter Massim
(Division Sign-Ofm
Division of Clinical Laboratory Devices
510(k) Numbor.
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.