K980356 · Canadian Medical Brush, Inc. · HHT · Apr 16, 1998 · Obstetrics/Gynecology
Device Facts
Record ID
K980356
Device Name
CERVICAL CYTOLOGY C-BRUSH MODEL CMB-1010
Applicant
Canadian Medical Brush, Inc.
Product Code
HHT · Obstetrics/Gynecology
Decision Date
Apr 16, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The Cervical Cytology C-Brush is intended for the collection of cervical cells for Pap smear analysis and /or for sexually transmitted disease (STD) testing. The Cervical Cytology C-Brush is not intended for use in pregnant women.
Device Story
Cervical Cytology C-Brush is a manual specimen collection tool; features truncated cone-shaped nylon brush head on wire shaft bonded to plastic handle. Clinician inserts brush 1.0-1.5 cm into cervical os; rotates two full turns counter-clockwise; withdraws. Collected cells are transferred to glass slide for Pap smear or STD testing. Device facilitates diagnostic sampling; aids in detection of cervical abnormalities or infections. Used in clinical settings by healthcare providers.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design similarity to predicate.
Technological Characteristics
Truncated cone-shaped nylon bristles; wire shaft; plastic handle. Manual operation. Non-sterile or sterile (implied by medical device classification).
Indications for Use
Indicated for endocervical specimen collection for Pap smear, chlamydia, and HPV testing. Contraindicated for use in pregnant women.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
Team Technologies, Inc. - Cytology Brush (K971849)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a black and white drawing of a caduceus. The caduceus is a symbol often associated with medicine and healthcare. It features a staff with two snakes coiled around it, and a pair of wings at the top.
Canadian Medical Brush Inc. Mississauga, Ontario L5S 1A3 Canada 2299 Drew Road, unit # 6
APR 1 6 1998
# 510(k) SUMMARY
K980356 PI 72
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared.
| Submitter: | Canadian Medical Brush Inc. |
|------------|------------------------------|
| | 2299 Drew road, Unit #6 |
| | Mississauga, Ontario L5S 1A3 |
| | Canada |
| Phone: | (905) 405-1075 |
| Fax: | (905) 405-1108 |
| E-mail: | cbrush@globalserve.net |
Contact Person: Stephen Jemuovic, Director, Regulatory Affairs Canadian Medical Brush Inc Phone: (905) 405- 1323 (905) 405- 1108 Fax:
# Name of Device and Name / Address of Sponsor
Name of Device: Cervical Cytology C-Brush
Sponsor: Canadian Medical Brush Inc. 2299 Drew Road, Unit #6 Mississauga, Ontario L5S 1A3 Canada
Common or Usual Name:
Specimen (cervical) collection brush
Classification Name:
C ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
DACE
Spatula, cervical, cytological
Phone: (905) 405-1075 Fax: (905) 405-1108 E-mail: cbrush@globalserve.net
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a caduceus symbol, which is often associated with medicine and healthcare. The symbol features a staff with two snakes coiled around it, and a pair of wings at the top. The staff is vertical and the snakes are intertwined, creating a symmetrical design. The wings are spread out, adding to the overall visual balance of the symbol.
Canadian Medical Brush Inc. 2299 Drew Road, unit # 6 Canada
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
1980356 02012
# 510(K) SUMMARY - PAGE 2
Predicate Device:
Team Technologies, Inc. - Cytology Brush (K971849)
#### Device Description:
The Cervical Cytology C-Brush has a truncated cone-shaped brush head composed of nylon bristles joined to a wire shaft. The brush head on the wire shaft is bonded to a plastic handle. To collect cervical cells, the brush end of the Cervical Cytology C-Brush is inserted 1.0 to 1.5 cm into the cervical os until the longest outer bristles of the brush touch the ectocervix. The brush is rotated two full turns in counter - clockwise direction and withdrawn. The brush is then used to spread the specimen onto a glass slide for a Pap smear.
# Intended Use:
The Cervical Cytology C-Brush is intended for the collection of cervical cells for Pap smear analysis and /or for sexually transmitted disease (STD) testing. The Cervical Cytology C-Brush is not intended for use in pregnant women.
### Technological Characteristics and Substantial Equivalence
The Cervical Cytology C-Brush is substantially equivalent in terms of design, materials, manufacturing processes, physical Intended use, properties, and principles of operation to the Team Technologies, Inc. Cytology Brush.
E-mail: cbrush@globalserve.net Phone: (905) 405-1075 Fax: (905) 405-1108 -1 -1 -5 1) 11
{2}------------------------------------------------
APR 1 6 1998
Re: K980356
Cytology Cervical C-Brush
Received: March 31, 1998
21 CFR 884.4530/Procode: 85 HHT
Dated: March 23, 1998
Regulatory Class: II
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Stephen Jemuovic Director Canadian Medical Brush, Inc. 2299 Drew Road, unit # 6 Mississauga, Ontario L5S 1A3 CANADA
Dear Mr. Jemuovic:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalience compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510%) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odrh/dsmaldsmamain.html".
Sincerely yours.
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):
Cervical Cytology C-Brush Device Name:
Indications For Use:
Indication for use are the endocervical collection of specimens used in Pap smear, chlamydia and HPV testing.
K980356
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Roder R. Rathbun
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 11 980356 510(k) Number _
Prescription Use > (Per 21 CFR 801.109)
PAGE 23
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.