WORKING ELEMENTS, RESECTOSCOPE SHEATHS, ELECTRODES, OBTURATORS, ENDOSCOPES
K980302 · Richard Wolf Medical Instruments Corp. · HIH · Jul 20, 1998 · Obstetrics/Gynecology
Device Facts
Record ID
K980302
Device Name
WORKING ELEMENTS, RESECTOSCOPE SHEATHS, ELECTRODES, OBTURATORS, ENDOSCOPES
Applicant
Richard Wolf Medical Instruments Corp.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jul 20, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Resectoscopes are used for endoscopically controlled ablation of tissue. They are used, in combination with endoscopic accessories, for examination, diagnosis, and / or therapy in various medical disciplines, such as urology and gynecology
Device Story
Resectoscopes consist of working elements, sheaths (standard/continuous-irrigation), electrodes (cutting/coagulation), obturators, and endoscopes. Device enables endoscopic visualization and tissue ablation via high-frequency (HF) energy. Physician inserts sheath into body opening using obturator; endoscope provides visualization; working element carries electrodes to target site. Electrodes perform mechanical/electrosurgical cutting and coagulation of soft tissue (e.g., prostate, myoma). Used in OR/clinic settings by urologists, gynecologists, and surgeons. Design improvements include reduced spring force for ergonomic operation, ceramic insulation for wear resistance, titanium construction for weight reduction, and Hytrel insulation for sealing accuracy. Output is visual feedback for the clinician to guide tissue resection, facilitating minimally invasive surgical intervention and patient recovery.
Clinical Evidence
No clinical tests were performed. Safety and effectiveness were established through bench testing, including endurance tests and steam sterilization validation. HF components were tested for conformity to ANSI/AAMI HF 18, IEC 601-2-2, and IEC 601-2-18 standards.
Technological Characteristics
Resectoscope system comprising working elements, sheaths, electrodes, and obturators. Materials: Titanium (housing), Ceramic (insulation), Hytrel (electrode insulation), Radel (temperature-resistant components). Sensing/Actuation: Mechanical working element with spring-loaded electrode movement. Energy: High-frequency (HF) electrosurgical energy. Connectivity: None. Sterilization: Steam sterilization. Standards: ANSI/AAMI HF 18, IEC 601-2-2, IEC 601-2-18, 93/42/EEC.
Indications for Use
Indicated for endoscopic examination, diagnosis, and therapy of tissue in urology, gynecology, and surgery, including TURP, TURB, adenoma/myoma resection, soft tissue tumor removal, endometrium ablation, bladder neck slitting, and prostate incision.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Continuous Irrigation Resectoscope System, Model 8655 (K895857/A)
{0}------------------------------------------------
JUL 201998 K980302
ﺳﻌﺮ
일
|
P 1 9 3
353 Corporate Woods Parkway Vernon Hills, Illinois 60064 Phone: 847.943.1113 Fax: 847.913.1488
RICHARD WOLF
MEDICAL INSTRUMENTS CORPORATION
な
| 510(k) Summary of Safety and Effectiveness | | | | |
|-----------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------|---------------------------|
| Submitter: | | Date of Preparation:<br>January 23, 1998 | | |
| Company / Institution Name:<br>Richard Wolf Medical Instruments Corporation | | FDA establishment registration number:<br>1418479 | | |
| Division Name (if applicable):<br>N.A. | | Phone number (include area code):<br>(847) 913-1113 | | |
| Street Address:<br>353 Corporate Woods Parkway | | FAX number (include area code):<br>(847) 913-0924 | | |
| City:<br>Vernon Hills | State/Province:<br>Illinois | Country:<br>USA | | ZIP/Postal Code:<br>60061 |
| Contact Name:<br>Mr. Robert Casarsa | | | | |
| Contact Title:<br>Quality Assurance Manager | | | | |
| Product Information | | | | |
| Trade Name:<br>Resectoscopes E-Line | | Model Number:<br>8654, 8655, 8656, 8657, 8658, 8661,<br>8663, 8666, 8668, 8407 to 8439 | | |
| Common Name:<br>Resectoscope | | Classification Name:<br>Resectoscope,<br>Resectoscope, Working Element | | |
| Information on devices to which substantial equivalence is claimed: | | | | |
| 510(k) Number | Trade or proprietary or model name | | Manufacturer | |
| 1 K895857/A | 1 Continuous Irrigation Resectoscope<br>System, Model 8655 | | 1 Richard Wolf | |
| 2 K953983 | 2 EVAP Electrodes Model 8423,<br>8427, 8410, 8413 | | 2 Richard Wolf | |
| 3 | 3 USA Elite System Resectoscopes | | 3 Circon ACMI | |
| 4 | 4 Resectoscopes 27040-27050, 27143 | | 4 Karl Storz | |
{1}------------------------------------------------
981307
#### 1.0 Description
The resectoscopes consist of working elements, standard and continuous-irrigation resectoscope sheaths, coagulation and cutting electrodes, standard, timerlake (deflecting) and visual obturators, and endoscopes. The working element is the carrier of the endoscope and guide element to the electrodes.
The standard resectoscope sheath is inserted together with obturators in to the body opening to enable an atraumatic passage. After removal of the obturators, the shaft provides a guidance for the working element and rinsing charge, while the distal sheath simultaneously serves as a cutting-edge.
The obturator fills the hollow sheaths. The visual obturator in combination with the endoscope is used to insert the sheath under vision. The standard electrodes are used for combined cutting and coagulation of soft tissue, e.g. prostate or myoma.
The inside of the patient is visualized when the endoscope is inserted into the body through natural or surgically generated access.
#### 2.0 Intended Use
Resectoscopes are used for endoscopically controlled ablation of tissue.
Indication: For the examination, diagnosis and/ or therapy in combination with endoscopic accessories in the various medical disciplines, such as, urology, gynecology, and surgery.
Field of Application: TURP (Transurethral Resection of Prostate), TURB (Transurethral Resection of Bladder Tumors), adenomas, myoma resection, soft tissue tumor, endometrium ablation, as well as slitting of the neck of the bladder and incision of the prostate.
Long sheath resectoscopes are used with long urethras.
### 3.0 Technological Characteristics
The resectoscopes improved design and material achieves a more efficient operation.
The new E-Line working elements have a spring forced reduced by 55% from the earlier generation, while the sliding ability is improved. The resulting operation shows less signs of fatigue.
- The material used for the new E-Line resectoscope sheath insulation at the cut-off edge is ceramic rather than GFK which results in higher wear resistance. The submitted sheaths are made from Titan rather than brass / new silver which produces a housing approximately 20% lighter than its predecessor. Some of the continuous irrigation sheaths are oval which improve the flow-off rate.
- The submitted E-Line electrode insulation is made from Hytrel instead of Teflon which produces size accuracy and improves sealing.
Improved temperature resistance and processing is achieved by replacing Maranyl with Radel.
{2}------------------------------------------------
P3093
#### Substantial Equivalence 4.0
The submitted devices pose the same type of questions about safety and effectiveness as the compared devices. The new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing 510(k) devices sold by Richard Wolf, Circon, and Karl Storz.
#### Performance Data 5.0
No performance standards are known.
The devices conform to the relevant provisions of European Device Directive 93/42/EEC.
Endurance tests and steam sterilization tests were performed to guarantee that the devices are safe and effective.
The HF devices were tested for conformity with the specified standards ANSI/AAMI HF 18, IEC 601-2-2 / IEC 601-2-18.
#### 6.0 Clinical Tests
No clinical tests were performed.
#### 7.0 Conclusions Drawn
The devices were designed and tested to guarantee the safety and effectiveness when used according to the instruction manuals.
By: Robert J. Casale
Robert L. Casarsa Quality Assurance Manager
Date: Jon 23, 98
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
# JUL 20 1998
Mr. Robert L. Casarsa Manager of Quality Assurance Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, IL 60061
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Re: K980302
Resectoscopes, Instruments and Accessories E-Line (working elements, standard and continuous irrigation resectoscope sheaths, coagulation and cutting electrodes, and obturators) Dated: April 21, 1998 Received: April 22, 1998 Regulatory Class: II 21 CFR 884.1690/Procode: 85 HIH 21 CFR 876.1500/Procodes: 78 FJL 78 FDC 78 FAS
Dear Mr. Casarsa:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we bave determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the ensciment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cometic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrl/dsmaldsmamain.html".
Sincerely yours
Lillian Yin, Ph.D.
Director, Division of Reproducti Abdominal. Ear. Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K980302
Device Name:
Resectospes, Instruments and Accessories E-Line, consisting of
working channels, standard and continuous irrigation, resectoscope sheaths,
coagulation and cutting electrodes, standard, timberlake (deflecting) and
visual obturators, and endoscopes
Indications for Use: Intended Use:
Resectoscopes are used for endoscopically controlled ablation of tissue. They are used, in combination with endoscopic accessories, for examination, diagnosis, and / or therapy in various medical disciplines, such as urology and gynecology
Revised 7/15/98
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Peter Satling/
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices K980302 510(k) Number _
Prescription Use
Per CRF 21 CFR 801.109
ાર
Over-The Counter Use
2 - 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.