Bd Becton Dickinson Vacutainer Systems Preanalytic
Product Code
LLH · Microbiology
Decision Date
Jul 17, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2660
Device Class
Class 1
Indications for Use
The ColorPAC™ Toxin A test is indicated for the qualitative detection of Clostridium difficile toxin A (enterotoxin) in stool specimens from patients suspected of having C. difficile associated disease. The test can also be used for the confirmation of suspect colonies of toxigenic C. difficile from agar plates or BHI broth. The assay is intended for use as an aid in the diagnosis of C. difficile associated disease. The test is not indicated for asymptomatic adults or children. This device is not indicated for use in a point of care setting.
Device Story
Rapid chromatographic immunoassay for qualitative detection of C. difficile toxin A. Input: stool specimens, bacterial colonies from agar plates, or BHI broth. Principle: lateral flow chromatographic test strip with immobilized monoclonal anti-toxin A antibody. Sample diluted in buffer, applied to test device; toxin A binds to capture antibody. Detection via pink dye-labeled liposomes (Detector A) and secondary liposome reagent (Detector B) to enhance signal. Output: visual pink test line (positive) or absence of line (negative) alongside control line. Used in clinical laboratories by trained personnel. Aids diagnosis of C. difficile associated disease; provides rapid confirmation of toxigenic isolates.
Clinical Evidence
Clinical study compared ColorPAC Toxin A to Cytotoxin B across 760 specimens (fresh and frozen) at four sites. Combined sensitivity 81% (95% CI: 72.4%-87.3%), specificity 97% (95% CI: 95.4%-98.2%). Colony confirmation study (n=111) showed sensitivity 89-95% and specificity 93-95% depending on media (BHI broth, selective agar, anaerobic blood agar). Analytical studies included cross-reactivity, limit of detection (1.4-22.6 ng/ml), and interference testing.
Technological Characteristics
Chromatographic lateral flow immunoassay. Components: test strip with monoclonal anti-toxin A antibody, sample buffer (saline, mucolytic agent, 1% detergent), wash reagent (saline, bovine protein stabilizer, 0.1% detergent), and liposome-based detection reagents (Detector A/B). Preservative: 0.2% sodium azide. Manual visual read. Standalone device.
Indications for Use
Indicated for qualitative detection of C. difficile toxin A in stool specimens from patients suspected of C. difficile associated disease and confirmation of suspect colonies from agar/broth. Not indicated for asymptomatic adults/children or point-of-care use.
Regulatory Classification
Identification
A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
Predicate Devices
Cytotoxin B
Submission Summary (Full Text)
{0}------------------------------------------------
K980185
# JU 17 1998
# SUMMARY OF SAFETY AND EFFECTIVENESS
SUBMITTED BY:
Colleen Rohrbeck Becton Dickinson Microbiology Systems 7 Loveton Circle Sparks, MD 21152
NAME OF DEVICE:
| Trade Name: | ColorPAC™ Toxin A |
|--------------------------|---------------------------------------|
| Common Name/Description: | Reagents, Clostridium difficile Toxin |
PREDICATE DEVICE:
Cytotoxin B
## DEVICE DESCRIPTION:
INTENDED USE: ColorPAC™ Toxin A is a rapid chromatographic assay for the qualitative detection of Clostrictium difficile toxin A (enterotoxin) in stool specimens from patients suspected of having C. difficile associated disease. The test can also be used for the confirmation of suspect colonies of toxigenic C. difficile from agar plates or BHI broth. This assay is intended for use as an aid in the diagnosis of C. difficile associated disease.
## KIT CONTENTS:
The following components are included in the ColorPAC™ Toxin A Test kit.
| Reagent 1 | (30 ml) | Sample Buffer: Buffered saline with a mucolytic<br>agent, 1% detergent. |
|-----------|----------|--------------------------------------------------------------------------------------------|
| Reagent W | (2.3 ml) | Wash Reagent: Buffered saline solution with<br>bovine protein stabilizer, 0.1 % detergent. |
| Reagent 2 | (1.5 ml) | Detector A: Rabbit C. difficile toxin A antibody<br>coated liposomes. |
| Reagent 3 | (1.5 ml) | Detector B: Rabbit antibody coated liposomes<br>specific for Detector Reagent A. |
{1}------------------------------------------------
| Control + | (1.0 ml) | Positive Control: Buffered saline solution with<br>bovine protein stabilizer, inactivated C. difficile<br>toxin A. |
|-----------------|----------|--------------------------------------------------------------------------------------------------------------------|
| Control - | (1.0 ml) | Negative Control: Buffered saline solution with<br>bovine protein stabilizer. |
| | | Reagents and controls each contain 0.2%<br>sodium azide (preservative) |
| 30 Test Devices | | Each containing a test strip coated with<br>monoclonal C. difficile toxin A antibody and C.<br>difficile toxin A. |
30 SQ-EASY™ Tubes and Filter Tips, 30 Applicator Sticks, 30 Transfer Pipets, 1 Dropper Rod
PRINCIPLE OF PROCEDURE: The ColorPAC Toxin A test is comprised of a capture line consisting of C. difficile toxin A antibody immobilized on a chromatographic test strip which is contained within a test device. A specimen (stool, colonies, or BHI broth) is diluted with sample buffer and added to the sample well of the test device. The diluted sample wicks up the test strip by capillary action. Toxin A, if present, binds to the capture antibody at the test line as the specimen migrates across the test strip. The wash and remaining reagents are added to the reagent well of the test device.
Upon addition of Detector A, toxin A antibody coated liposomes containing a pink dye migrate across the test strip and attach to the C. difficile toxin A antigen that was bound to the test strip in the previous step. Color formation is enhanced upon addition of a second liposome reagent, Detector B, that binds to the complex formed between Detector A and sample antigen at the test line. Following the final wash, the reactions are read visually. If toxin A antigen is detected in the patient specimen, a pink test line and a pink control line will appear, indicating a positive result. In the absence of toxin A, only the pink control line will appear, indicating a negative result. The result is uninterpretable if there is no pink control line or if the background obscures reading of the control line.
62
{2}------------------------------------------------
# PERFORMANCE DATA:
# ANALYTICAL STUDIES:
- Cross-Reactivity 1.
The ColorPAC™ Toxin A test was evaluated for cross reactivity by The ColorPACT® Toxin A tost wateria, yeast, viruses and parasites) into toxin A positive and toxin A negative stool specimens to a final concentration of 107-10° CFUlmi for bacteria and yeast, 1082- - 1082 Concentration of 10 -10 -10 -10 - 10 - narasites/ml. The only organism shown to I City of the viruses and 10 penic isolate of Clostridium sordellii (VPI a oss-leact was a highly toxigente is Cowan strain ATCC 12598, which 9040). The Staphyloooooo aan cross reactivity. In addition, Escherichia produces protein A, and 43895 which produce Shiga-like toxins, did not show any cross reactivity.
#### 2. Limit of Detection
The analytical sensitivity of the ColorPAC Toxin A test was performed by The analytical sensitivity of a lo of toxin A into 5 speciments each of liquid, seemi-solid, and solid stools. Seeded specimens were tested in triplicate. The limit of detection range for liquid, semi-solid, and solid stool specimens was 1.4-5.2, 1.6-18.7, 3.2-22.6 ng/ml, respectively.
- Interfering Substances 3.
The evaluation of interfering substances was performed by seeding substances that may be present in stool specimens from patients substarious a lat may be provins associated disease. ColorPAC Toxin A test results are not affected by ampicillin, cephalexin, metronidazole, vancomycin, barium sulfate, laxatives, blood, and antidiarrheal medications.
- Stability of Stool Specimens with Known Amounts of Toxin A 4.
Multiple freeze thaw cycles (up to three) had no significant effect on the assay. A subset of clinical trial samples that were tested at one site demonstrated that specimens can be stored up to 72 hours at 2-8°C . If specimens are to be tested after 72 hours, they should be frozen at -70°C immediately upon receipt in the laboratory. Specimens may be frozen up to 2 months.
{3}------------------------------------------------
# CLINICAL STUDIES:
### ColorPAC Toxin A vs. Cytotoxin B 1.
The performance of the ColorPAC™ Toxin A test was determined in The performance of the Color Ao - Cannotest medical content . The sites were located in geographically diverse areas of the United States. A
sites were located in geographically diverse areas of the United with sites were located in geographically diversions from patients with total of 598 fresh and 162 frozen stocked disease were tested. Each site intigl its routine cytotoxin B assay to ColorPAC Toxin A. Results of this initial
its routine cytotoxin B assay to ColorPAC Toxin of this initial its routine cylotoxin D docuy to site are summarized below.
| Site | No. | ColorPAC<br>Toxin A | Positive | Negative | | 95%<br>Confidence<br>Intervals |
|----------|-----|---------------------|----------|----------|-----------------|--------------------------------|
| 1 | 240 | Positive | 28 | 6 | Sensitivity 82% | (65.5%,93.2%) |
| | | Negative | 6 | 200 | Specificity 97% | (93.8%,98.9%) |
| 2 | 135 | Positive | 18 | 1 | Sensitivity 90% | (68.3%,98.8%) |
| | | Negative | 2 | 114 | Specificity 99% | (95.3%,99.9%) |
| 3 | 207 | Positive | 21 | 11 | Sensitivity 81% | (60.7%,93.5%) |
| | | Negative | 5 | 170 | Specificity 94% | (89.4%,96.9%) |
| 4 | 178 | Positive | 29 | 1 | Sensitivity 74% | (57.9%,87.0%) |
| | | Negative | 10 | 138 | Specificity 99% | (96.1%,100%) |
| Combined | 760 | Positive | 96 | 19 | Sensitivity 81% | (72.4%,87.3%) |
| | | Negative | 23 | 622 | Specificity 97% | (95.4%,98.2%) |
13n
{4}------------------------------------------------
### Reproducibility: 2.
Reproducibility of the ColorPAC Toxin A test was evaluated by testing stool panels consisting of three levels of toxin A and a toxin A negative stool panels Consisting of through were tested in triplicate on each of antigen Control. Billious specifical trial sites. ColorPAC Toxin A
three days at each of four different clinical trial situsiity three days at back of loar and inter assay reproducibility.
### Colony Confirmation: 3.
A study was conducted to evaluate the ColorPAC™ Toxin A test A study was concident to evalual confirmation from suspected colony
performance for providing culture confirmation schoological va accr performance for providing strains of C. difficile using selective agar,
growth for toxin producing strains of C. difficile using broth . A tatal grown for toxin producing strains of O. Chilision (BH), broth. A fotal of 111
anaerobic blood agar, and Brain Heart Infusion (BH), broth, a ff. differily vore anaerobic blood agar, and Drain Houseteristics of C. difficile were
clinical isolates that morphological characteristics of C. difficile were cirflical (Solates that mor moring the ColorPAC Toxin A test. After identification of the colonies by biochemical methods, toxiganic status was determined by a cytotoxin B assay. Results from this initial was dotennino each medium are summarized below.
| Media Type | No.<br>Specimens | Sensitivity<br>95% Confidence<br>Intervals | Specificity<br>95% Confidence<br>Intervals |
|-------------------------|------------------|--------------------------------------------|--------------------------------------------|
| BHI Broth | 111 | 95% (84.9%,98.8%) | 93% (82.7%,98.0%) |
| Selective Agar | 111 | 95% (84.9%, 98.8%) | 93% (82.7%,98.0%) |
| Anaerobic Blood<br>Agar | 111 | 89% (77.7%,95.9%) | 95% (85.1%,98.9%) |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
July 17 1998
Colleen Rohrbeck Regulatory Affairs Associate Becton Dickinson Becton Dickinson Microbiology Systems 7 Loveton Circle Sparks, Maryland 21152
K980185 Re: Trade Name: ColorPAC™ Toxin A Regulatory Class: I Product Code: LLH Dated: May 13, 1998 Received: May 14, 1998
Dear Ms. Rohrbeck:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Nutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
### K980185
Device Name:
ColorPAC™ Toxin A
Indications for Use:
The ColorPAC™ Toxin A test is indicated for the qualitative detection of Clostridium difficile toxin A (enterntaxin) in stool specimens from patients used for the seefing C. difficile associated disease. The test can also be used for the confirmation of suspect colonies of toxigenic C. difficile from agar plates or BHI broth. The assay is intended for use as an aid in the diagnosis of C. difficile associated disease. The test is not indicated for asymptomatic adults or children. This device is not indicated for use in a point of care setting.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | 980185 |
| Prescriptive Use | X |
|---------------------|---|
| (Per 21CFR 801.109) | |
OR
| Over-the-Counter-Use |
|----------------------|
|----------------------|
| odiff.ind.rev. | (Optimal Format 1-2-96) |
|----------------|-------------------------|
|----------------|-------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.