# FLOW RIDER FLOW DIRECTED MICRO CATHETER (K980104)

_Micro Therapeutics, Inc. · KRA · Apr 10, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K980104

## Device Facts

- **Applicant:** Micro Therapeutics, Inc.
- **Product Code:** [KRA](/productcode/KRA.md)
- **Decision Date:** Apr 10, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The catheter is intended to be used for the controlled selective infusion of physician specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuro anatomy.

## Device Story

Flow Rider is a single-lumen infusion catheter designed for flow-directed navigation into distal vasculature. Features include a hydrophilic coating for lubricity and a radiopaque marker at the tip for fluoroscopic visualization. A proximal luer lock adapter allows for accessory attachment. Used by physicians in clinical settings to deliver pharmacologic agents or contrast media. The device facilitates targeted delivery, potentially improving therapeutic outcomes or diagnostic imaging in peripheral and neurovascular anatomy.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (dimensional, tensile, burst, flow, torque, simulated use) and animal studies assessing performance.

## Technological Characteristics

Single-lumen infusion catheter with hydrophilic coating and radiopaque tip marker. Proximal luer lock adapter. Biocompatibility evaluated per ISO 10993-1:1992(E).

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- Target Spinnaker™ Flow Directed Infusion Catheter ([K965189](/device/K965189.md))
- Balt Magic™ Infusion Catheter ([K923368](/device/K923368.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K980104

PR 10 1998

Image /page/0/Picture/2 description: The image shows the logo for Micro Therapeutics, Inc. The logo consists of the letters "MTI" in a bold, sans-serif font. To the left of the letters, there is a series of small squares that appear to be arranged in a grid-like pattern. Below the letters, the words "MICRO THERAPEUTICS, INC." are written in a smaller, serif font.

# 510(k) Summary of Safety and Effectiveness Flow Rider™ Flow Directed Micro Catheter

Prepared January 8, 1998

| Trade Name:   | Flow RiderTM Flow Directed Micro Catheter                                  |                 |                                                       |
|---------------|----------------------------------------------------------------------------|-----------------|-------------------------------------------------------|
| Generic Name: | Percutaneous Intravascular Catheter                                        | Classification: | Class II                                              |
| Submitted By: | Micro Therapeutics, Inc.<br>1062-F Calle Negocio<br>San Clemente, CA 92673 | Contact:        | Tom Daughters<br>Regulatory Affairs<br>(714) 361-0610 |

#### Predicate Devices

Target Spinnaker™ Flow Directed Infusion Catheter, K965189 Balt Magic™ Infusion Catheter, K923368

#### Device Description

The Flow Rider Flow Directed Micro Catheter is a single lumen infusion catheter designed to be flow directed into the distal vasculature. The Flow Directed Micro Catheter has a hydrophilic coating to provide lubricity for navigation of vessels. A standard luer lock adapter on the proximal hub is used for the attachment of accessories. The catheter tip is indicated by a radiopaque marker to facilitate fluoroscopic visualization.

#### Intended Use

The catheter is intended to be used for the controlled selective infusion of physician specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuro anatomy.

## Testing

Biocompatibility testing was performed on the Flow Directed Micro Catheter in accordance with the International Standard for the Biological Evaluation of Medical Devices. Part 1: Guidance on Selection of Tests (ISO 10993-1:1992(E)). Results of the tests showed that the device passed Biocompatibility testing and is suitable for its application.

Physical testing of the product included dimensional inspection, tensile strength tests, burst pressure tests, flow rate tests, torque tests and performance under simulated conditions. All testing of the product yielded acceptable results.

Animal studies were performed to assess the performance of the Flow Rider Catheter. The studies demonstrated that the Flow Rider Catheter is substantially equivalent to the predicate devices.

## Summary of Substantial Equivalence

The Flow Rider Flow Directed Micro Catheter is substantially equivalent to the predicate devices in intended use and principles of operation.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# APR 1 0 1998

Mr. Tom Daughters Requlatory Affairs Micro Therapeutics, Inc. 1062-F Calle Negocio San Clemente, CA 92673

Re: K980104 Flow Rider™ Flow Directed Micro Catheter Trade Name: Requlatory Class: II Product Code: KRA Dated: January 9, 1998 January 13, 1998 Received:

Dear Mr. Daughters:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device _ can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any

{2}------------------------------------------------

Page 2 - Mr. Tom Daughters

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

Device Name:

Flow Rider™ Flow Directed Micro Catheter

Indications for Use:

The catheter is intended to be used for the controlled selective infusion of physician specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuro anatomy.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over the Counter Use _________________________________________________________________________________________________________________________________________________________ OR

(Per 21 CFR 801.109)

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices

| 510(k) Number | K980604 |
|---------------|---------|
|---------------|---------|

Exhibit 6, Page 4

---

**Source:** [https://fda-staging.innolitics.com/device/K980104](https://fda-staging.innolitics.com/device/K980104)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com