STORZ DP4210 VENTURI ECONOMY ANTERIOR PACK AND DP5000 ASC DAYPACK
Applicant
Storz Instrument Co.
Product Code
MSR · Ophthalmic
Decision Date
Jun 1, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Storz DP4210 Venturi Economy Anterior Pack and DP5000 ASC Daypack are intended to be used as accessory devices with the Storz DAISY® (K854508), PREMIERE® (K894278, K921460 and K946227), Protege® (K921758 and K950114), and Millennium™ (K961310) Ophthalmic Microsurgical Systems. Both accessory packs facilitate use of Storz phaco handpieces (K935926) and Storz I/A handpieces (K951463) with the above referenced microsurgical systems in various anterior segment ophthalmic surgical procedures, e.g., phacoemulsification and residual cortical material removal following a planned Extracapsular Cataract Extraction.
Device Story
Accessory packs for ophthalmic microsurgical systems; facilitate irrigation/aspiration and phacoemulsification. DP4210 contains reusable anterior collection cassette, I/A tubing, and test chambers. DP5000 contains reusable cassette, test chambers, needle wrench, infusion sleeve; includes single-use I/A tube and BSS administration tube assemblies. Used in clinical settings by surgeons/staff during anterior segment procedures. Components connect to Storz phaco and I/A handpieces to manage fluidics and tissue removal. DP4210 provided non-sterile; DP5000 provided sterile. Reusable components limited to 10-20 uses. Supports surgical workflow by providing necessary fluidic pathways and instrumentation for cataract extraction.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and comparison of device characteristics, materials, and intended use to legally marketed predicate devices.
Technological Characteristics
Materials include silicone (reflux tubes). Components include collection cassettes, I/A tubing, test chambers, needle wrenches, and infusion sleeves. DP4210 is non-sterile; DP5000 is sterile. Reusable components limited to 10-20 uses. System functions as a fluidic accessory for phacoemulsification microsurgical systems.
Indications for Use
Indicated for patients undergoing anterior segment ophthalmic surgical procedures, including phacoemulsification and residual cortical material removal following planned Extracapsular Cataract Extraction. For prescription use only.
Regulatory Classification
Identification
A vitreous aspiration and cutting instrument is an electrically powered device, which may use ultrasound, intended to remove vitreous matter from the vitreous cavity or remove a crystalline lens.
Special Controls
*Classification.* Class II (special controls). The device, when it is phacofragmentation unit replacement tubing, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Storz DP4210 Universal I/A and Phaco Pack (K882240)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of characters that appear to be handwritten. The characters include 'K', 'G', '8', '0', '1', '0', and '0'. The characters are written in a dark color, possibly black ink, against a white background. The handwriting style is somewhat stylized, with distinct shapes for each character.
1 1998
Image /page/0/Picture/2 description: The image shows the word "storz" in white letters on a black background. The letters are stylized and have a unique font. There is a registered trademark symbol to the right of the word.
ﻣﺨﺘﻠﻔﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ
## Storz DP4210 Venturi Economy Anterior Pack and DP5000 ASC Daypack Premarket Notification
## -510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
The following information is submitted in accordance with the requirements of 21 CFR 807.92:
Contact Person: Patrick G. Balsmann, Regulatory Affairs Associate. (314) 225-5051 STORZ, 3365 Tree Court Industrial Blvd., St. Louis, MO 63122-6694
Date Prepared: January 9, 1998.
Proprietary Name: Storz DP4210 Venturi Economy Anterior Pack and DP5000 ASC Daypack.
Common/Usual Name: Ophthalmic microsurgical system irrigation/aspiration and phacoemulsification accessory packs.
Classification Name: Unit, Phacofragmentation (86HQC).
Device Description/Intended Use: The Storz DP4210 Venturi Economy Anterior Pack and DP5000 ASC Daypack are intended to be used as accessory devices with the Storz DAISY® (K854508), PREMIERE® (K894278, K921460 and K946227), Protege® (K921758 and K950114), and Millennium™ (K961310) Ophthalmic Microsurgical Systems. Both accessory packs facilitate use of Storz phaco handpieces (K935926) and Storz I/A handpieces (K951463) with the above referenced microsurgical systems in various anterior segment ophthalmic surgical procedures, e.g., phacoemulsification and residual cortical material removal following a planned Extracapsular Cataract Extraction.
The Storz DP4210 Venturi Economy Anterior Pack is provided non-sterile and contains reusable components consisting of an anterior collection cassette, I/A tubing set, and I/A test chambers. The Storz DP5000 ASC Daypack is provided sterile and contains both reusable and single use components. The DP5000 reusable items include an anterior collection cassette, I/A test chambers, needle wrench, and infusion sleeve. The DP5000 single use items include an I/A tube assembly and a BSS administration tube assembly.
{1}------------------------------------------------
Predicate Device: The Storz DP4210 Venturi Economy Anterior Pack is substantially equivalent to the Storz DP4210 Universal I/A and Phaco Pack (K882240.) The Storz DP5000 ASC Daypack is substantially equivalent to the Storz DP4210 Universal I/A and Phaco Pack (K882240) and the Storz DP4310 Deluxe Phaco Pack (K955901.)
Predicate Comparison: A chart comparing the Storz DP4210 Venturi Economy Anterior Pack and DP5000 ASC Daypack to their respective predicate devices, demonstrating substantial equivalence, is attached.
Submitted by;
W. H. Johnson
Patrick G. Balsmann Regulatory Affairs Associate Storz Instrument Company
{2}------------------------------------------------
## Storz DP4210 Venturi Economy Anterior Pack Device Comparison Chart
| Device Characteristic | Storz DP4210 (modified) | Storz DP4210 (existing) |
|----------------------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k): | Present. | Storz K882240. Referenced as accessory device in Storz DAISY® (K854508), PREMIERE® (K894278, K921460 and K946227), Protegé® (K921758 and K950114), and Millennium™ (K961310) Ophthalmic Microsurgical Systems. |
| Pack Name: | DP4210 Venturi Economy<br>Anterior Pack | DP4210 Universal I/A and<br>Phaco Pack |
| Intended Use: | Anterior segment irrigation/<br>aspiration and<br>phacoemulsification<br>ophthalmic surgical<br>procedures. | Anterior segment irrigation/<br>aspiration and<br>phacoemulsification<br>ophthalmic surgical<br>procedures. |
| Individual Pack<br>Components: | (1) Anterior<br>Collection Cassette<br>(1) I/A Tubing Set<br>(2) I/A Test Chamber | (1) Anterior<br>Collection Cassette<br>(1) I/A Tubing Set<br>(2) I/A Test Chamber |
| Anterior Collection<br>Cassette: | Cylindrical silicone reflux<br>tube. | Rectangular silicone reflux<br>tube. |
| Reusable Components: | Yes, all components. | Yes, all components. |
| Limited Reuse: | Not to exceed 20 times. | Not to exceed 20 times. |
| Pack Provided Sterile: | No. | No. |
·
ుర్త
{3}------------------------------------------------
## Storz DP5000 ASC Daypack Device Comparison Chart
1
| Device<br>Characteristic | Storz DP5000 | Storz DP4210 | Storz DP4310 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k): | Referenced as accesory<br>device in Storz DAISY®<br>(K854508), PREMIERE®<br>(K894278, K921460 and<br>K946227), Protegé®<br>(K921758 and K950114),<br>and Millennium™<br>(K961310) Ophthalmic<br>Microsurgical Systems.<br>Present. | Storz K882240. | Storz K955901. |
| Pack Name: | DP5000 ASC Daypack. | DP4210 Universal I/A and<br>Phaco Pack | DP4310 Deluxe Phaco<br>Pack. |
| Intended Use: | Anterior segment<br>irrigation/<br>aspiration and<br>phacoemulsification<br>ophthalmic surgical<br>procedures. | Anterior segment<br>irrigation/<br>aspiration and<br>phacoemulsification<br>ophthalmic surgical<br>procedures. | Anterior segment<br>irrigation/<br>aspiration and<br>phacoemulsification<br>ophthalmic surgical<br>procedures. |
| Individual<br>Pack<br>Components: | (1) Anterior<br>Collection Cassette<br>(2) I/A Test Chamber<br>(1) Needle Wrench<br>(2) Infusion Sleeve<br>(1) I/A Tube Assembly<br>(1) BSS Administration<br>Tube Assembly | (1) Anterior<br>Collection Cassette<br>(1) I/A Tubing Set<br>(2) I/A Test Chamber | (1) Anterior<br>Collection Cassette<br>(2) I/A Test Chamber<br>(1) Needle Wrench<br>(2) Infusion Sleeve<br>(1) I/A Tube Assembly<br>(1) BSS Administration<br>Tube Assembly<br>(1) Auxiliary Drape<br>(1) Mayo Arm Drape<br>(1) Male Luer Cap<br>(1) Tube Stopper |
| Reusable<br>Components: | Yes, anterior collection<br>cassette, I/A test chamber,<br>needle wrench, and<br>infusion sleeve. | Yes, all components. | No. |
| Limited Reuse<br>Components: | Not to exceed 10 times. | Not to exceed 20 times. | Not applicable. |
| Pack Provided<br>Sterile: | Yes. | No. | Yes. |
| Pack<br>Sterilization: | Not applicable. | Gamma irradiation. | Gamma irradiation. |
:
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 1998
Mr. Gary M. Rauvola Group Manager, Bausch & Lomb Surgical, Storz Products 3365 Tree Court Industrial Blvd. St. Louis, MO 63122-6694
Re: K980100 Trade Name: Storz DP4210 Venturi Economy Anterior Pack and DP500 ASC Daypack Regulatory Class: II Product Code: 86 MSR Dated: April 24, 1998 Received: April 27, 1998
Dear Mr.Rauvola:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{5}------------------------------------------------
Page 2 - Mr. Gary M. Rauvola
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page / of /
510(k) Number (if known): K 980100 Recess
Device Name: DR4211 Venture Frenchise Anterior Pack DP SCCC Indications For Use:
> The Storz DP4210 Venturi Economy Anterior Pack and DP5000 ASC Daypack The Storz Dr 4210 - Penian Loodlection cassettes, needle sleeves, and tip testers for Contain replacedore tabiligs, sents and Millennium™ Ophthalmic Microsurgical the Dioriz Dato, O, I reed in anterior segment ophthalmic surgical procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Quynh Hoang
sion Sion-Off) Division of Ophthalmic Devices K980100 510(k) Number _
Prescription Use_ X (Per 21 CFR 801.109)
્ત્વ
Over-The-Counter Use_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.