← Product Code [MEG](/productcode/MEG) · K980069

# VANISHPOINT SYRINGE (K980069)

_Retractable Technologies, Inc. · MEG · Feb 12, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K980069

## Device Facts

- **Applicant:** Retractable Technologies, Inc.
- **Product Code:** [MEG](/productcode/MEG.md)
- **Decision Date:** Feb 12, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The function of the VanishPoint® Syringe is to provide a safe and reliable method of injecting medication into a patient or withdrawing blood from a patient. The VanishPoint® Syringe works like a conventional hypodermic syringe except for its ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user. Because the contaminated needle is automatically withdrawn into the syringe barrel, the syringe user is protected from accidental needle sticks. These accidental needlesticks would occur between removing the needle from the patient and disposing of the syringe in a sharps disposable container.

## Device Story

VanishPoint® Syringe functions as a standard hypodermic syringe for medication delivery or blood collection; features integrated manual needle retraction mechanism. User activates retraction immediately post-injection; needle automatically withdraws into syringe barrel. Designed for clinical use to mitigate risk of accidental needlestick injuries during disposal. Device provides physical containment of contaminated needle; reduces exposure risk for healthcare personnel.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Manual retractable needle syringe; piston-style; plastic barrel and plunger components. Operates via user-activated retraction mechanism. No electronic components or software.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- 3cc VanishPoint Syringe (marketed Dec 1995)
- 5cc VanishPoint Syringe (marketed May 1997)
- 10cc VanishPoint Syringe (marketed May 1997)
- 1cc Allergy and Insulin Syringes (marketed Nov 1997)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K980069

FE

January 15, 1998

### 510(k) SAFETY and EFFECTIVENESS SUMMARY page 1 of 2 Re:

Pursuant to Section 510(k) of the Federal Food, Drug and Cosmetic Act, Safety and Effectiveness information is enclosed for the following device:

| Device Name:              | Automated Retractable VanishPoint Syringe                                                                                                                                                                                                                                                                                                                                                                      |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:               | VanishPoint® Syringe                                                                                                                                                                                                                                                                                                                                                                                           |
| Common Name:              | Safety Syringe                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name:      | Piston Syringe/Anti-Stick Syringe                                                                                                                                                                                                                                                                                                                                                                              |
| Device Class:             | Class II Medical Device                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Panel:     | Panel 80 (General Hospital and Personal Use)                                                                                                                                                                                                                                                                                                                                                                   |
|                           | Procode: MEG - Anti-Stick Syringe                                                                                                                                                                                                                                                                                                                                                                              |
| Performance Standard:     | None established for hypodermic syringes.                                                                                                                                                                                                                                                                                                                                                                      |
| Safety and Effectiveness: | No new issues of safety or effectiveness when prepared.<br>The 1cc Tuberculin VanishPoint® Syringe is a smaller version and<br>is substantially equivalent to the original 3cc VanishPoint® which<br>was introduced into the market in December 1995, the 5cc and<br>10ccVanishPoint® Syringes which were introduced May 1997 and<br>the 1cc Allergy and Insulin Syringes were introduced in November<br>1997. |
| Facility Address:         | Retractable Technologies, Inc.<br>511 Lobo Lane, P.O. Box 9<br>Little Elm, TX 75068-0009                                                                                                                                                                                                                                                                                                                       |

### Establishment Registration Number: 1647137

Indications for Use: The function of the VanishPoint® Syringe is to provide a safe and reliable method of injecting medication into a patient or withdrawing blood from a patient. The VanishPoint® Syringe works like a conventional hypodermic syringe except for its ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user. Because the contaminated needle is automatically withdrawn into the syringe barrel, the syringe user is protected from accidental needle sticks. These accidental needlesticks would occur between removing the needle from the patient and disposing of the syringe in a sharps disposable container.

{1}------------------------------------------------

# 510(k) SAFETY and EFFECTIVENESS SUMMARY __page 2 of 2

# 4年 10月 10月 18:58:58 AM 10:00 10:00 1

Substantial Equivalence The 1 cc Tuberculin VanishPoint® is a smaller version of the original 3 cc VanishPoint® originally called a Pop n' Lok Syringe.

| Contact Information: |                                                                                          |
|----------------------|------------------------------------------------------------------------------------------|
| Name:                | Terance Grisso                                                                           |
|                      | Director, Regulatory Affairs and Quality                                                 |
| Address:             | Retractable Technologies, Inc.<br>511 Lobo Lane, P.O. Box 9<br>Little Elm, TX 75068-0009 |
| Telephone:           | (972) 294-1010                                                                           |
| Facsimile:           | (972) 294-4400                                                                           |

... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2) 294-4400

The 1 cc Tuberculin Vanis
nt® originally called a Pop n

**MARY page 2 of 2**

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB | 2 |998

Mr. Terance Grisso Director of Regulatory Affairs and Quality Retractable Technologies, Incorporated 511 Lobo Lane 75068 Little Elm, Texas

Re : K980069 Trade Name: Vanishpoint Syringe Regulatory Class: II Product Code: MEG Dated: January 7, 1998 Received: January 8, 1998

Dear Mr. Grisso:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with ----the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Grisso

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to one began and fication" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K980069

## INDICATIONS FOR USE

510 (k) Number (if known):

Device Name:

## Indications for use:

The function of the VanishPoint® Syringe is to provide a safe and reliable method of injecting medication into a patient or withdrawing blood from a patient. The VanishPoint® Syringe works like a conventional hypodermic syringe except for its ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user. Because the contaminated needle is automatically withdrawn into the syringe barrel, the syringe user is protected from accidental needlesticks. These accidental needlesticks would occur between removing the needle from the patient and disposing of the syringe in a sharps disposable container.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Pataria Crescenti

510(k) Numb

Prescription Use

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K980069](https://fda-staging.innolitics.com/device/K980069)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
