← Product Code [GCJ](/productcode/GCJ) · K980023

# ENDOPATH AND PROXIMATE LINEAR CUTTERS (K980023)

_Ethicon Endo-Surgery, Inc. · GCJ · Mar 18, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K980023

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Mar 18, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The ENDOPATH® and PROXIMATE® Linear Cutters are intended for transection, resection, and/or creation of anastomoses.

## Device Story

Linear cutters are surgical instruments used for tissue transection, resection, and anastomosis creation. Devices function by mechanically stapling and cutting tissue; some models include bipolar coagulation features. Used in open and minimally invasive surgical procedures by surgeons. Output is the physical transection and stapling of internal tissues, facilitating surgical reconstruction. Benefits include standardized tissue management and reduced operative time. Operation is manual; some models require specific electrosurgical generators for coagulation functions.

## Clinical Evidence

No clinical data provided; performance characteristics are stated to be unchanged from previously cleared predicate devices.

## Technological Characteristics

Mechanical surgical stapling and cutting instruments. Materials, design, and physical properties are identical to previously cleared predicate devices. Some models include bipolar electrosurgical coagulation capabilities requiring external generators.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Linear Stapler (LSR) - Rigid ([K821994](/device/K821994.md))
- Disposable Linear Cutter ([K843034](/device/K843034.md))
- PROXIMATE® Linear Cutter Thick Tissue Instrument ([K892927](/device/K892927.md))
- ENDOPATH® Endoscopic Linear Cutter with Safety Lock-Out and Reloading Unit ([K915099](/device/K915099.md))
- ENDOPATH® Reloadable Linear Cutter with Safety Lock-Out and Reloading Unit ([K930934](/device/K930934.md))
- Modified ENDOPATH® Endoscopic Linear Cutter with Safety Lock-Out and Reloading Unit (No-Row) ([K935064](/device/K935064.md))
- Modified ENDOPATH® Endoscopic Reloadable Linear Cutter with Safety Lockout and Reload Unit ([K945202](/device/K945202.md))
- New Indication for ENDOPATH® and PROXIMATE® Linear Cutters and Staplers ([K951546](/device/K951546.md))
- ENDOPATH® EZ-RF Linear Cutter and Coagulation Device ([K961264](/device/K961264.md))
- ENDOPATH® ETS Endoscopic Linear Cutter and ETS-FLEX Endoscopic Articulating Linear Cutter Devices ([K961390](/device/K961390.md))
- ENDOPATH® EZ45 Endoscopic Linear Cutter and ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler ([K970720](/device/K970720.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a black and white graphic of a circle with lines emanating from the upper right quadrant. The lines are thick and appear to be converging towards a central point outside the circle. The lower left quadrant of the circle is shaded with a stippled effect, creating a contrast with the rest of the circle.

ETHICON ENDO-SURGERY, INC. a Johnson Johnson company

# 980023

MAR 1 8 1998

4545 CREEK ROAD
CINCINNATI, OH 45242-2839

### 510(k) Summary of Safety and Effectiveness

| Contact                        | Lorri Chavez, Project Manager<br>Ethicon Endo-Surgery, Inc.<br>4545 Creek Road<br>Cincinnati, Ohio 45242<br>Telephone: (513) 483-3513                                                                                                                                                                                                                                                                                                                                                                                                                           |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date prepared                  | January 2, 1998                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device                         | Name: ENDOPATH® and PROXIMATE® Linear Cutters<br>Classification Name: Endoscope and Accessories<br>Common Name: Linear Cutter<br>Trade Name/Proprietary Name: ENDOPATH® and PROXIMATE® Linear<br>Cutters                                                                                                                                                                                                                                                                                                                                                        |
| Legally<br>marketed<br>devices | Linear Stapler (LSR) - Rigid (K821994);<br>Disposable Linear Cutter (K843034);<br>PROXIMATE® Linear Cutter Thick Tissue Instrument (K892927);<br>ENDOPATH® Endoscopic Linear Cutter with Safety Lock-Out and<br>Reloading Unit (K915099);<br>ENDOPATH® Reloadable Linear Cutter with Safety Lock-Out and<br>Reloading Unit (K930934);<br>Modified ENDOPATH® Endoscopic Linear Cutter with Safety Lock-Out and<br>Reloading Unit (No-Row) (K935064);<br>Modified ENDOPATH® Endoscopic Reloadable Linear Cutter with Safety<br>Lockout and Reload Unit (K945202); |

Continued on next page

{1}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness, Continued

| Legally<br>marketed<br>devices, cont'd | New Indication for ENDOPATH® and PROXIMATE® Linear Cutters and<br>Staplers (K951546);<br>ENDOPATH® EZ-RF Linear Cutter and Coagulation Device (K961264);<br>ENDOPATH® ETS Endoscopic Linear Cutter and ETS-FLEX Endoscopic<br>Articulating Linear Cutter Devices (K961390); and,<br>ENDOPATH® EZ45 Endoscopic Linear Cutter and ENDOPATH® EZ45 No<br>Knife Endoscopic Linear Stapler (K970720). |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>description                  | All device functions, scientific concepts, significant physical and performance<br>characteristics (i.e. device design, materials, physical properties, etc.) have not<br>changed from the above-mentioned 510(k)s.                                                                                                                                                                             |
| Intended use                           | The ENDOPATH® and PROXIMATE® Linear Cutters are intended for<br>transection, resection, and/or creation of anastomoses.                                                                                                                                                                                                                                                                         |
| Indications<br>statement               | The ENDOPATH® and PROXIMATE® Linear Cutter indication statements<br>have not changed from the above-mentioned 510(k)s.                                                                                                                                                                                                                                                                          |
| Technological<br>characteristics       | The ENDOPATH® and PROXIMATE® Linear Cutter technological<br>characteristics have not changed from the above-mentioned 510(k)s.                                                                                                                                                                                                                                                                  |
| Performance<br>data                    | The ENDOPATH® and PROXIMATE® Linear Cutters' performances have<br>not changed from the above-mentioned 510(k)s.                                                                                                                                                                                                                                                                                 |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a bird with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 8 1998

Ms. Lorri Chavez Project Manager Requlatory Affairs Ethicon Endo-Surgery, Inc. 4545 Creek Road 45242-2839 Cincinnati, Ohio

K980023 Re: Endopath® and Proximate® Linear Cutters Trade Name: Requlatory Class: II Product Code: GCJ Dated: January 2, 1998 Received: January 5, 1998

Dear Ms. Chavez:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸੇ substantially equivalent determination assumes compliance with ....... the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Chavez

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Stephen Rhodes

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use Statement (App. B)

Indications for Use Statement: Statement

510(k) Number: K 980023

Device Name: ENDOPATH® and PROXIMATE® Linear Cutters

K821994: The Disposable Linear Stapler has application in abdominal, thoracic, and pediatric surgery for resection and transection of internal tissues. The total thickness of the tissues being connected must be within the range of 1.0 to 3.0 mm.

K843034: The PLC50 has application in abdominal, pediatric and thoracic surgery for resection, transection and creation of anastomoses.

K892927: The LCT75 has application in gastrointestinal and thoracic surgery for resection, transection and creation of anastomoses. Cartridges for the standard (PLC75) and Thick Tissue (LCT75) are interchangeable within these instruments.

K915099: The ENDOPATH Endoscopic Linear Cutter with Safety Lock-out has applications multiple open or endoscopic procedures for the transection, resection and creation of anastomosis.

K930934: The ENDOPATH Endoscopic Reloadable Linear Cutter with Safety Lock-Out has application in gastrointestinal and thoracic surgery for transection, resection and creation of anastomoses.

K935064: The ENDOPATH Endoscopic Linear Cutter with Safety Lock-Out has application in multiple open or endoscopic procedures for the transection, resection, and creation of anastomoses.

K945202: The ENDOPATH Endoscopic Reloadable Linear Cutter with ----------Safety Lock-Out has multiple open or endoscopic procedures for transection. resection and/or creation of anastomoses.

Continued on next page

{5}------------------------------------------------

### Indications for Use Statement (App. B), Continued

K951546: [Verbiage for all of the various indication statements remain the Statement. continued same, plus] ... and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

> K961264: The ENDOPATH® EZ-RF Linear Cutter and Coagulation device has applications in open and minimally invasive surgical procedures for coagulation, transection, and resection of tissue. The bipolar feature is to be used only with the ERBE ERBOTOM ICC 350 or the Valleylab Force 2 electrosurgical generators.

> K961390: The ENDOPATH ETS Endoscopic Linear Cutter and ETS-FLEX Endoscopic Articulating Linear Cutter have application in general, urologic, gynecologic, and thoracic surgery for transection, resection, and/or creation of anastomoses.

K970720: The ENDOPATH EZ45 Endoscopic Linear Cutter has application in multiple open or other minimally invasive surgical procedures for transection, resection, and/or creation of anastomoses and can be used with staple line or tissue buttressing materials, such as bovine pericardium. and

The ENDOPATH EZ45 No Knife Endoscopic Linear Stapler has application in multiple open or other minimally invasive surgical procedures for transection and resection, and can be used with staple line or buttressing materials, such as bovine pericardium.

Stipta Plwelu

ivision Sign-Off) ision of General Restorative Devices 980023 10(k) Number

Prescription Use
X
(Per 21 CFR 801.109)

Over-the-Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K980023](https://fda-staging.innolitics.com/device/K980023)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).