K974878 · Martin Uram, MD · GEX · Mar 27, 1998 · General, Plastic Surgery
Device Facts
Record ID
K974878
Device Name
E3 MICROPROBE
Applicant
Martin Uram, MD
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The E3 MicroProbe delivers up to 60 Watts of continuous wave radiation to a flexible optical fiber for use in open and endoscopic procedures.
Device Story
E3 MicroProbe is an integrated surgical laser and endoscope system. It utilizes a solid-state diode laser (810 nm) for therapeutic energy delivery and a visible diode laser (640 nm) for aiming. The endoscope features fiberoptics for viewing, laser energy delivery, and illumination. The system includes a console and footpedal; the operator presets laser power and pulse duration at the console and initiates delivery via the footpedal. It is used by clinicians in open and endoscopic procedures to perform soft tissue ablation, incision, excision, coagulation, and vaporization. The device provides visual access to the surgical site while delivering laser energy to the target tissue, facilitating precise surgical intervention and potentially reducing trauma compared to traditional methods.
Clinical Evidence
Bench testing only. The device underwent unit and system-level validation to confirm that performance specifications are satisfied and that the modified device is substantially equivalent to the predicate.
Technological Characteristics
Solid-state diode laser system; 810 nm therapeutic wavelength; 640 nm aiming beam; continuous wave output up to 60 Watts; fiberoptic delivery (0.25 mm viewing, 0.10 mm laser delivery); console-based user interface; footpedal activation; cooling system scaled for power output.
Indications for Use
Indicated for contact and non-contact excision, hemostasis, incision, and vaporization of soft tissue in general surgery, ophthalmology/oculoplastic, urology, gastroenterology, gynecology, otorhinolaryngology, pulmonary/thoracic, dermatology/plastic surgery, neurosurgery, and orthopedic applications.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
MARTIN URAM, M.D. Disorders of the Retina
39 Sycamore Avenue Little Silver, New Jersey 07739 (908) 530-7730 Fax (908) 530-5344
225 East 64th Street New York, New York 10021 (212) 628-0032
K97487P
MAR 2 7 1998
#### Summary 510 (k) For the E3 MicroProbe
1. - Date Summary Prepared:
December 15, 1997
2. - Submitter's Name and Address:
Martin Uram, MD
39 Sycamore Avenue
Little Silver, NJ 07739-1208
Contact Person: Keith Hertz
Fax: 732-530-5344 Tel.: 732-530-6762
E-mail: mail@monmouth.com
### 3. - Device Name:
Trade / Proprietary Name: E3 MicroProbe / 2
e
David Zoser
Common Name: Integrated Surgical Laser and Endoscope System
Classification Name: Surgical Diode Laser Fiber Optic Endoscope
{1}------------------------------------------------
39 Sycamore Avenue Little Silver, New Jersey 07739 (908) 530-7730 Fax (908) 530-5344
225 East 64th Street New York, New York 10021 (212) 628-0032
#### 10 (K) Probe For
## 4. - Predicate Devices:
The legally marketed devices to which equivalence is being claimed are:
Pentax Video Colonoscope MicroProbe Pentex Video Duodenoscope Family of Endoscopes Pentax Video Gastroscope Value Pack Pentax Video Sigmoidoscope Endoscope w. Cannula Pentax Bronchofiberscope Transtympanic Endoscope Storz Flexible Fiberscope Diomed (15, 30, 60) Surgical Diode Laser Storz Flexible Rhino-Pharyngo-Coherent Surgical Laser Laryngo-Fiberscope Storz Flexible Hysteroscope ASI Uroplasty Cystoscope Storz Hopkins Telescope Olympus Arthroscopy System Stryker Arthroscopy System Olympus Laparo-Thoraco Videoscope Stryker Sinuscope Olympus Oratracheal Intubation Endoscope Olympus Endoscope System
{2}------------------------------------------------
39 Sycamore Avenue Little Silver, New Jersey 07739 (908) 530-7730 Fax (908) 530-5344
225 East 64th Street New York, New York 10021 (212) 628-0032
# 510 (k) Summarv For the E3 MicroProbe
# 5. - Device Description:
The Endo Optiks E3 MicroProbe System consists of a laser, endoscope, console and footpedal. The treatment laser is a solid state diode laser system operating at 810 nanometers ± 2 nanometers and continuous wave power of up to 60 Watts. A visible solid state diode laser operating at a nominal wavelength of 640 nm and a variable continuous wave output from 0 to 1.5 milliwatts is incorporated as an aiming device. The end of the endoscope consists of a 0.25 milimeter diameter fiberoptic for viewing, a 0.10 millimeter diameter fiberoptic for delivery of the laser energy and fiberoptics that transmit the illumination light. Laser energy is delivered via a single timed continuous wave pulse with the time preset at the operator console and the pulse initiated by the footpedal. Laser power is also preset at the console.
The E3 MicroProbe is a modification to the maximum power output rating of the original MicroProbe.
Labeling: The MicroProbe will be renamed (relabeled) the E3 MicroProbe. This change is being made to be consistant with the industry practice of designating a laser's power based on the output power at the laser port rather than that at the tissue site. There is no change in construction or performance of this device.
Power Output: An additional model will be created by increasing the output power to produce the E3 MicroProbe. The same basic design is retained. The 60 Watt model uses sixteen 4 Watt laser diodes. The power supply and cooling designs of the new model have been appropriately scaled, while the mechanical design and user interface remain unchanged.
{3}------------------------------------------------
#### MARTIN URAM, M.D. Disorders of the Retina
39 Sycamore Avenue Little Silver, New Jersey 07739 (988) 530-7730 Fax (998) 530-5344
225 East 64th Street New York, New York 10021 (212) 628-0032
737
#### 510 (k) Summary For the E3 MicroProbe
Indication For Use: The indications are expanded to include both contact and non-contact ablation, incision, excision, coagulation and vaporization for the following soft tissue applications:
- General Surgery ●
- . Ophthalalmology / Oculoplastic
- . Urology
- . Gastroenterology
- . Gynecology
- Otorhinolaryngology .
- . Pulmonary / Thoracic
- Dermatology / Plastic Surgery �
- Neurosurgery ●
- . Orthopedic
# 6. - Intended Use:
The E3 MicroProbe delivers up to 60 Watts of continuous wave radiation to a flexible optical fiber for use in open and endoscopic procedures.
# 7 . - Comparison of Technological Charateristics
This modification replaces the original MicroProbe with four models, the Full Featured MicroProbe, the E3 MicroProbe - 15, the E3 MicroProbe - 30 and the E3 MicroProbe - 60. Modifications are to offer models with different maximum output powers (15, 30 and 60 Watts at the output port), in order to provide sufficient power when it is required, and to provide a more cost effective device for applications that do not require the higher power capabilities. The original MicroProbe currently operates with a maximum power of 15 Watts. In addition, indications for use have been expanded to include both contact and non-contact treatment of soft tissue applications for which leqally marketed lasers are currently available.
{4}------------------------------------------------
39 Svcamore Avenue Little Silver, New Jersey 07739 (908) 530-7730 Fax (908) 530-5344
225 East 64th Street New York, New York 10021 (212) 628-0032
#### Summarv 510 (k) E3 MicroProbe For
## 8. - Nonclinical Tests Used in Determination of Substancial Equivalence
The design of the E3 MicroProbe has been thoroughly validated at the unit and system level. The tests showed that all system specifications are satisfied.
# 9. - Conclusions From Nonclinical Testing
The testing of the modified devices demonstrates that the performance is substantially equivalent to the predicate prior to the modifications.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized human figure with three arms or lines extending from the head. The symbol is enclosed in a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. The text is in all caps and is arranged to follow the curve of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 27 1998
Martin Uram, M.D. Disorders of the Retina 39 Sycamore Avenue 07739 Little Silver, New Jersey
Re: K974878 E3 MicroProbe Series of Diode Lasers Trade Name: and Accessories Regulatory Class: II Product Code: GEX Dated: December 22, 1997 December 30, 1997 Received:
Dear Dr. Uram:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 - Dr. Uram
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
f
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
K974878/12
DUPLICATE
225 East 64th Street
New York, New York 10021
(212) 628-0032
39 Sycamore Avenue a Silver, New Jersey 07739 (908) 530-7730 Fax (908) 530-5344
### 510(k) Number: K974878
### Device Name: E3 MicroProbe
### Indications For Use
Contact and non-contact excision, hemostasis, incision and vaporization for the following soft tissue applications:
General Surgery Ophthalmology / Oculoplastic Urology Gastroenterology Gynecology
Otorhinolaryngology Pulmonary / Thoracic Dermatology / Plastic Surgery Neurosurgery Orthopedic
Concurrent of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Prescription Use (Per21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-96)
(Division Sign-Off)
Division of General Restorative Devices צרף רק צוצרף 510(k) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.