ABUSCREEN ONLINE II FOR OPIATES 300/2000, ABUSCREEN ONLINE OPIATES CALIBRATION PACK, ABUSCREEN ONLINE OPIATES CONTROL
Applicant
Roche Diagnostic Systems, Inc.
Product Code
DJG · Clinical Toxicology
Decision Date
Apr 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Indications for Use
The Abuscreen OnLine II for Opiates 300/2000 Reagent is an in vitro diagnostic test for the qualitative (at 300 and 2000 ng/mL cutoff) and semi-quantitative (at 2000 ng/mL cutoff) detection of morphine and its metabolites in human urine. The Abuscreen ONLINE Opiates Calibration Pack is an in vitro diagnostic device designed for the calibration of the Roche reagent assays for opiates. The Roche Abuscreen ONLINE Opiates Control Pack is an assayed quality control sample for use with the Roche assays for opiates.
Device Story
Abuscreen OnLine II for Opiates 300/2000 is an in vitro diagnostic reagent system for detecting morphine and metabolites in human urine. It utilizes kinetic interaction of microparticles in solution (KIMS) technology. The system includes an antibody/microparticle reagent (mouse monoclonal antibody) and a conjugate reagent (opiate derivative). The device is used in clinical laboratory settings by trained personnel. The assay provides qualitative results at 300 and 2000 ng/mL cutoffs and semi-quantitative results at 2000 ng/mL. Results are interpreted by healthcare providers to assist in clinical toxicology assessments. The system includes dedicated calibration and control packs to ensure assay accuracy and quality control.
Clinical Evidence
Bench testing only. Accuracy evaluated by comparing results to GC/MS (gold standard). For the 300 ng/mL qualitative application, 100% agreement (N=40) was observed. For the 2000 ng/mL quantitative application, 100% agreement (N=42) was observed. Precision studies showed within-run CVs ranging from 0.7% to 4.2% across various concentrations. Analytical sensitivity reported as <12 ng/mL.
Technological Characteristics
Methodology: Kinetic interaction of microparticles in solution (KIMS). Reagents: Mouse monoclonal anti-opiate antibody attached to microparticles and opiate-conjugated derivative. Sample: Human urine. Assay type: Qualitative and semi-quantitative. Connectivity: Not applicable (reagent system).
Indications for Use
Indicated for the qualitative and semi-quantitative detection of morphine and its metabolites in human urine samples. Used for clinical toxicology testing.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
APR 2 4 1998
Image /page/0/Picture/3 description: The image shows the Roche logo. The logo consists of the word "Roche" written in bold, sans-serif font. The word is enclosed in a hexagon shape with rounded corners. The logo is black and white.
K974840
# 510(k) Summary
# Abuscreen OnLine II for Opiates 300/2000 Reagent Abuscreen OnLine Opiates Calibration Pack Abuscreen OnLine Opiates Control Pack
In accordance with the Sufe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.
12974842 The assigned 510(k) number is :_
#### I. Identification of 510(k) Sponsor:
Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jetsey 08876-3771
510(k) Submission dated December 23, 1997
Contact: James W. Haynes Regulatory Affairs Associate Phone: (908) 253-7569 Fax: (908) 253-7547
Roche Diagnostic Systems, Inc.
Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
{1}------------------------------------------------
Roche Diagnostic Systems, Inc. Somerville, NJ אַפּל 1998 April
Abuscreen OnLine II Opiates 300/2000 Reagent, Calibrator & Controls Summary
#### II. Device Name:
The device name, including both the trade/proprietary name and the classification name are provided in the table below.
### Table I
| Proprietary Name | Classification Name | Product Code | Regulation Number |
|-----------------------------------------------------|-----------------------------------------|--------------|-------------------|
| Abuscreen OnLine II for Opiates<br>300/2000 Reagent | Opiate test<br>system | DJG | 862.3650 |
| Abuscreen OnLine Opiates<br>Calibration Pack | Clinical toxicology calibrator | DLJ | 862.3200 |
| Abuscreen OnLine Opiates Control<br>Pack | Clinical toxicology control<br>material | DIF | 862.3280 |
#### III. Identification of the legally marketed device to which the 510(k) sponsor claims equivalence:
The following table identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.
| Product Name | Predicate Product Name | K number | Date of<br>substantial<br>equivalence |
|-----------------------------------------------------|-------------------------------|----------|---------------------------------------|
| Abuscreen OnLine II for Opiates<br>300/2000 Reagent | Abuscreen OnLine for Opiates | K951319 | 8/2/95 |
| Abuscreen OnLine Opiates Calibration<br>Pack | Abuscreen ONLINE MiL Cal Pack | K935550 | 1/3/94 |
| Abuscreen OnLine Opiates Control<br>Pack | Abuscreen OnLine Controls | K962280 | 8/16/96 |
| 1<br>S | 17<br>11<br>1 | 1<br>œ<br>11<br>V<br>1 | 0 |
|--------|---------------|------------------------|---|
| | | | |
{2}------------------------------------------------
Roche Diagnostic Systems, Inc. Somerville, NJ April 1998
#### Description of the Device/Statement of Intended Use: IV.
### Abuscreen OnLine II for Opiates 300/2000 Reagent:
The Abuscreen OnLine II for Opiates 300/2000 Reagent is an in vitro diagnostic test for the qualitative (at 300 and 2000 ug/mL cutoff) and serni-quantitative (at 2000 ng/mL cutoff) detection of morphine and its metabolites in human urine.
### Abuscreen OnLine Opiates Calibration Pack:
The Abuscreen ONLINE Opiates Calibration Pack is an in vitro diagnostic device designed for the calibration of the Roche reagent assays for opiates.
## Abuscreen ONLINE Opiates Control Pack:
The Roche Abuscreen ONLINE Opiates Control Pack is an assayed quality control sample for use with the Roche assays for opiates.
The intended use, clinical utility and incthodology of cach device is further described in the package inserts, contained in the test specific sections of this submission.
#### Summary of the technological characteristics of the new device in comparison to V. those of the predicate.
Tables 3-5 outline the technological characteristics (methodologies) of cach device in comparison to those of legally marketed predicate products.
#### VI. Brief discussion of the clinical and nonclinical tests relied on for a determination of substantial equivalence:
Tables 3-5 demonstrate the results of clinical and nonclinical studies performed using the Abuscreen OnLine II for Opiales 300/2000 Reagent. Abuscreen OnLine Opiates Calibration Pack and the Abuscreen ONLINE Opiates Control Pack, respectively. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in these charts. This information concludes that the performance of these devices are essentially equivalent to other legally marketed devices of a similar kind.
{3}------------------------------------------------
Roche Diagnostic Systems, Inc. Somerville, NJ April 1998
# Table 3 - Abuscreen OnLine II for Opiates 300/2000 Reagent
| | Abuscreen OnLine II Opiates<br>Reagent | Abuscreen OnLine Opiates<br>Reagent |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Methodology | Kinetic interaction of<br>microparticles in a solution | Kinetic interaction of<br>microparticles in a solution |
| Sample type | urine | urine |
| Calibrator | For 300ng/mL cutoff:<br>Abuscreen OnLine Calibrator<br>Level 3<br>For 2000 ng/mL cutoff:<br>Abuscreen OnLinc Opiates<br>Calibration Pack | Abuscreen OnLine Calibration<br>Pack<br>Abuscreen ONLINE Mil<br>Calibration Pack |
| Cutoff | Qualitative:<br>300 and 2000 ng/mL<br>Semi-quantitative:<br>2000 ng/mL | Semi-quantitative:<br>2000 ng/mL |
| Reagent<br>(active ingredients) | 1. Ab/Microparticle reagent:<br>Microparticles attached with<br>opiate monoclonal antibody<br>(mouse)in buffer<br>2. Conjugate reagent:<br>Opiate conjugated derivative<br>in buffer | 1. Ab reagent: Opiate<br>monoclonal antibody in buffer<br>2. Microparticle reagent:<br>Conjugated opiate derivative<br>microparticles in buffer<br>3. Diluent |
| Performance Characteristics: With the 300 ng/mL qualitative application | | |
| Assay range | Up to 2000 ng/mL | Up to 600 ng/mL |
| Precision: | | |
| Negative reading | > 95% confidence at 80% cutoff | >95% confidence at 80% cutoff |
| Positive reading | > 95% confidence at 120% cutoff | > 95% confidence at 120% cutoff |
| Accuracy<br>(% Agreement) | N = 40<br>100 % vs. GC/MS | N = 49<br>100 % vs. GC/MS |
| Sensitivity<br>(Analytical) | < 12 ng/mL | < 5.0 ng/mL |
| Performance Characteristics: With the 2000 ng/mL quantitative application | | |
| Assay range | Up to 8000 ng/mL | Up to 4000 ng/mL |
| Precision<br>(Within-run) | 2.2 % at 1000 ng/mL<br>3.3 % at 1600 ng/mL<br>2.8 % at2000 ng/mL<br>3.8 % at 2400 ng/mL<br>4.2 % at 4000 ng/mL | 1.7 % at 246 ng/mL<br>1.7 % at 297 ng/mL<br>0.7 % at 359 ng/mL |
| Accuracy<br>(% Agreement) | N = 42<br>100 % vs. GC/MS | N = 49<br>100 % vs. GC/MS |
| Sensitivity<br>(Analytical) | < 12 ng/mL | < 5.0 ng/mL |
{4}------------------------------------------------
Roche Diagnostic Systems, Inc. Somerville, NJ April 1998
Abuscreen OnLine II Opiates 300/2000 Reagent, Calibrator & Controls Summary
Table 3 - (Cont.')
| Table 4 - Abuscreen OnLine Opiates Culibration Pack | | | |
|-----------------------------------------------------|--|--|--|
|-----------------------------------------------------|--|--|--|
| | Abuscreen ONLINE Opiates<br>Calibration Pack | Current Abuscreen<br>ONLINE MiL Cal Pack |
|-----------------------------|----------------------------------------------|------------------------------------------|
| Matrix | human urine | human urine |
| Constituents | morphine HCL | morphine sulfate |
| Calibrator Values:<br>ng/mL | 0 | 0 |
| | 600 | 1000 |
| | 1000 | 2000 |
| | 2000 | 4000 |
| | 4000 | |
| | 8000 | |
{5}------------------------------------------------
Roche Diagnostic Systems, Inc. Somerville, NJ April 1998
| | Abuscreen ONLINE<br>Opiates Control Pack | | Abuscreen ONLINE<br>Controls | |
|--------------------|------------------------------------------|----------------|------------------------------|-------------|
| | Negative | Positive | Negative | Positive |
| Matrix | human<br>urine | human<br>urine | human urine | human urine |
| Drugs: | ng/mL | ng/mL | ng/mL | ng/mL |
| Amphetamines | | | 500 | 1500 |
| Barbiturates | | | 100 | 300 |
| Benzodiazepines | | | 50 | 150 |
| Cannabinoids | | | 20 | 75 |
| Cocaine | | | 150 | 450 |
| Methadone | | | 150 | 450 |
| Methaqualone | | | 150 | 450 |
| Opiates (Morphine) | 1000 | 3000 | 150 | 450 |
| PCP | | | 12.5 | 38 |
| Propoxyphene | | | 150 | 450 |
# Table 5 - Abuscreen OnLine Opiates Control Pack
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized eagle-like symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement around the emblem. The emblem itself consists of three stylized lines forming the shape of a bird or wing.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR 2 4 1998
James W. Haynes Requlatory Affairs Associate Roche Diagnostics Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
K974840 Re : Abuscreen OnLine II for Opiates 300/2000 Reagent, Calibration Pack and Control Pack II - - - - - - -Regulatory Class: Product Code: DJG, DJG, DJJ, DLR Dated: April 3, 1998 April 6, 1998 Received:
Dear Mr. Haynes:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਕੇ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Page 1 _of 1
510(k) Number (if known)
Device Name: Abuscreen OnLine II for Opiates 300/2000 Abuscreen OnLine Opiates Calibration Pack Abuscreen OnLine Opiates Control Pack
Indications for Use:
The Abuscreen OnLine II for Opiates 300/2000 Reagent is an in vitro diagnostic test for the qualitative (at 300 and 2000 ng/mL cutoff) and semi-quantitative (at 2000 ng/mL cutoff) detection of morphine and its metabolites in human urine.
The Abuscreen ONLINE Opiates Calibration Pack is an in vitro diagnostic device designed for the calibration of the Roche reagent assays for opiates.
The Roche Abuscreen ONLINE Opiates Control Pack is an assayed quality control sample for use with the Roche assays for opiates.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
ﺼﺎ، ﻓ
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division:
Division:
510(k) Number: K974840
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.