ALPHA DX ANALYZER 01-0100, ALPHA DX MYO/CK/CK-MB/TNI CARDIAC PANEL TEST KIT 03-0001, ALPHA DX CK/CK-MB PANEL TEST KIT
K974839 · First Medical, Inc. · MMI · Feb 4, 1998 · Clinical Chemistry
Device Facts
Record ID
K974839
Device Name
ALPHA DX ANALYZER 01-0100, ALPHA DX MYO/CK/CK-MB/TNI CARDIAC PANEL TEST KIT 03-0001, ALPHA DX CK/CK-MB PANEL TEST KIT
Applicant
First Medical, Inc.
Product Code
MMI · Clinical Chemistry
Decision Date
Feb 4, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
The Alpha Dx Cardiac Panel Test Kit is intended for use with the Alpha Dx Analyzer to provide rapid, quantitative and simultaneous measurement of myoglobin, creatine kinase (CK), creatine kinase MB isoenzyme (CK-MB) and cardiac troponin I (TnI) in whole blood or serum to aid in the diagnosis of acute myocardial infarction (AMI). The Analyzer and Cardiac Panel Test Kit combine ease of use and rapid turnaround time with laboratory-quality performance and reliability. With factory calibration, built-in bilevel quality controls and closed tube sampling, the system can be used in the central laboratory, STAT lab, emergency department, coronary care unit, chest pain center and other point of care locations.
Device Story
Alpha Dx System is a benchtop, fully integrated fluorescence immunoassay analyzer used with specific cardiac panel test kits. It accepts whole blood or serum samples to measure myoglobin, CK, CK-MB, and TnI. The system utilizes immunofluorometric procedures and fluorometric detection to provide quantitative results. Designed for use in central labs, STAT labs, emergency departments, coronary care units, and chest pain centers by healthcare professionals. It features factory calibration, built-in bilevel quality controls, and closed-tube sampling. The output provides rapid, laboratory-quality diagnostic data to clinicians to assist in the diagnosis of acute myocardial infarction (AMI), facilitating faster clinical decision-making in acute care settings.
Clinical Evidence
Clinical evaluation included testing 410 chest pain patients and 134 healthy individuals. Correlation studies between Alpha Dx and predicate methods (Stratus, Vitros, Access) showed high concordance (r=0.88 to 0.99). Clinical sensitivity for TnI was 93% (95% CI) and specificity was 94% (95% CI). Concordance between Alpha Dx TnI and Stratus CK-MB was 92%, with no statistically significant difference in diagnostic performance per McNemar's test.
Technological Characteristics
Benchtop fluorescence immunoassay analyzer. Uses immunofluorometric procedure with monoclonal and polyclonal antibodies. Employs fluorometric detection. Features factory calibration, built-in bilevel quality controls, and closed-tube sampling. Designed for point-of-care and laboratory environments.
Indications for Use
Indicated for the rapid, quantitative, and simultaneous measurement of myoglobin, CK, CK-MB, and TnI in whole blood or serum to aid in the diagnosis of acute myocardial infarction (AMI) in patients presenting with chest pain.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Dade Stratus Cardiac Troponin I Fluorometric Enzyme Immunoassay
Beckman Access Troponin I Immunoassay System
Dade Stratus II Fluorometric Immunoassay System
Submission Summary (Full Text)
{0}------------------------------------------------
### FB
Image /page/0/Picture/1 description: The image shows a black and white drawing of a swirling vortex. The vortex is composed of several curved lines that spiral inward towards a central point. The lines are thick and bold, creating a sense of motion and energy. The overall impression is one of a dynamic and powerful force.
### MEDICAL, INC. FIRST
## X 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
- Alpha Dx System
- A. Name and Address of Submitter
- . Company Name and Address:
- Telephone: ●
- FAX: .
Contact Person: ●
- . Date 510(k) summary was prepared:
- Device Names B.
- 1. Trade Names
The Alpha Dx System consists of:
- The Alpha Dx Analyzer, and .
- The Alpha Dx Myo/CK/CK-MB/TnI Cardiac Panel Test Kit, or .
- The Alpha Dx CK/CK-MB Panel Test Kit, or .
- The Alpha Dx Myo/TnI Panel Test Kit .
- Common Names 2.
- a. Alpha Dx Analyzer
Fully integrated benchtop analyzer capable of performing in vitro diagnostic measurement of a panel of tests.
- b. Alpha Dx Panel Test Kits
Rapid, quantitative and simultaneous measurement of myoglobin (Myo), creatine kinase (CK), creatine kinase MB isoenzyme (CK-MB) and cardiac troponin I (TnI) in whole blood or serum to aid in the diagnosis of acute myocardial infarction (AMI).
- 3. Classification Names
- a. Alpha Dx Analyzer
Fluorescence Immunoassay Analyzer for use with the Alpha Dx Panel Test Kits for the rapid measurement of a panel of tests (Clinical Chemistry Panel Classification Device List).
- First Medical, Inc. 530 Logue Avenue Mountain View, CA 94043 (650) 903-5974 (650) 903-9040 Van N. Johnson December 23, 1997
{1}------------------------------------------------
- b. Alpha Dx Panel Test Kits
Fluorescence Immunoassays for myoglobin, CK, CK-MB and TnI (Clinical Chemistry Panel and Immunology Classification Device List).
- C. Legally Marketed Devices
- 1. Panel Test Kits
- Dade Stratus Myoglobin Fluorometric Enzyme Immunoassay .
- Johnson & Johnson Vitros CK Slide .
- Dade Stratus CK-MB Fluorometric Enzyme Immunoassay ●
- Dade Stratus Cardiac Troponin I Fluorometric Enzyme Immunoassay ●
- . Beckman Access Troponin I Immunoassay System
- 2. Analyzer
- Dade Stratus II Fluorometric Immunoassay System ●
- D. Device Description
The Alpha Dx Cardiac Panel Test Kits, when used with the Alpha Dx Analyzer, are fluorescence immunoassays for the rapid measurement of myoglobin, CK, CK-MB and TnI in human whole blood or serum.
- E. Intended Use
The Alpha Dx Cardiac Panel Test Kit is intended for use with the Alpha Dx Analyzer to provide rapid, quantitative and simultaneous measurement of myoglobin, creatine kinase (CK), creatine kinase MB isoenzyme (CK-MB) and cardiac troponin I (TnI) in whole blood or serum to aid in the diagnosis of acute myocardial infarction (AMI). The Analyzer and Cardiac Panel Test Kit combine ease of use and rapid turnaround time with laboratory-quality performance and reliability. With factory calibration, built-in bilevel quality controls and closed tube sampling, the system can be used in the central laboratory, STAT lab, emergency department, coronary care unit, chest pain center and other point of care locations.
- F. Comparison of Methodology with Predicate Devices
The Alpha Dx Cardiac Panel Test Kit, when used with the Alpha Dx Analyzer, provides quantitative determinations of human myoglobin, CK, CK-MB and TnI in whole blood or serum. The current Stratus Myoglobin, CK-MB and Troponin I Fluorometric Enzyme Immunoassays when used with the Stratus II Fluorometric System Analyzer, the Vitros CK Slides when used with the Vitros 700 and the Access Troponin I Assay when used with the Access Immunoassay System are examples of other devices currently marketed for the same uses.
The Alpha Dx Cardiac Panel Test Kits are two-site fluorescence immunoassays. The current Stratus Myoglobin, CK-MB and Cardiac Troponin I Fluorometric Enzyme Immunoassays and the Stratus II Fluorometric Immunoassay Analyzer are examples of currently available products that employ the immunofluorometric procedure and fluorometric detection.
{2}------------------------------------------------
Monoclonal and polyclonal antibodies are used in the Alpha Dx Cardiac Panel Test Kits. The use of monoclonal and polyclonal antibodies for immunoassay procedures has been clearly demonstrated in other currently commercially available assays and in numerous scientific reports.
- G. Summary of Analytical Data
- 1. Dilution Recoveries
The mean recoveries by diluting serum samples with high myoglobin, CK, CK-MB or TnI are as follows:
| Analyte | VIVO | AV | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>AV 120<br>JA-MD | 1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTRACTORIAL COLLEGICAL COLLEGION CONSULTION OF<br>. Mean % Recoveries<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | <br>C | AND LIES GENER<br>1<br>A SECT ANNUAL<br>------------------------------------------------------------------------------------------------------------------------------------------------ | 103<br>Carmer Research and Concell a co | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-----------------------------------------------------------------------------------------------<br>A CONSULTION COLLEGION OF COLLEGION |
Table 43 Dilution Recoveries
- Analytical Sensitivity 2.
The analytical sensitivity for the Alpha Dx myoglobin, CK, CK-MB and TnI assays are 1.1 ng/mL, 1.8 ng/mL, 0.04 ng/mL, 0.017 ng/mL, respectively.
- 3. Precision
Median %CV of the assays are as follows:
| %CV | Myo | CK | CK-MB | Tnl |
|---------------------|-----|-----|-------|-----|
| Whole Blood Samples | 3.9 | 5.3 | 3.9 | 7.3 |
| Serum Samples | 4.0 | 7.0 | 2.7 | 2.4 |
Table 44 Precision
- H. Summary of Clinical Data
- 1. Correlation Between Paired Whole Blood and Serum Samples Tested with the Alpha Dx System
Paired whole blood and serum samples from serial draws of 72 chest pain patients and 24 healthy individuals were tested. Linear regression analyses, after exclusion of samples with levels outside the reportable ranges of the test methods, are as follows:
| Method | Myo | CK | CK-MB | Tnl |
|-----------|------|------|-------|------|
| N | 322 | 309 | 294 | 73 |
| Slope | 1.05 | 1.04 | 1.07 | 0.90 |
| Intercept | -3 | 5 | -0.1 | 0.08 |
| r | 0.99 | 0.99 | 1.00 | 0.99 |
Table 45 Correlation Between Whole Blood and Serum
{3}------------------------------------------------
- Correlations Between Whole Blood Samples Assayed with the Alpha Dx System and 2. Paired Serum Samples Assayed with the Predicate Methods
The relation between myoglobin, CK, CK-MB and TnI measurements with the Alpha Dx System and the predicate devices was determined in serial samples from 410 chest pain patients and 134 healthy individuals, after exclusion of samples with levels outside the reportable range of the test methods. Results of the regression analyses are as follows:
| | Myo | CK | CK-MB | TnI | TnI |
|-------------------|---------|--------|---------|---------|--------|
| Predicate | Stratus | Vitros | Stratus | Stratus | Access |
| N | 955 | 923 | 676 | 226 | 244 |
| Slope | 0.81 | 0.77 | 1.07 | 0.18 | 1.44 |
| Intercept (ng/mL) | -2 | -1 | 1.1 | 0.56 | 0.96 |
| r | 0.98 | 0.97 | 0.97 | 0.93 | 0.88 |
Table 46 Correlation Between Alpha Dx System and Predicate Methods
- 3. Clinical Sensitivity and Specificity of the Alpha Dx TnI, Stratus TnI, Access TnI and Stratus CK-MB Assays
The clinical cutoffs, sensitivity and specificity for the Alpha Dx TnI, Stratus TnI, Access TnI and the Stratus CK-MB are as follows:
| Method | Cutoff Value<br>(ng/mL) | Sensitivity<br>(95% Cl) | Specificity<br>(95% Cl) |
|---------------|-------------------------|-------------------------|-------------------------|
| Alpha Dx Tnl | 0.40 | 93 ± 6 | 94 ± 3 |
| Stratus Tnl | 1.50 | 93 ± 6 | 93 ± 3 |
| Access Tnl | 0.15 | 92± 6 | 88 ± 4 |
| Stratus CK-MB | 7.0 | 89± 7 | 90 ± 4 |
Table 47 Tnl and CK-MB Sensitivity and Specificity Comparison
Sensitivity and specificity for the Alpha Dx TnI and Stratus CK-MB are comparable within the 95% confidence intervals of the two methods. The Alpha Dx TnI assay results were 92% concordant versus the Stratus CK-MB results in 362 subjects tested with both devices. McNemar's test indicates no statistically significant difference between the two assays in diagnostic performance.
- Conclusion I.
The Alpha Dx Cardiac Test Discs are used with the Alpha Dx Analyzer for the rapid, quantitative and simultaneous measurement of myoglobin, creatine kinase (CK), creatine kinase MB isoenzyme (CK-MB) and cardiac troponin I (TnI) in whole blood or serum to aid in the diagnosis of acute myocardial infarction (AMI). The data shows that the Alpha Dx Cardiac Test Discs are substantially equivalent to other tests currently in commercial distribution for the purpose of analyzing unknown concentrations of myoglobin, CK, CK-MB and TnI.
{4}------------------------------------------------
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
### 1 1338 FR
Van N. Johnson . Vice President, Clinical and Regulatory Affairs First Medical, Inc. 530 Logue Avenue Mountain View, California 94043
Re : K974839 Alpha DX System Requlatory Class: I, II Product Code: MMI,JHX, DDR, CGX, KHO December 23, 1997 Dated: Received: December 24, I997
Dear Ms. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set """" forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note:
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as The FDA described in your 510 (k) premarket notification. finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page **_ / of _** /
K97483 510(k) Number (if known):_
Device Name:
Indications For Use:
Statement of Indication for Use
The Alpha Dx Cardiac Panel Test Kit is intended for use with the Alpha Dx Analyzer to provide rapid, quantitative and simultaneous measurement of myoglobin, creatine kinase (CK), creatine kinase MB isoenzyme (CK-MB) and cardiac troponin I (TnI) in whole blood or serum to aid in the diagnosis of acute myocardial infarction (AMI). The Analyzer and Cardiac Panel Test Kit combine ease of use and rapid turnaround time with laboratory-quality performance and reliability. With factory calibration, built-in bilevel quality controls and closed tube sampling, the system can be used in the central laboratory, STAT lab, emergency department, coronary care unit, chest pain center and other point of care locations.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qyruic
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. 4974839
Over-The-Counter Use
Prescription Use_ 2 (Per 21 CFR 801.109) OR
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.