K974820 · Sunrise Medical, Inc. · IOR · Jan 27, 1998 · Physical Medicine
Device Facts
Record ID
K974820
Device Name
BREEZY SERIES
Applicant
Sunrise Medical, Inc.
Product Code
IOR · Physical Medicine
Decision Date
Jan 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
Pediatric
Indications for Use
Quickie manual wheelchairs empower physically challenged persons by providing a means of mobility. This includes temporary and permanent conditions in all ages such as: Arthritis, Amputee, Paraplegic, Cerebral Palsy, Hemiplegic, Tetraplegic, Quadriplegic, Spina Bifida, Head Injury or Trauma, Muscular Dystrophy, Multiple Sclerosis, Polio, Geriatric conditions And other immobilizeing or debilitating condition. A wheelchair with Bus Transport Option may be used for motor vehicle transportation, with the use of wheelchair tie down and occupant restraint system (WTORS) that meet the requirements of SAE J2249
Device Story
Breezy series are lightweight manual wheelchairs providing mobility for physically challenged users. Devices consist of seat frame, backrest, cushion, footrest, and casters; accessories include armrests, positioning belts, oxygen tank holders, and IV poles. Key update is addition of reclining back model. Chairs are configured to customer orders based on user needs/capabilities; dimensions (width, depth, back height) are adjustable. Used in indoor/outdoor environments with firm surfaces. Healthcare professionals and users determine appropriate configuration. Reconfiguration possible as user needs evolve. Benefits include improved mobility and maneuverability.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Manual wheelchair; lightweight frame construction. Adjustable dimensions (width, depth, back height). Components include backrest, seat frame, cushion, footrest, casters. Accessories include positioning belts, IV poles, oxygen tank holders. No electronic components or software.
Indications for Use
Indicated for physically challenged individuals of all ages with temporary or permanent mobility-limiting conditions, including arthritis, amputation, paraplegia, cerebral palsy, hemiplegia, tetraplegia, quadriplegia, spina bifida, head injury/trauma, muscular dystrophy, multiple sclerosis, polio, or geriatric conditions.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Sunrise Invacare's Everest & Jennings
Breezy Family 9000 Universal
EZ LITE
Submission Summary (Full Text)
{0}------------------------------------------------
510k Notification - Breezy 510 Recliner and Model Number Updates
#### 13.0 510(k) SUMMARY OF SAFETY AND EFFICACY
Rebecca Andersen Submitter: Date: December 18, 1997
R974820
JAN 27 1998
Name(s) of the device(s):
## Breezy Series
Identification of predicate device(s):
Sunrise Invacare's Everest & Jennings' Everest and Jennings
Breezy Family 9000 Universal EZ LITE
# Description of the device:
The Breezy series wheelchairs are light weight manual chairs. These chairs are intended to provide mobility based on an individual users' needs and capabilities. They are available in a range of sizes (dimensions) to allow fit to a particular user, and is adjustable in various ways, including: The key change covered by this submission is the addition of a reclining back model to the Breezy product line.
Chairs are configured to customer orders when sold. Configuration to order allows flexibility. Features such as width, depth, and back height are specified. Specific configuration includes chair accessories and desired safety features. Any combination of features from the approved matrix may be configured to create the chair features needed by a given rider It is the rider and the health care professional that determine the appropriate configuration for the users needs. As the individuals condition or size change, other components or accessories may be ordered and the chair reconfigured to meet the evolving needs of its user.
Breezy wheelchairs with or without the reclining back option, consists of typical components found on most wheelchairs, such as backrest, seat frame, cushion, footrest and casters. Accessories include items such as armrests, positioning belts, backpacks, seat pouches, oxygen tank holders, IV poles, etc.
Many of these components may become available in a range of sizes, shapes, angles, forms, materials or coverings. These variations allow the chairs to be configured to meet the specific desires and needs of the user.
The chairs have excellent performance indoors and are very good outdoors over surfaces that are firm ad free of large obstacles and long steep inclines. That makes them an ideal maneuverable, light weight "Get out there chair".
Warnings, cautions and contraindications are detailed in the users manual.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized image of an eagle. The eagle is depicted with three curved lines representing its body and wings. The text is in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 27 1998
Ms. Rebecca Andersen · Vice President, Quality and Requlatory Affairs Sunrise Medical 7477 East Dry Creek Parkway Longmont, Colorado 80503
Re: K974820 Breezy Series Regulatory Class: I Product Code: IOR Dated: December 16, 1997 Received: December 23, 1997
Dear Ms. Andersen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{2}------------------------------------------------
#### Page 2 - Ms. Rebecca Andersen
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Acotello
CelNa M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# 12.2 Indications for Use
### Intended use:
Quickie manual wheelchairs empower physically challenged persons by providing a means of mobility. This includes temporary and permanent conditions in all ages such as:
| Arthritis |
|----------------|
| Amputee |
| Paraplegic |
| Cerebral Palsy |
| Hemiplegic |
Tetraplegic Quadriplegic Spina Bifida Head Injury or Trauma Muscular Dystrophy
Multiple Sclerosis Polio Geriatric conditions And other immobilizeing or debilitating condition
A wheelchair with Bus Transport Option may be used for motor vehicle transportation, with the use of wheelchair tie down and occupant restraint system (WTORS) that meet the requirements of SAE J2249
510(k) number: Not assigned as of this time
Device name: Breezy Series
Which today consist of : Breezy 500 Breezy 510 Breezy 510 Recliner Breezy 600
Concurrence of CDRH, Office of Device Evaluation (ODE)
[ Prescription use (per 21 CFR801.109)
Over-the-counter use
coop
(D
· Restorative Devic 5 Tulk) Number
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three waves emanating from its body, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 7 1998
Mr. Zhou Shun Kao Wenzhou Hindar Optical Co., Ltd. No. 46-5 Liming Mid Road Wenzhou City, China
Re: K974861 Trade Name: Non-prescription Sunglasses Regulatory Class: I Product Code: 86 HQY Dated: December 17, 1997 Received: December 29, 1997
Dear Mr. Kao:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21-CFR-Part 820) and that, through ' periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Zhou Shun Kao
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of
510(k) Number (if known): K974861
Device Name: Sunglasses
Indications For Use:
Sunglasses (nonprescription) are devices that consist of spectacle frames or clips with absorbing, reflective, tinted, polarizing, or photosensitized lenses intended to be worn by a person to protect the eyes from bright sunlight but not to provide refractive corrections. This device is usually available over-the-counter.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Am Williams
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K974861
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use XXX
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.