← Product Code [DRC](/productcode/DRC) · K974742

# MANAN POTTS-COURNAND NEEDLE (K974742)

_Medical Device Technologies, Inc. · DRC · Apr 6, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K974742

## Device Facts

- **Applicant:** Medical Device Technologies, Inc.
- **Product Code:** [DRC](/productcode/DRC.md)
- **Decision Date:** Apr 6, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1390
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Manan™ Potts-Cournand Needle is intended for use for carotid angiography, direct arterial pressure monitoring, blood sampling, and percutaneous catheterization.

## Device Story

Manan™ Potts-Cournand Needle is a three-piece manual surgical instrument used for vascular access. Device consists of an outer cannula with a short, tapered bevel, a matching inner cannula with a chisel bevel, and an obturator with a rounded, blunt tip. The inner cannula extends past the outer cannula; the obturator extends past the inner cannula. Used by clinicians in clinical settings for carotid angiography, arterial pressure monitoring, blood sampling, and percutaneous catheterization. Device facilitates vessel entry and catheter placement.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Three-piece assembly: outer cannula, inner cannula, and obturator. Gauges: 15 to 21 ga. Lengths: 2 to 2-1/2 inches. Manual operation. Non-powered.

## Regulatory Identification

A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.

## Special Controls

*Classification.* Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Potts-Cournand needle ([K851832](/device/K851832.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K974742

Image /page/0/Picture/1 description: The image shows the logo for MDTECH MEDICAL DEVICE TECHNOLOGIES INC. The logo is in black and white, with the text "MDTECH" in large, bold letters. Below that, in smaller letters, is the text "MEDICAL DEVICE TECHNOLOGIES INC.". The logo is simple and professional, and it clearly communicates the company's name and industry.

4445-310 S.W. 35th Terrace Gainesville, Florida 32608 TEL: 352/338-0440 FAX: 352/338-0662 APR - 6 1998

## 510(k) SUMMARY

| APPLICANT:               | Medical Device Technologies, Inc.<br>4445-310 SW 35th Terrace<br>Gainesville, FL 32608 |              |           |
|--------------------------|----------------------------------------------------------------------------------------|--------------|-----------|
| CONTACT:                 | Karl Swartz<br>Quality Assurance Manager                                               |              |           |
| TELEPHONE:               | (352)338-0440<br>fax (352)338-0662                                                     |              |           |
| TRADE NAMES:             | Manan™ Potts-Cournand Needle                                                           |              |           |
| COMMON NAME:             | Potts-Cournand Needle                                                                  |              |           |
| CLASSIFICATION NAME:     | Angiographic needle - 74HAQ                                                            |              |           |
| SUBSTANTIAL EQUIVALENCE: | Company Name                                                                           | Product Name | 510(k) No |

| Manan Medical Products | Potts-Cournand needle | K851832 |
|------------------------|-----------------------|---------|
|------------------------|-----------------------|---------|

## DESCRIPTION OF DEVICE:

This needle is made in various gauges from 15 to 21 ga, and in various lengths from 2 to 2-1/2 in. It is a three piece assembly made up of an outer cannula with a short, tapered bevel which is matched to the inner cannula. The inner cannula has chisel bevel which extends past the outer cannula. An obturator with a rounded, blunt tip fits inside of the inner cannula, and extends past the tip of the inner cannula.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, arranged in a stacked formation. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 6 1998

Mr. Karl Swartz Quality Assurance Manager Medical Device Technologies, Inc. 4445-310 SW 35th Terrace Gainsville, FL 32608

K974742, K974743 and K974745 Re : Trade Name: Manan™ Potts-Cournand Needle, Manan™ Seldinger Needle and Manan™ GWI Guide Wire Introducer Requlatory Class: II Product Code: DRC Dated: March 18, 1998 Received: March 19, 1998

Dear Mr. Swartz:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) intoceither classella (specialment mu Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In

{2}------------------------------------------------

Page 2 - Mr. Karl Swartz

addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahon

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and ... ...... Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MDTECH Medical Device Technologies Inc. The logo is in black and white, with the letters "MDTECH" in a stylized font. The words "MEDICAL DEVICE TECHNOLOGIES INC" are printed in a smaller font below the letters "MDTECH". The logo is rectangular in shape.

-310 S.W. 35th Gainesville, Florida 32608 TEL: 352/338-0440 FAX: 352/338-0662

Page_1_of_1

510(k) Number (if known): ___ K97 47 42

Device Name: Manan™ Potts-Cournand Needle

Indications for Use:

The Manan™ Potts-Cournand Needle is intended for use for carotid angiography, direct arterial pressure monitoring, blood sampling, and percutaneous catheterization.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ta. A. he for

---

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 14974742 510(k) Number .

Prescription Use_ (Per 21 CFR 801.109)

OR

Over-The-Counter Use_____________________

(Optional Format 1-2-96)

11. 11. 11. 11.

---

**Source:** [https://fda-staging.innolitics.com/device/K974742](https://fda-staging.innolitics.com/device/K974742)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com