HEARTPORT DIRECT AORTIC RETURN CANNULA WITH INTRODUCER

K974736 · Heartport, Inc. · DWF · Jul 15, 1998 · Cardiovascular

Device Facts

Record IDK974736
Device NameHEARTPORT DIRECT AORTIC RETURN CANNULA WITH INTRODUCER
ApplicantHeartport, Inc.
Product CodeDWF · Cardiovascular
Decision DateJul 15, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Direct Aortic Return Cannula is indicated for patients undergoing cardiopulmonary bypass. It is intended to deliver oxygenated blood for cardiopulmonary bypass during surgery. The Direct Aortic Return Cannula also allows the hemostatic introduction and removal of vascular catheters such as the Heartport® Endoaortic Clamp™ catheter. The cannula may be introduced through a 12mm or larger trocar or incision. The Cannula Introducer is intended for use with Heartport arterial return cannulae. It is intended for incising the vessel and introducing the cannula into the vessel.

Device Story

Direct Aortic Return Cannula with Introducer; used during cardiopulmonary bypass surgery. Cannula delivers oxygenated blood to patient; facilitates hemostatic introduction/removal of vascular catheters (e.g., Endoaortic Clamp catheter). Introducer used to incise vessel and guide cannula placement. Operated by surgeons in OR setting. Device provides mechanical access and fluid delivery path; benefits patient by enabling minimally invasive or standard cardiopulmonary bypass procedures.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Cardiovascular surgical cannula and introducer. Materials verified for biocompatibility. Mechanical device; no energy source or software components.

Indications for Use

Indicated for patients undergoing cardiopulmonary bypass requiring delivery of oxygenated blood during surgery and hemostatic introduction/removal of vascular catheters.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ K974736 510(k) Notification ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ APPENDIX B. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS {1}------------------------------------------------ This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ (650) 482-4346 ### Applicant Information: | Date Prepared: | December 18, 1997 | |-----------------|------------------------------------------------| | Name: | Heartport, Inc. | | Address: | 200 Chesapeake Drive<br>Redwood City, CA 94063 | | Contact Person: | Marianne C. Drennan | | Phone Number: | (650) 482-4405 | ### Device Information: Fax Number: | Classification | Class II | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------| | Trade Name: | Heartport® Direct Aortic Return Cannula with Introducer | | Classification Name: | Cardiovascular Surgical Devices - Cardiopulmonary bypass vascular<br>cannula (21 CFR 870.4210); Catheter introducer (21 CFR 870.1340) | # Equivalent Devices: The subject device is substantially equivalent in intended use and method of operation to the currently marketed Heartport® Direct Aortic Return Cannula, the Heartport® Endoarterial Return™ Cannula and the Cardiovascular Research, Inc. Cannula Introducer. #### Intended Use: The Direct Aortic Return Cannula is indicated for patients undergoing cardiopulmonary bypass. It is intended to deliver oxygenated blood for cardiopulmonary bypass during surgery. The Direct Aortic Return Cannula also allows the hemostatic introduction and removal of vascular catheters such as the Heartport® Endoaortic Clamp™ catheter. The cannula may be introduced through a 12mm or larger trocar or incision. The Cannula Introducer is intended for use with Heartport arterial return cannulae. It is intended for incising the vessel and introducing the cannula into the vessel. ## Non-Clinical Test Results: Performance testing demonstrated that the Heartport Direct Aortic Return Cannula with Introducer meets established specifications. The materials used in the Heartport Direct Aortic Return Cannula with Introducer have proven biocompatibility. #### Summary: Based on the intended use, product performance and biocompatibility information provided in this notification, the subject device has been shown to be substantially equivalent to currently marketed predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three heads facing right, representing the department's focus on health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 5 1998 Ms. Marianne C. Drennan Regulatory Affairs Manager HeartPort, Inc. 200 Chesapeake Drive Redwood City, CA 94063 Re : K974736 HeartPort® Direct™ Aortic Return Cannula with Introducer Regulatory Class: II (Two) Product Code: DWF Dated: May 22, 1998 May 26, 1998 Received: Dear Ms. Drennan: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of The general controls provisions of the Act include the Act. requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth-in the Quality -System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Ms. Marianne C. Drennan This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding described in your stority promazion nevice to a legally marketed or babbuantana a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance Also, please note the regulation entitled, at (301) 594-4639. "Misbranding by reference to premarket notification" (21 CFR Other general information on your responsibilities under 807.97) . . the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Division Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): _ K 97 4736 Heartport® Direct Aortic Return Cannula with Introducer Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications for Use: The Direct Aortic Return Cannula is indicated for patients undergoing cardiopulmonary bypass. It is intended to deliver oxygenated blood for cardiopulmonary bypass during surgery. The Direct Aortic Return Cannula also allows the hemostatic introduction and removal of vascular catheters such as the Heartport® Endoaortic Clamp™ catheter. The cannula may be introduced through a 12mm or larger trocar or incision. The Cannula Introducer is intended for use with Heartport arterial return cannulae. It is intended for incising the vessel and introducing the cannula into the vessel. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bore L. Spencer (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K974736 Prescription Use V (Per 21 CFR 801.109), OR Over- The Counter Use (Optional Format 1-2-96) Heartport Confidential
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...