MILLENNIA 3500 SERIES MONITOR WITH ANTHESTHETIC AGENT
K974581 · Invivo Research, Inc. · DRT · Oct 28, 1998 · Cardiovascular
Device Facts
Record ID
K974581
Device Name
MILLENNIA 3500 SERIES MONITOR WITH ANTHESTHETIC AGENT
Applicant
Invivo Research, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Oct 28, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitoring is intended for general hospital or clinical use by medical professionals whenever it is required to monitor concentrations of anesthetic gases. The agents monitored include halothane, enflurane, isoflurane, sevoflurane and desflurane. Additionally, this monitor also monitors end-tidal carbon dioxide, nitrous oxide and oxygen gas concentrations. The need to monitor these anesthetic agents is most commonly encountered in the operating room during administration of anesthesia. This device is available for sale only upon the order of a physician or other related licensed medical professional.
Device Story
Patient monitor for anesthesia gas concentrations; draws continuous gas sample from breathing circuit; uses NDIR (non-dispersive infrared) optical bench to measure infrared light absorption for agent/CO2/N2O identification and concentration; uses polarographic cell for oxygen measurement; operated by medical professionals in hospital/clinical settings (primarily OR); displays numerical values and waveforms on front panel; provides real-time monitoring to assist clinical decision-making during anesthesia administration; benefits patient by ensuring appropriate gas delivery and ventilation monitoring.
Clinical Evidence
Bench testing only. Device conforms to national and international product safety standards for electrical, electromagnetic compatibility, and anesthetic agent monitoring.
Technological Characteristics
Polychromatic non-dispersive infrared (NDIR) system for gas analysis; polarographic oxygen cell; resistive infrared light source; optical filters for gas identification; analog and digital circuit boards; AC/battery power; moisture collection system.
Indications for Use
Indicated for patients requiring monitoring of anesthetic gases (halothane, enflurane, isoflurane, sevoflurane, desflurane), end-tidal carbon dioxide, nitrous oxide, and oxygen concentrations, typically in an operating room setting.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Criticare Poet II Anesthesia Agent Monitor (K892646, K942737)
Submission Summary (Full Text)
{0}------------------------------------------------
#### OCT 28 1998
# 510(k) Summary of Safety and Effectiveness
#### Submitter Information:
Invivo Research Inc. 12601 Research Parkway Orlando, FL 32826 407-275-3220 Contact: Mr. Francis Casey
#### Product Name:
| Proprietary: | Millennia 3500 Series Monitor with Anesthetic Agent Monitoring |
|-----------------|-----------------------------------------------------------------------|
| Common: | Anesthetic Gas Analyzer |
| Classification: | Anesthetic Gas Analyzer, enflurane - (73 CFR/21CFR 868.1500 Class II) |
| | Anesthetic Gas Analyzer, halothane - (73 CFR/21CFR 868.1620 Class II) |
| | Oxygen Gas Analyzer - (73 CFR/21CFR 868.1720 Class II) |
#### Predicate Devices:
The predicate device(s) are the Datex AS/3 Anesthesia Agent Monitor (510(k) numbers K832113, K903104, K923276, K932098), the Datascope Multinex Anesthetic Agent Monitor (510(k) numbers K883254, K903104), and the Criticare Poet II Anesthesia Agent Monitor (510(k) numbers K892646, K942737). These predicate devices have the same performance specifications as the Millennia 3500 Series Monitor with Anesthetic Agents Monitor.
#### Device Description:
The Millennia 3500 Series Monitor with Anesthetic Agents Monitor is a patient monitor which draws a continuous flow of sample gas from the patient's anesthesia breathing circuit. The sample is then drawn through a measurement sample cell where the agent(s) and other patient gas concentrations are measured. Measured values are displayed as numbers and waveforms on the main display of the monitor. The monitor is operated by controls on the front panel of the monitor. The major functional subsystems of the device are the power supply, the pneumatic system, the measurement sample cell, the system electronics and product software. Power is supplied by an AC power cord, or an internal set of batteries. AC power is internally conditioned by the power supply subassembly, which powers the system and charges the internal batteries. The pneumatic system draws sample gas into the device through a moisture collection system. The sample then enters the sample cell, which is a cavity within the optical bench assembly. The sample is then exhausted out the rear of the unit. The sample cell cavity allows infrared light to be passed through the sample so a detector on the other side of the sample can measure the amount of infrared light absorption. The infrared light source consists of a resistive
{1}------------------------------------------------
element which radiates a relatively wide spectrum of infrared light. The amount of transmitted infrared light is measured by an optical detector, and several optical filters are used to determine the identity and concentration of the anesthetic agents, nitrous oxide and carbon dioxide. The oxygen measurements are obtained using a polarographic oxygen cell which also receives sample gas from the sample cell. The electronic system consists of an analog circuit board, and a digital circuit board. These boards are added to the existing standard Millennia 3500 Series Patient Monitoring System. The analog board takes the electrical signals from the sample cell assembly and conditions them so that measurement calculations can be performed. The digital circuit board digitizes the conditioned analog signals and performs measurement calculations. Proprietary software algorithms are used to calculate measured concentrations of anesthetic agents and patient gases in the sample gas.
## Intended Uses:
The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitoring is intended for general hospital or clinical use by medical professionals whenever it is required to monitor concentrations of anesthetic gases. The agents monitored include halothane, enflurane, isoflurane, sevoflurane and desflurane. Additionally, this monitor also monitors end-tidal carbon dioxide, nitrous oxide and oxygen gas concentrations. The need to monitor these anesthetic agents is most commonly encountered in the operating room during administration of anesthesia. This device is available for sale only upon the order of a physician or other related licensed medical professional.
## Technological Comparison to Predicate Device(s):
The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitor uses a polychromatic non-dispersive infrared light (NDIR) system to identify and measure the anesthetic gases and carbon dioxide. This technology is identical to the other cited predicate devices listed above.
## Summary of Performance Testing:
The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitor conforms with national and available international product safety standards for electrical, electromagnetic compatibility, and anesthetic agent monitoring.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 28 1998
Mr. Francis X. Casey Invivo Research, Inc. 12601 Research Parkway Orlando, FL 32826
Re : K974581 Millennia 3500 Series Patient Monitor with Anesthetic Agent Regulatory Class: II (two) Product Code: 74 DRT Dated: August 3, 1998 Received: August 4, 1998
Dear Mr. Casey:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Mr. Francis X. Casey
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labelinq requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Intended Uses
The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitoring is intended for general hospital or clinical use by medical professionals whenever it is required to monitor concentrations of anesthetic gases. The agents whonitored include halothane, enflurane, isoflurane, sevoflurane and desflurane. Additionally, this monitor also monitors end-tidal carbon dioxide, nitrous oxide and Additionally, this more and to monitor these anesthetic agents is most commonly encountered in the operating room during administration of anesthesia. This device is available for sale only upon the order of a physician or other related licensed medical professional.
Kramer
(Division Sign-Division of Care and Neurologic 510(k) Num
prescriptions use ✓
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.