K974533 · Gerard Medical, Inc. · LJT · May 7, 1998 · General Hospital
Device Facts
Record ID
K974533
Device Name
TRIMPORT
Applicant
Gerard Medical, Inc.
Product Code
LJT · General Hospital
Decision Date
May 7, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Our TrimPort Systems are indicated for use whenever patient therapy requires repeated intravascular injection or continuous infusion of fluids, medications, antibiotics, nutritionals and the withdrawal of venous blood samples.
Device Story
TrimPort is an implanted subcutaneous vascular access system; consists of a portal, radiopaque catheter, and locking catheter connector. Implanted into a large blood vessel to provide repeated access; avoids damage to small peripheral veins from frequent needle punctures. Operated by physicians or nurses; site located by palpating the portal rim; needle inserted through skin into portal for fluid delivery or blood withdrawal. System designed for two-way flow; includes locking mechanism to prevent catheter disconnection. Requires use of 10cc or larger syringes to prevent excessive pressure; limits pressure to 40 psi. Benefits include reduced patient discomfort and protection of venous integrity.
Clinical Evidence
Bench testing only. No clinical data provided. Testing focused on system integrity, including catheter locking mechanism performance and leak-tightness under extreme misuse conditions.
Technological Characteristics
Implantable subcutaneous vascular access system comprising a portal, radiopaque catheter, and locking connector. Designed for two-way flow. Mechanical design features a locking mechanism for catheter security. No electronic components, software, or energy sources.
Indications for Use
Indicated for patients requiring repeated intravascular access for fluid/medication delivery, antibiotic administration, nutritional support, or venous blood withdrawal.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
MiniPort Implantable Vascular Access System (K942623)
Submission Summary (Full Text)
{0}------------------------------------------------
7 1998 MAY
C. 1
Summary of Safety and Effectiveness Information
Gerard Medical, Inc. Company Name and Address: 6 City Depot Rd. P.O. Box 940 ー Charlton City, MA 01508
Company Telephone No.: (508) 248-1562
Richard Cayer, Jr. Company Contact Name:
November 29, 1997 Summary Date:
Proprietary Name: TrimPort
Implanted Vascular Access System Common or Usual Name:
Implanted Subcutaneous IV Catheter Classification Name:
Class: '
Legally Marketed Substantially Equivalent Devices:
MiniPort Implantable Vascular Access System (Re: K942623). Manufactured by Gerard Medical, Inc.
Device Description and Intended Usage:
The Gerard Medical, Inc. TrimPort Implantable Access System is precision engineered to provide repeated access to the vascular system for both parenteral delivery of fluids and the withdrawal of venous blood.
The implantable segments of the TrimPort System and comparable devices are the portal, the radiopaque catheter, and a locking catheter connector.
The small veins in one's arms and hands can become severely damaged when subjected to frequent needle punctures and/or irritating fluids. Usage of a TrimPort device, which is implanted beneath the skin, frees patients from the discomfort and potential damage to veins from repeated injections into them.
The TrimPort catheter is implanted into a large blood vessel so that fluids can be introduced or blood withdrawn without the necessity for repeated injections into the patient's vein.
{1}------------------------------------------------
Once the implanted TrimPort is ready for use, the physician or nurse will cleanse the skin at the implantation site and will pinpoint the device's location by simply feeling the unobtrusive but clear outline of its top rim. A needle specific to this purpose will be inserted gently through the skin into the portal of the TrimPort, where it will remain throughout the fluid delivery-withdrawal process.
Device Design, Testing, and Potential Adverse Occurrences:
We have reviewed a great quantity of information regarding our own and many competitive implantable vascular access systems. Based on our analysis, we feel that there are specific occurrences that would negatively affect the safety and effectiveness of these devices. The following is a list of these potential adverse occurrences and a summary of what we have done to minimize the risks.
- 1. Catheter Disconnection:
We designed an excellent catheter locking mechanism that remains intact, even under extreme conditions. To ensure that the surgeon implanting our TrimPort assembles the locking device properly, we provide clear and complete instructions and diagrams in our "Instructions for Use".
- 2. System Leak:
We precisely engineered our TrimPort vascular access systems so that they perform properly (Re: two-way flow, leak tight), even under extreme misuse conditions. To ensure this, we mandate that all new designs must be able to pass each of our standard usage and extreme misuse tests. To minimize the possibility of misuse, we instruct those who
access the device to use syringes that are 10cc or larger. This helps to prevent the emission of damaging pressure into Our recommendations to use 10cc or larger the system. syringes and not to exceed 40 psi when using the system is stated several times in our "Instructions for Use". Note that this is also a standard warning in the competitive instruction manuals we have reviewed.
3. System Occlusions:
We make numerous references, and provide much direction, in our "Instructions for Use" relative to flushing our device and clearing system occlusions.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1 1888 MAT
Mr. Richard Cayer, Jr. ·President Gerard Medical, Incorporated 6 City Depot Road P.O. Box 940 Charlton City, Massachusetts 01508
Re : K974533 TrimPort® Implantable Access System Trade Name: Regulatory Class: Unclassified Product Code: LJT Dated: April 23, 1998 April 27, 1998 Received:
Dear Mr. Cayer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the TrimPort® Implantable Access System have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਨੇ substantially equivalent determination assumes compliance with
{3}------------------------------------------------
## Page 2 - Mr. Cayer
the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such Failure to comply with the GMP regulation may assumptions. result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## RE: REVISED INDICATIONS FOR USE
1 Page of
K974533 510(k) Number (if known):
Device Name: TrimPort Implantable Access System
Indications For Use:
Our TrimPort Systems are indicated for use whenever patient therapy requires repeated intravascular injection or continuous infusion of fluids, medications, antibiotics, nutritionals and the withdrawal of venous blood samples.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Palacio Crescente
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K974533
Prescription Use ) (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Formal 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.