K974504 · E For M Imaging Systems Div., Marquette Electronic · LLZ · Feb 27, 1998 · Radiology
Device Facts
Record ID
K974504
Device Name
EPS
Applicant
E For M Imaging Systems Div., Marquette Electronic
Product Code
LLZ · Radiology
Decision Date
Feb 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The EPS system is a digital recording system designed for use as an electronic photospot device on OEM manufactured Radiographic & Fluoroscopic ("R/F") systems. The system is modular in configuration allowing single and dual room configurations for acquisition and/or review, providing assistance to Radiologic Technologists in routine diagnostic examinations and special procedures in R/F. The EPS system may be used to acquire, display, record, archive, and disseminate digital R/F image data.
Device Story
EPS is a 10-bit PC-based digital recording system for R/F imaging; functions as an electronic photospot device on OEM R/F systems. Inputs: X-ray image data from R/F systems. Processing: Simultaneous acquisition, review, post-processing, and archival of raw data. Outputs: Digital images displayed on monitors; archived to MOD media; disseminated via LAN. Used in clinical settings (single/dual room configurations) by Radiologic Technologists. Facilitates diagnostic examinations and special procedures; supports clinical review, reporting, and billing. Benefits: Enables digital capture, storage, and efficient management of R/F images.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
10-bit, 1024 digital acquisition/display system; PC-based architecture (Pentium processors); Windows NT OS; DICOM 3.0 compliant; modular configuration; MOD archival media; connectivity via LAN. Complies with UL 2601.1, UL 1950, and 21 CFR 1020.10/1040.10.
Indications for Use
Indicated for use by Radiologic Technologists to assist in routine diagnostic examinations and special procedures in Radiographic and Fluoroscopic (R/F) imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
The following summary is submitted pursuant to 21 CFR 807.92:
FES 27 1998
| [1] Submitter: | Marquette Medical Systems |
|-----------------------------|--------------------------------------------------------------------------------------|
| | E for M Imaging Systems Division |
| | 625 Alaska Avenue |
| | Torrance, CA 90503 USA |
| Date Prepared: | November 26, 1997 |
| [2] Trade/Proprietary Name: | EPS |
| Common/Usual Name: | R/F Digital Image Acquisition and Review System |
| Classification: | Class II per 21 CFR 892.1620 |
| Classification Name: | (Accessory to) Cine or Spot Fluorographic X-Ray Camera |
| Panel: | Radiology |
| Performance Standards: | 21 CFR 1020.10, Performance Standards for Ionizing Radiation Emitting |
| | Products; |
| | 21 CFR 1040.10, Performance Standards for Light-Emitting Products; |
| | DICOM 3.0 Standard for Digital Exchange Media (Voluntary); |
| | UL 2601.1, Standard for Safety, Medical Electrical Equipment, Part 1: General |
| | Requirements for Safety [Underwriters Laboratories] (Voluntary); |
| | UL 1950, Safety of Information Technology Equipment, Including Electrical Business |
| | Equipment [Underwriters Laboratories] (Voluntary); |
| | SMPTE (Society of Motion Picture and Television Engineers) Test Patterns (Voluntary) |
| Reason for 510(k): | New product equivalency. |
| [3] Predicate Device: | EPS-20, K922240, manufactured by Toshiba America Medical Systems, Inc. |
EPS-20, K922240, manufactured by Toshiba America Medical Systems, Inc. [3] Predicate Device:
#### [4] Description:
The Electronic PhotoSpot (EPS) is a 10 bit PC based multitasking acquisition/display system configured for use in R/F applications. This system is designed for use by a qualified Radiologic Technologist.
More specifically, the EPS system is a digital recording system designed for use as an electronic photospot device on OEM manufactured R/F systems. This system consists of independent and simultaneous acquisition of data, review, post processing of raw data and archive transfer functions. The system is modular in configuration allowing single and dual room configurations for acquisition and/or review. The system has 10 bit, 1024 digital acquisition, display and storage capability and uses a Windows based operator interface on the Operator Console/Review Station. It includes the following major components (minimally):
- Pentium computers a.
- b. Image monitor(s)
- Camera X-Ray Interface C.
- MOD media for archival d.
- Software program developed by Marquette Medical Systems to run in conjunction with Microsoft Windows NT e.
The EPS provides image acquisition, display, archival, and exchange media compliant with the DICOM 3.0 standard. Furthermore, the EPS allows connection to the medical facility's existing LAN network for data dissemination to support reviews, reports, and patient billing.
{1}------------------------------------------------
The following questions have been addressed:
| • Is the device life supporting or life sustaining? | No |
|------------------------------------------------------------------------|-------|
| • Is the device implanted (short term or long term)? | No |
| • Does the device design use software? | Yes |
| • Level of concern for software (if applicable)? | Minor |
| • Is the device sterile? | No |
| • Is the device for single use? | No |
| • Is the device for home use or prescription use? | No |
| • Does the device contain drugs or biological products as a component? | No |
| • Is the device a diagnostic kit? | No |
### [5] Intended Use:
The EPS system is a digital recording system designed for use as an electronic photospot device on OEM manufactured Radiographic & Fluoroscopic ("R/") systems. The system is modular in configuration allowing single and dual room configurations for acquisition and/or review, providing assistance to Radiologic Technologists in routine diagnostic examinations and special procedures in R/F. The EPS system may be used to acquire, display, record, archive, and disseminate digital R/F image data.
## [6] Comparison to Predicate Devices:
The Toshiba EPS-20 is a legally marketed predicate device with the same intended use as Marquette Medical Systems' EPS System. The Toshiba EPS-20 was successfully cleared through the FDA 's premarket notification process under file number K922240. A comparison of device specifications and principles of operation indicates no new questions of safety or efficacy, or substantial risk are raised.
#### Official Correspondent: Stu Bush
Signature: Ster Back 11/26/97
Telephone: Fax
(310) 320-8334 (310) 618-9031
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human figure with three faces in profile, stacked on top of each other. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 7 1998
Stu Bush Director, Regulatory Affairs & Quality Assurance Official Correspondent Marquette Medical Systems 625 Alaska Avenue Torrance, CA 90503-5124
Dear Mr. Bush:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Coometic Act (Act). You may, therefore, narket the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
Re:
K974504
Review System
Regulatory class: II
Dated: November 26, 1997
Received: December 1, 1997
EPS (R/F Digital Image Acquisition and
21 CFR 892.1650/Procode: 90 JAA/90 LLZ
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Feleval Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through persons, the Food and Drug Administration (FDA) will verify systems. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdr/dsmaln.html".
Sincerely yours,
Lillian Yin, Ph.D. Director, Division of Reproductiv Abdominal, Ear, Nose and Thros and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page__________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
### B. INTENDED USE
The EPS system is a digital recording system designed for use as an electronic photospot device on OEM manufactured Radiographic & Fluoroscopic ("R/F") systems. The system is modular in configuration allowing single and dual room configurations for acquisition and/or review, providing assistance to Radiologic Technologists in routine diagnostic examinations and special procedures in R/F.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K974504
V Prescription Use_ (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.