ACCU-CHEK A1C HEMOGLOBIN TEST

K974491 · Boehringer Mannheim Corp. · LCP · May 13, 1998 · Hematology

Device Facts

Record IDK974491
Device NameACCU-CHEK A1C HEMOGLOBIN TEST
ApplicantBoehringer Mannheim Corp.
Product CodeLCP · Hematology
Decision DateMay 13, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Indications for Use

The Accu-Chek A₁c Hemoglobin Test is designed for use with BMC Hitachi reagents that measure glycosylated hemoglobin. The test has not been evaluated using any other methodology. The device is intended for at-home or in-office use to collect a whole blood sample. The product will be marketed over-the-counter. Our proposed labeling contains instructions for patients that test results should be evaluated together with a personal doctor. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient’s diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient. This product is not indicated for the diagnosis of diabetes mellitus.

Device Story

Kit for collection of whole blood samples for Hemoglobin A1c testing; includes test strips, disposable lancets, instructions, return envelope, and patient ID card. Used at home or in-office by patients or clinicians. Patient collects blood sample on test strip; mails sample to laboratory for analysis using BMC Hitachi reagents. Results returned to patient/doctor for assessment of diabetes control and insulin dosage adjustment. Facilitates remote sample collection for laboratory processing; improves access to testing.

Clinical Evidence

Bench testing included reagent and sample stability studies under abusive environmental conditions. User studies (human factors) were conducted with lay persons to evaluate ease of use and kit acceptability. Results demonstrated that samples collected by untrained users correlate well with samples collected by healthcare professionals.

Technological Characteristics

Kit includes test strips, disposable lancets, and mailing materials. Device functions as a pre-analytical sample collection accessory for laboratory-based glycosylated hemoglobin assays. No electronic, software, or energy-based components.

Indications for Use

Indicated for patients with diabetes to assess glycemic control and assist in insulin dosage determination. Not indicated for the diagnosis of diabetes mellitus. Suitable for at-home or in-office use.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Submission Summary (Full Text)

{0} MAY 13 1998 K974491 # 510(k) Summary | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | --- | --- | | 1) Submitter name, address, contact | Boehringer Mannheim Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 845-2000 Contact Person: Mike Flis Date Prepared: November 14, 1997 | | 2) Device name | Proprietary name: Accu-Chek A_{1c}^{TM} Hemoglobin Test Common name: Hemoglobin A_{1c} Test Classification name: Glycosylated Hemoglobin Assay | | 3) Predicate device | The device is an accessory to our Tina Quant Hemoglobin A_{1c} Assay. We claim our own device as the predicate device (#k934070). In addition, a secondary predicate device may be considered. The role of the Accu-Chek A_{1c} Hemoglobin Test in the overall process is substantially equivalent to the intended use of the Flexsite Diagnostics EZCHEK/HBA Blood Collection Kit (#k971919). | | 4) Device Description | The device is a kit containing the materials required to collect a whole blood sample on a test strip and mail it to a laboratory for determination of Hemoglobin A_{1c}. The kit contains test strips, disposable lancets, instructions booklet, other related labeling, a return envelope, and a patient identification card for returning a result to the patient and/or his/her doctor. | Continued on next page {1} 510(k) Summary, Continued 5) Intended use The Accu-Chek A₁c Hemoglobin Test is designed for use with BMC Hitachi reagents that measure glycosylated hemoglobin. The test has not been evaluated using any other methodology. The device is intended for at-home or in-office use to collect a whole blood sample. The product will be marketed over-the-counter. Our proposed labeling contains instructions for patients that test results should be evaluated together with a personal doctor. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient’s diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient. This product is not indicated for the diagnosis of diabetes mellitus. 6) Comparison to predicate device Use of the Accu-Chek A₁c Hemoglobin Test with the Tina Quant assay only affects the pre-analytical stage of the testing process. All procedural steps and performance specifications associated with the analytical and post analytical stages of the assay are unaffected by the introduction of this accessory device. The Tina Quant Hemoglobin A₁c Test was evaluated for several performance characteristics, including precision and correlation. Additionally, the Accu-Chek A₁c Hemoglobin Test was evaluated for reagent and sample stability when exposed to abusive environmental conditions. Actual use situations (user studies) were also performed. Study participants filled out questionnaires regarding the ease of use. The evaluation studies provide evidence that results produced using samples collected by untrained lay persons using the accessory device correlate well with whole blood samples collected and handled directly by health care professionals following the procedure cleared in the original 510(k) file. Study participants indicated that the kit is acceptable as designed. {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service MAY 13 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mike Flis Regulatory Affairs Specialist Boehringer Mannheim Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, Indiana 46250-0457 Re: K974491 Accu-Chek A₁c™ Hemoglobin Test Regulatory Class: II Product Code: LCP Dated: March 27, 1998 Received: March 30, 1998 Dear Mr. Flis: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} 510(k) Number (if known): Device Name: Accu-Chek A₁c™ Hemoglobin Test Indications for Use: The Accu-Chek A₁c™ Hemoglobin Test is designed for use with BMC Hitachi reagents that measure glycosylated hemoglobin. The test has not been evaluated using any other methodology. The device is intended for at-home or in-office use to collect a whole blood sample. The product will be marketed over-the-counter. Test results should be evaluated together with a personal doctor. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient’s diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient. This product is not indicated for the diagnosis of diabetes mellitus. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Tecnica J. Caluia for A & M. Montgomery (Division of ODT) Division of Clinical Laboratory Devices 510(k) Number K974491 Prescription Use ☐ (Per 21 CFR 801.109) OR Over-The-Counter Use ☑ (Optional Format 1-2-96) 030
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...