K974452 · Beckman Instruments, Inc. · JJY · Dec 17, 1997 · Clinical Chemistry
Device Facts
Record ID
K974452
Device Name
VIGIL LIPID CONTROL
Applicant
Beckman Instruments, Inc.
Product Code
JJY · Clinical Chemistry
Decision Date
Dec 17, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
Beckman's Vigil Lipid Control is intended for use in monitoring the reliability of automated in vitro diagnostic assays of total cholesterol, HDL cholesterol (HDLc), triglycerides, apolipoprotein A-1 and apolipoprotein B in the clinical laboratory. The use of three or more levels of control allows the laboratorian to monitor change in calibration linearity along with analytical error and imprecision.
Device Story
VIGIL Lipid Controls are human serum-based liquid quality control materials used in clinical laboratories. The device consists of four levels of ready-to-use controls, each containing specific concentrations of cholesterol, HDL cholesterol, triglycerides, apolipoprotein A-1, and apolipoprotein B. Laboratory personnel use these controls to monitor the performance of automated diagnostic assays. By analyzing these controls, clinicians can detect systematic analytical deviations, assess imprecision, and verify calibration linearity of the test system. The controls are stored at -15°C to -20°C to maintain stability. The use of multiple levels allows for comprehensive monitoring of assay reliability, helping ensure accurate patient test results.
Clinical Evidence
Bench testing only. Stress stability studies were conducted at 25°C, 32°C, 37°C, and 41°C for durations ranging from 20 to 45 days to support a 24-month shelf life claim.
Technological Characteristics
Human serum-based liquid control material. Four levels of analyte concentration (cholesterol, HDL cholesterol, triglycerides, apolipoprotein A-1, apolipoprotein B). Storage temperature: -15°C to -20°C. Shelf life: 24 months. Class I medical device.
Indications for Use
Indicated for use in clinical laboratories to monitor the reliability, precision, and calibration linearity of automated in vitro diagnostic assays for total cholesterol, HDL cholesterol, triglycerides, apolipoprotein A-1, and apolipoprotein B.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}
Beckman Instruments, Inc., Section 510(k) Notification
VIGIL Lipid Controls
Summary of Safety & Effectiveness
K974452
DEC 17 1997
# Summary of Safety & Effectiveness
## Beckman VIGIL™ Lipid Controls
### 1.0 Submitted By:
Richard T. Ross
Staff Regulatory Specialist, Product Submissions
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-104
Brea, California 92822-8000
Telephone: (714) 961-4912
FAX: (714) 961-4541
### 2.0 Date Submitted:
21 November 1997
### 3.0 Device Name(s):
#### 3.1 Proprietary Names
VIGIL™ Lipid Control
#### 3.2 Classification Name
Quality Control Material (assayed and unassayed) (21 CFR § 862.1660)
### 4.0 Predicate Device(s):
| BECKMAN Reagent | Predicate | Predicate Company | Docket Number |
| --- | --- | --- | --- |
| VIGIL™ Lipid Controls | VIGIL PR_{x}^{TM} Controls | Beckman Instruments, Inc. | K936184 |
### 5.0 Description:
The VIGIL™ Lipid Controls are four level ready-to-use human serum-based liquid controls manufactured by Beckman Instruments, Inc. Each kit contains 4 X 4 mL bottles of a single level of control.
### 6.0 Intended Use:
Beckman's Vigil Lipid Control is intended for use in monitoring the reliability of automated in vitro diagnostic assays of total cholesterol, HDL cholesterol (HDLc), triglycerides, apolipoprotein A-1 and apolipoprotein B in the clinical laboratory. The use of three or more levels of control allows the laboratorian to monitor change in calibration linearity along with analytical error and imprecision.
file: Lipid_510k.sse
{1}
Beckman Instruments, Inc., Section 510(k) Notification
VIGIL Lipid Controls
Summary of Safety & Effectiveness
# 7.0 Comparison to Predicate(s):
The following tables show similarities and differences between the predicate identified in Section 4.0 of this summary.
## SIMILARITIES to the PREDICATE
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| Vigil Lipid Control | Intended use | Same as the predicate |
| | Value Assignment | Same process as the predicate |
| | Storage Temperature | Same as predicate at -15 °C to -20°C |
## DIFFERENCES from the PREDICATE
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| Vigil Lipid Control | Formulation | Vigil Lipid: stabilized by storage at -15°C to -20°C
Vigil PR_{s}: stabilized with ethylene glycol and storage at -15°C to -20°C |
| | Levels of analyte | Vigil Lipid: 4 levels
Vigil PR_{s}: 3 levels |
| | Analytes | Vigil Lipid: Contains: cholesterol, HDL cholesterol, triglycerides, apolipoprotein A-1 and apolipoprotein B fortified to attain various levels.
Vigil PR_{s}. Contains: apolipoprotein A-1, apolipoprotein B and endogenous cholesterol, HDL cholesterol and triglycerides. |
# 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence of the Vigil Lipid Controls to the Vigil PRs. Stress stability studies of the Vigil Lipid Controls support the Beckman stability claim of 24 months.
## Vigil Lipid Controls Shelf Life Stability Study Summary
| Stress Temperature | Duration of Incubation | Beckman Stability Claim* |
| --- | --- | --- |
| 25°C | 42 Days | 24 months |
| 32°C | 45 Days | 24 months |
| 37°C | 27 Days | 24 months |
| 41°C | 20 Days | 24 months |
*Expiration dating placed on the package based on date of manufacture
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
file: Lipid_510k.sse
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
DEC 17 1997
Richard T. Ross
Staff Regulatory Specialist
Beckman Instruments, Inc.
200 S. Kraemer Boulevard, M/S W-104
P.O. Box 8000
Brea, California 92822-8000
Re: K974452
VIGIL™ Lipid Controls
Regulatory Class: I
Product Code: JJY
Dated: November 21, 1997
Received: November 25, 1997
Dear Mr. Ross:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
K974452
page ___ of ___
510(k) Number (if known): Not yet assigned
Device Name: VIGIL™ Lipid Control
Indications for Use:
Beckman's Vigil Lipid Control is intended for use in monitoring the reliability of automated in vitro diagnostic assays of total cholesterol, HDL cholesterol (HDLc), triglycerides, apolipoprotein A-1 and apolipoprotein B in the clinical laboratory. The use of three or more levels of control allows the laboratorian to monitor change in calibration linearity along with analytical error and imprecision.
## 21 CFR 862.1660 Quality Control Material (assayed and unassayed)
(a) Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency studies in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
(b) Classification. Class I.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑
(per 21 CFR 801.109)
OR
(Division Sign-Off)
Division of Control Laboratory Devices
510(k) Number 12974452
Over-the-Counter Use
Optional Format 1-2-96
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.