← Product Code [DZE](/productcode/DZE) · K974280

# ABUTMENT - COMPLETE (K974280)

_The Implant Center of the Palm Beaches · DZE · Feb 20, 1998 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K974280

## Device Facts

- **Applicant:** The Implant Center of the Palm Beaches
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Feb 20, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The abutment submitted is for use in multiple unit, partial, or full edentulous case types. The underlying surface mates with the external hex of the dental implant, and the set screw down the middle locks it into place. Its purpose is to provide a gentle contain of more ing from the top of the implant into the oral cavity, ending with a metal margin or a cosmetic margin that can be established at the laboratory phase. The laboratory technician with this abutment has coping devices to place over the abutment or the abutment analog to fabricate the prosthetic design in the laboratory, which will then be brought to the patient for these clinical indications. It is not intended for single-tooth application, as there is no built-in, anti-rotational features. It is, therefore, recommended only as a multiple unit; partially edentulous; or a multiple unit, totally edentulous abutment design. It functions similarly to other designs in this implant system for similar indications. The components are made of biocompatible ASTM standard titanium/titanium alloy, and it is not considered an implant, as is an abutment piece placed above the top of the implants.

## Device Story

Dental abutment device; interfaces with external hex of dental implants via central set screw. Used by laboratory technicians to fabricate prosthetic designs for multiple-unit, partial, or full edentulous cases. Provides transition from implant top to oral cavity; supports metal or cosmetic margins. Not for single-tooth use. Clinical benefit: enables multi-unit prosthetic restoration in edentulous patients.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design and intended use.

## Technological Characteristics

Biocompatible titanium/titanium alloy per ASTM standards. Mechanical interface: external hex with central set screw. Non-rotational design. Sterile/non-sterile status not specified.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Jack T. Krauser, D.M.D. The Implant Center of The Palm Beaches 824 US Highway I, Suite 370 North Palm Beach, Florida 33408

FEB 20 1998

K974280 Re : Abutment - Complete Trade Name: Regulatory Class: III Product Code: DZE October 20, 1997 ... .. . Dated: November 5, 1997 Received:

## Dear Dr. Krauser:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set --forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does

{1}------------------------------------------------

Page 2 - Dr. Krauser

not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of ... . Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaggov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

{2}------------------------------------------------

1974280 of Page

| 510(k) Number (if known): | K97-4280 |
|---------------------------|----------|
|---------------------------|----------|

Abutment Device Name:

Indications For Use:

The abutment submitted is for use in multiple unit, partial, or full edentulous case types. The underlying surface mates with the external hex of the dental implant, and the set screw down the middle locks it into place. Its purpose is to provide a gentle contain of more ing from the top of the implant into the oral cavity, ending with a metal margin or a cosmetic margin that can be established at the laboratory phase. The laboratory technician with this abutment has coping devices to place over the abutment or the abutment analog to fabricate the prosthetic design in the laboratory, which will then be brought to the patient for these clinical indications. It is not intended for single-tooth application, as there is no built-in, anti-rotational features. It is, therefore, recommended only as a multiple unit; partially edentulous; or a multiple unit, totally edentulous abutment design. It functions similarly to other designs in this implant system for similar indications. The components are made of biocompatible ASTM standard titanium/titanium alloy, and it is not considered an implant, as is an abutment piece placed above the top of the implants.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Conci rrence of CDRH, Office of Device Evaluation (ODE)

2974280 - Suan Runas
(Division Sign-Off)

(Division Sign-Off, Division of De and General I 510(k) Numb

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K974280](https://fda-staging.innolitics.com/device/K974280)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).