K974241 · Scimed Life Systems, Inc. · DQX · Nov 12, 1998 · Cardiovascular
Device Facts
Record ID
K974241
Device Name
SCIMED INFORMER PRESSURE WIRE SYSTEM
Applicant
Scimed Life Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Nov 12, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The SCIMED INFORMER™ Pressure Wire System is intended to monitor the mean intravascular blood pressure and to facilitate the placement of balloon dilatation catheters or other therapeutic devices during PTCA or other intravascular interventional procedures. The phasic wave signal of the INFORMER Pressure Wire System is intended to facilitate placement of the wire. The INFORMER Pressure Wire System is not intended for use in the cerebral vasculature. The SCIMED INFORMER Pressure Wire System is not intended to be used in combination with any H.F. surgical equipment. The SCIMED INFORMER Pressure Wire System is indicated for use with all physiological monitors that meet IEC 601-1, IEC 601-2-34 standards. These monitors are required to meet electrical isolation for Type CF applied parts.
Device Story
System comprises steerable 0.014-inch coronary guide wire, disposable transducer with cable, and permanent cable. Wire monitors intravascular pressure; transducer converts hemodynamic pressure signals into electrical signals. System interfaces with standard physiological monitors to display phasic waveforms. Used by clinicians during PTCA or interventional procedures to facilitate balloon dilatation catheter placement. Provides real-time pressure data to assist navigation and clinical decision-making during vascular interventions.
Clinical Evidence
In-vitro bench testing included standard guide wire evaluations, ANSI/AAMI BP-22 transducer testing, and mechanical cable testing (tensile, cycling, strain relief). System response and shelf life were validated. In-vivo testing was conducted to compare pressure reading capabilities and guide wire performance against currently marketed devices.
Technological Characteristics
System components: 0.014-inch steerable guide wire, disposable transducer, permanent cable. Guide wire features radiopaque tips (3cm/5cm) and variable flexibility. Transducer converts hemodynamic pressure to electrical signals. Compatible with IEC 601-1 and IEC 601-2-34 compliant physiological monitors requiring Type CF electrical isolation.
Indications for Use
Indicated for patients undergoing PTCA or other intravascular interventional procedures requiring intravascular blood pressure monitoring and guide wire placement. Not for use in cerebral vasculature or with H.F. surgical equipment.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Cardiometrics WaveWire/WaveMap Pressure System (K965140)
Submission Summary (Full Text)
{0}------------------------------------------------
## NOV 1 2 1998
## Summary of Safety and Effectiveness
| General<br>Information | Submitter's Name and Address | SCIMED Life Systems, Inc.<br>One SCIMED Place<br>Maple Grove, MN 55311 |
|------------------------|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person | Jill Munsinger<br>(612) 494-2359 |
| | Classification Name(s) | Catheter Guide Wire, 21 CFR Part<br>870.1330; Transducer, Blood<br>Pressure, Extra Vascular, 21 CFR<br>Part 870.2850; Patient Transducer<br>and Electrode Cable, 21 CFR Part<br>870.2900 |
| | Common or Usual Name(s) | PTCA Guide Wire<br>Disposable Pressure Transducer<br>Cable, Patient Transducer |
| | Proprietary Name(s) | SCIMED® Informer™ Pressure<br>Wire System |
| | Product Code(s) | 74 DQX (Catheter Guide Wire)<br>74 DRS (Transducer, Blood<br>Pressure, Extra Vascular)<br>74 DRA (Patient Transducer and<br>Electrode Cable |
Predicate Devices
The SCIMED® Informer™ Pressure Wire is substantially equivalent to the products listed on the following page.
{1}------------------------------------------------
.
| Summary of Safety and Effectiveness, continued | |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The SCIMED INFORMER™ Pressure Wire System is intended to<br>monitor the mean intravascular blood pressure and to facilitate the<br>placement of balloon dilatation catheters or other therapeutic devices<br>during PTCA or other intravascular interventional procedures. The<br>phasic wave signal of the INFORMER Pressure Wire System is<br>intended to facilitate placement of the wire. The INFORMER Pressure<br>Wire System is not intended for use in the cerebral vasculature. |
| | The SCIMED INFORMER Pressure Wire System is not intended to be<br>used in combination with any H.F. surgical equipment. |
| | The SCIMED INFORMER Pressure Wire System is indicated for use<br>with all physiological monitors that meet IEC 601-1, IEC 601-2-34<br>standards. These monitors are required to meet electrical isolation for<br>Type CF applied parts. |
| Summary of<br>Technological<br>Characteristics | The Informer™ System is substantially equivalent to the currently<br>marketed SCIMED ChoICE™ and Sceptor PTCA Families of Guide<br>Wires, the Namic Perceptor™ DT Disposable Transducer, the MEDEX<br>TranStar Pressure Transducer, the BAXTER Edwards Permanent Cable<br>Model PX-1800 and the Cardiometrics WaveWire/WaveMap Pressure<br>System. |
| Non-Clinical<br>Test Summary | In-Vitro testing and evaluations were performed on the Informer™ Guide<br>Wire, the Informer™ Disposable Transducer with Attached Cable, the<br>Informer™ Permanent Cable and the complete Informer™ System.<br>Standard guide wire tests were used to evaluate the Informer™ Guide<br>Wire. The Informer™ Disposable Transducer experienced standard<br>ANSI/AAMI BP-22 testing while the cable testing included Cable<br>Connection Cycling, Disposable Transducer Cable Connector Tensile,<br>Transducer Housing/Cable Tensile, Strain Relief and Permanent Cable<br>Connector Tensile. System Response and Shelf Life testing was<br>conducted on complete Informer™ System units. |
| | Additionally, In-Vivo testing was conducted to compare pressure<br>reading capabilities and guide wire performance characteristics to<br>currently marketed devices. |
## Summary of Safety and Effectiveness, continued
{2}------------------------------------------------
| Product | 510(k) | Clearance Date |
|------------------------------------------------------|---------|-------------------|
| ChoICE Guide Wire | K943192 | November 22, 1994 |
| ChoICE Guide Wire | K945129 | March 3, 1995 |
| ChoICE Guide Wire | K950141 | March 3, 1995 |
| ChoICE Guide Wire | K950113 | March 31, 1995 |
| ChoICE Guide Wire | K961015 | May 15, 1996 |
| ChoICE Guide Wire | K965063 | March 4, 1997 |
| ChoICE Guide Wire | K964551 | May 21, 1997 |
| Sceptor Guide Wire | K942333 | August 31, 1994 |
| Sceptor Guide Wire | K946240 | March 31, 1995 |
| Sceptor Guide Wire | K950534 | April 28, 1995 |
| Sceptor Guide Wire | K960563 | April 29, 1996 |
| Namic Perceptor DT<br>Disposable Transducer | K910764 | April 22, 1991 |
| Medex TranStar<br>Pressure Transducer | K942377 | August 16, 1994 |
| Baxter Edwards<br>Reusable PX-1800 | K925638 | October 18, 1993 |
| Cardiometrics<br>WaveWire/WaveMap<br>Pressure System | K965140 | August 18, 1997 |
## Summary of Safety and Effectiveness, continued
Device Description
Predicate Devices, continued
> The SCIMED® Informer™ Pressure Wire System consists of the following components:
- · a guide wire,
- · a disposable transducer with attached cable, and
- · a permanent cable.
The guide wire is a steerable coronary guide wire available in a nominal diameter of 0.014 inches, overall lengths of 185 and 300 centimeters, radiopaque tip lengths of 3 and 5 centimeters, straight and J-tip configurations and one tip flexibility that falls between the ChoICE Floppy and Intermediate models, similar to the ChoICE Extra Support. The guide wire is designed to monitor intravascular pressures and to facilitate the placement of balloon dilatation catheters.
The disposable transducer with attached cable converts hemodynamic pressure signals from the guide wire into electrical signals. This transducer interfaces with standard physiologic monitors to display the electrical signals in phasic waveforms via the permanent cable.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, depicted in a minimalist, abstract design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 2 1998
Ms. Jill Munsinger Regulatory Affairs Associate SCIMED Life Systems, Inc. One Scimed Place Maple Grove, MN 55311-1566
Re: K974241 SCIMED® Informer™ Pressure Wire System Trade Name: Requlatory Class: II Product Code: DQX Dated: August 13, 1998 Received: August 14, 1998
Dear Ms. Munsinger:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation
{4}------------------------------------------------
Page 2 - Ms. Jill Munsinger
you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahar , Ph.D Director Division of Cardiovascular, Respiratory And Neurolgoical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| 510(k) Number | K974241 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SCIMED® Informer™ Pressure Wire System |
| Indications for Use | The SCIMED Informer™ Pressure Wire System is intended to monitor the mean intravascular blood pressure and to facilitate the placement of balloon dilatation catheters or other therapeutic devices during PTCA or other intravascular interventional procedures. The phasic wave signal of the INFORMER Pressure Wire System is intended to facilitate placement of the wire. The INFORMER Pressure Wire is not intended for use in the cerebral vasculature. The SCIMED INFORMER Pressure Wire System is not intended to be used in combination with any H.F. surgical equipment. The SCIMED INFORMER Pressure Wire System is indicated for use with all physiological monitors that meet IEC 601-1, IEC 601-2-34 standards. These monitors are required to meet electrical isolation for Type CF applied parts. |
fy for Gantf
(Division
Divis
and IVC.
510(k) Number K974241
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)Prescription Use (Per 21 CFR 801.109) OR
Over The Counter Use__________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.