← Product Code [LYY](/productcode/LYY) · K974218

# DERMACLEAN STERILE (PROTEIN LABELING CLAIM) EXAM GLOVES (K974218)

_Ansell Perry · LYY · Dec 18, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K974218

## Device Facts

- **Applicant:** Ansell Perry
- **Product Code:** [LYY](/productcode/LYY.md)
- **Decision Date:** Dec 18, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A disposable device intended for medical purpose that is worn on the examiners hand to prevent contamination between patient and examiner.

## Device Story

DermaClean Sterile is a disposable latex patient examination glove. It is worn by clinicians to provide a protective barrier, preventing cross-contamination between the patient and the examiner. The device is sterile and powder-free, featuring a protein label claim of 50 micrograms or less of total water-extractable protein per gram. It is intended for general medical use.

## Clinical Evidence

No clinical data. Bench testing only, including physical property testing per ASTM D 3578, freedom from holes per ASTM D 5151, and biocompatibility testing (primary skin irritation and guinea pig sensitization).

## Technological Characteristics

Latex examination glove; powder-free; sterile. Meets ASTM D 3578 for dimensions and physical properties. Meets ASTM D 5151 for freedom from holes. Protein content limited to ≤ 50 micrograms/gram. Biocompatibility: passes primary skin irritation and guinea pig sensitization tests.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## DEC 18 1997

DermaClean Sterile (Protein Labeling Claim) Ansell Perry 1875 Harsh Avenue SE Massillon, Ohio 44646 330-833-2811 Telephone: 330-833-6213 Fax:

## Checklist Section 21.0

K974218

- [1] 510 (k) Summary
- [2] Ansell Perry Inc. 1875 Harsh Avenue SE Massillon, Ohio 44646

330-833-2811 Telephone: Fax: 330-833-6213

Contact: James R. Chatterton Telephone: 330-833-2811 330-833-6213 Fax:

October 28, 1997

·

- DermaClean Sterile (Protein Labeling Claim) Trade Name: [3] Examination Gloves Common Name: Classification Name: Patient Examination Glove
- DermaClean Sterile (Protein Labeling Claim) examination gloves, meet all of the requirements of [4] ASTM D 3578.
- DermaClean Sterile (Protein Labeling Claim) examination gloves meet all the current [ર] specifications for ASTM D 3578 Rubber Examination Gloves.
- DermaClean Sterile (Protein Labeling Claim) examination gloves are disposable device intended [6] for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
- DermaClean Sterile (Protein Labeling Claim) examination gloves are summarized with the [7] following technological characteristics compared to ASTM or equivalent standards.

| Characteristics     | Standard          |
|---------------------|-------------------|
| Dimensions          | Meets ASTM D 3578 |
| Physical Properties | Meets ASTM D 3578 |

{1}------------------------------------------------

| DermaClean Sterile (Protein Labeling Claim)<br>Ansell Perry<br>1875 Harsh Avenue SE<br>Massillon, Ohio 44646<br>Telephone: 330-833-2811<br>Fax: 330-833-6213 |                                                                                                    |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Freedom from holes                                                                                                                                           | Meets ASTM D 3578<br>Meets ASTM D 5151                                                             |
| Powder-Free<br>Meets described test in Attachment VI                                                                                                         | Not more than 2 mg residue by mass.                                                                |
| Protein Label Claim                                                                                                                                          | This latex glove contains 50 micrograms or<br>less of total water extractable protein per<br>gram. |
| Biocompatability                                                                                                                                             |                                                                                                    |
| Primary Skin Irritation in Rabbits                                                                                                                           | Passes                                                                                             |
| Guinea Pig Sensitization                                                                                                                                     | Passes                                                                                             |

- The performance test data of the non clinical tests are the same as mentioned immediately above. [8]
- Clinical data is not needed for medical gloves or for most devices cleared by the 510(k) process. િંત્ર
- It is concluded that DermaClean Sterile (Protein Labeling Claim) examination gloves are as safe, [10] as effective, and perform as well as the glove performance standards referenced in Section 7 above and therefore meet:

ASTM listed standards, FDA hole requirements, and labeling claims for the product.

- This summary will include any other information reasonably deemed necessary by The FDA. [11]

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized lines forming its body and wings.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 18 1997

Mr. James R. Chatterton Vice President Regulatory Affairs/Technical Ansell Perry Incorporated 1875 Harsh Avenue, S.E. Massillon, Ohio 44646-7199

K974218 Re : Dermaclean Sterile Latex Examination Gloves Trade Name: with Protein Label Claim (50 micrograms or less of total water extractables protein per gram.) Regulatory Class: I Product Code: LYY October 28, 1997 Dated: November 10, 1997 Received:

Dear Mr. Chatterton:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does

{3}------------------------------------------------

Paqe 2 - Mr. Chatterton

not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Diredtor Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

Ansell Perry Inc. Applicant:

510(K) Number (if known): _ K974218

Latex Patient Examination Glove Sterile, Powder Free with Protein Label Claim (50 Device Name: micrograms or less of total water extractables protein per gram.

Indications For Usc.

A disposable device intended for medical purpose that is worn on the examiners hand to prevent contamination between patient and examiner.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

*Chin S. Lin*
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices,
510(k) Number *K974218*

Prescription Use
Per 21 CFR 801.109

OR

Over-The-Counter *V*
(Optional Format 1-2-96)

3

ANSWELL PARKS

---

**Source:** [https://fda-staging.innolitics.com/device/K974218](https://fda-staging.innolitics.com/device/K974218)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com