K974176 · Drager, Inc. · CBK · Jun 17, 1998 · Anesthesiology
Device Facts
Record ID
K974176
Device Name
BABYLOG 8000 PLUS
Applicant
Drager, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jun 17, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Babylog 8000 plus is a long term ventilator to be used in intensive care of premature and newborn babies and of infants up to 10 kg (22 lbs).
Device Story
Microprocessor-controlled intensive care ventilator for premature/newborn babies and infants up to 10 kg; operates on continuous-flow principle with time or volume control and pressure limitation. Inputs include air/O2 gas supply; device uses integrated air/O2 mixer. Outputs include mechanical ventilation support; monitoring of airway pressure, PEEP, tidal volume, minute volume, and lung mechanics (compliance, resistance). Used in clinical intensive care settings; operated by healthcare professionals. Optional analog/digital interface allows data transfer to patient monitors or computers. Device includes medicament nebulizer compatibility. Provides multiple ventilation modes (CMV, AIC, SIMV, PSV, CPAP, Volume Guarantee, VIVE). Safety features include continuous monitoring for failures and audible alarms for pressure, O2 concentration, minute volume, apnea, and technical errors. Benefits include precise respiratory support and monitoring for neonatal/infant patients.
Clinical Evidence
Bench testing only. Device performance compared against predicate devices (Babylog 8000 and Servo 300) regarding ventilation modes, monitoring parameters, and physical characteristics. No clinical trial data presented.
Technological Characteristics
Microprocessor-controlled intensive care ventilator; valve array (2x10 valves) for gas delivery; air/O2 mixer. Dimensions: 212x280x390 mm; weight: 14.5 kg. Connectivity: RS232 serial port and analog outputs. Power: 85-140 V, 47-63 Hz. Monitoring: peak/mean airway pressure, PEEP, tidal volume, minute volume, FiO2, compliance, resistance. Software-controlled ventilation modes.
Indications for Use
Indicated for intensive care ventilation of premature babies, newborns, and infants up to 10 kg (22 lbs).
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K974176" in a handwritten style on the top line. Below that is the word "Dräger" in a bold, sans-serif font. The word "Dräger" has two dots above the letter "a".
Datei: Sumstate.RTF Date: 10/08/1997 Author: Frank Clanzett
### 510(k) SUMMARY Summary of Safety and Effectiveness
### APPLICANTS NAME AND ADDRESS:
Dräger Inc. Critical Care Systems Ships Fleasent Valley Road 3136 QuARLY Road Ghantilly, VA 20154 Telford, PA 18969
APPLICANTS TELEPHONE NUMBER: (703)-817-0100
APPLICANTS FACSIMILE NUMBER: (703)-817-0101
APPLICANTS CONTACT PERSON: Harald Kneuer
Requlatory Affairs Manager
DATE THE SUMMARY WAS PREPARED: Oct. 8, 1997
DEVICE NAME: Trade Name: Classification Name:
Babylog 8000 plus Intensive Care Ventilator (per 21 CFR 868.5895)
LEGALLY MARKETED DEVICE TO WHICH DRAGER IS CLAIMING SUBSTANTIAL EQUIVALENCE:
Babylog 8000 - Manufactured by Dräger Medizintechnik GmbH. Lübeck. Germany and sold in the United States by Dräger Inc.
Servo 300 - Sold in the United States by Siemens
JUN 17 1998
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The "a" in "Dräger" has two dots above it, which is called an umlaut. The text is black and the background is white. The font is a bit blocky.
### Brief description of the Babylog 8000 plus
The Babylog 8000 plus is an enhanced version of its predecessor Babylog 8000. It is a long-term, microprocessor-controlled, intensive care ventilator for premature and newborn babies, as well as for infants weighing up to about 10 kg.
The Babylog 8000 plus ventilates according to the continuous-flow principle with either time control or volume control, depending on the ventilation mode, and pressure limitation. Oxygen is metered out by the integrated air/O2 mixer.
A variety of the device functions are monitored in operation to make sure that safety-relevant failures are detected and safety for the patient is maintained.
The analog and digital interface is optionally available for connecting the Babylog 8000 plus to a device like a patient monitor or computer so that measured values and settings can be transferred to such equipment.
The ventilator can also be combined with a medicament nebulizer.
### Indented Use of the Babylog 8000
The Babylog 8000 plus is a long term ventilator to be used in intensive care of premature and newborn babies and of infants up to 10 kg (22lbs).
The device provides the following ventilation modes:
CMV Continuous Mandatory Ventilation
AIC Assist Control Ventilation
SIMV Synchronized Intermittent Mandatory Ventilation
Pressure Support Ventilation (available option) PSV
CPAP Continuous Positive Airway Pressure
Ventilation mode extension:
VG Volume Guarantee (available option), Volume controlled ventilation
VIVE Variable Inspiratory, Variable Expiratory Flow
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The "a" in the word has an umlaut, which is a diacritical mark consisting of two dots placed over the letter. The word is in black and the background is white.
11.1 2.1 3.1 3.1 3. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1
### Basis for the S&E - Decision
ﺍﻟﻤﻮﺿﻮﻋﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴ
ﺔ ﺍﻟﺘﻲ ﺗﻘﺮﻳﺒﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ....... .
1
| Characteristics | Servo 300 | Babylog 8000 | Babylog 8000 plus | |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------|
| 510(k) number | K902859 | K926292 | under review | |
| Manufacturer | Siemens | Dräger | Dräger | |
| Classification | Intensive Care<br>Ventilator | Intensive Care<br>Ventilator | Intensive Care<br>Ventilator | |
| Control Principle | pressure-controlled<br>or volume-<br>controlled, time-<br>cycled or flow-<br>cycled;<br>pressure-limited | time-controlled, time-<br>cycled, pressure-<br>limited | time-controlled or<br>volume-controlled;<br>time-cycled or flow-<br>cycled;<br>pressure-limited | |
| Software controlled | yes | yes | yes | |
| Ventilation<br>modes: | | | | |
| • CPAP | yes | yes | yes | |
| • CMV | yes | yes | yes | |
| • Assist Control | yes | yes | yes | |
| • SIMV | yes | yes | yes | |
| • PSV | yes | no | substantially equivalent<br>to Servo 300 | |
| • VSV | yes | no | substantially equivalent<br>to Servo 300 | |
| • PRVC | yes | no | substantially equivalent<br>to Servo 300 | |
| • Manual<br>inspiration | no | yes | like Babylog 8000 | |
| Method of triggering | Flow or pressure | Volume | like Babylog 8000 | |
| Ventilator<br>settings: | | | | |
| • Tin | 10 to 80% of breath<br>cycle time | 0.1 to 2 s | like Babylog 8000 | |
| • Tex | via rate and Tin% | 0.2 to 30 s | like Babylog 8000 | |
| • CMV rate | 5 to 150 bpm | 2 to 150 bpm<br>(indirectly set via Tin<br>and Tex) | like Babylog 8000 | |
| • SIMV rate | 0.5 to 40 bpm | 2 to 150 bpm<br>(indirectly set via Tin<br>and Tex) | like Babylog 8000 | |
| Characteristics | Servo 300 | Babylog 8000 | Babylog 8000 plus | |
| I:E ratio | via rate and Tin% | 1:300 to 3:1 (indirectly<br>set via Tin and Tex) | like Babylog 8000 | |
| O2<br>concentration | 21 to 100 Vol%±3<br>Vol% | 21 to 100 Vol%±3<br>Vol% | like Babylog 8000 | |
| Pinsp | 10 to 120 cmH2O | 10 to 80 cmH2O | 5 to 80 cmH2O | |
| PEEP | 0 to 50 cmH2O | 0 to 25 cmH2O | like Babylog 8000 | |
| Insp. Flow | no | 1 to 30 L/min<br>continuous flow | like Babylog 8000 | |
| Tidal Volume | adult: 2 to 4000 mL<br>pediatric: 0 to 400<br>mL<br>neonate: 0 to 40 mL | no | 2 to 100 mL (when VG<br>option is activated) | |
| Minute Volume | 0.2 to 60 L/min | cannot be set; due to<br>the control principle<br>minute volume is result<br>of ventilator settings<br>and the characteristics<br>of the respiratory<br>system | like Babylog 8000 | |
| Trigger<br>sensitivity | 0 to -17 cmH2O, or<br>0.7 to 2 L/min<br>(adult);<br>0.3 to 1 L/min<br>(pediatric);<br>0.17 to 0.5 L/min<br>(neonate);<br>(effective in all<br>ventilation modes) | 1 to 10 (=0 to 3 mL) | like Babylog 8000 | |
| Trigger<br>response time | not specified | 40 to 60 ms at max<br>sensitivity | like Babylog 8000 | |
| Rise time | 0 to 10% of breath<br>cycle time | no | like Babylog 8000 | |
| Front panel controls | rotary dials and<br>push buttons | rotary dials and push<br>buttons;<br>screen menu | like Babylog 8000 | |
| Display technology | LED indicators;<br>LED array for<br>pressure bargraph | LCD or ELD graphics<br>screen;<br>LED indicators;<br>LED array for pressure<br>bargraph | like Babylog 8000 | |
| Audible alarm | yes | yes | yes | |
| Alarm sound level | adjustable | adjustable | like Babylog 8000 | |
| Pressure relief<br>valve | 120 cmH2O | 120 cmH2O | like Babylog 8000 | |
| Characteristics | Servo 300 | Babylog 8000 | Babylog 8000 plus | |
| <b>Alarms:</b> | | | | |
| Airway pressure high/low | yes | yes | yes | |
| O2 concentration low/high | yes | yes | yes | |
| Expired minute volume low/high | yes | yes | yes | |
| Apnea | yes | yes | yes | |
| Gas supply low/high | yes | yes | yes | |
| Breathing frequency low/high | no | no | yes (upper alarm limit only) | |
| Power failure | yes | yes | yes | |
| Battery supply voltage low/high | yes | NA | like Babylog 8000 | |
| O2 cell disconnect | yes | yes | yes | |
| flow sensor error | yes | yes | yes | |
| pressure transducer error | yes | yes | yes | |
| Ventilator inop (technical alarms) | yes | yes | yes | |
| <b>Monitoring Parameters</b> | | | | |
| peak airway pressure | yes | yes | yes | |
| mean airway pressure | yes | yes | yes | |
| PEEP | yes | yes | yes | |
| inspired tidal volume | yes | no | yes (not displayed) | |
| expired tidal volume | yes | yes | yes | |
| expired minute volume | yes | yes | yes | |
| spontaneous fraction of expired minute volume | no | yes | yes | |
| percent leakage | no | yes | yes | |
| | Characteristics | Servo 300 | Babylog 8000 | Babylog 8000 plus |
| • | breathing<br>frequency | yes | yes | yes |
| • | rate-volume-<br>ratio | no | no | yes |
| • | compliance | no | no | yes |
| • | resistance | no | no | yes |
| • | FiO2 | yes | yes | yes |
| • | supply<br>pressures | yes | yes | yes |
| Physical<br>Characteristics: | | | | |
| • | dimensions<br>WxDxH | 242x370x240 mm<br>(patient unit)<br>431x150x325 mm<br>(control unit) | 212x280x390 mm | like Babylog 8000 |
| • | weight | 24 kg | 14.5 kg | like Babylog 8000 |
| Gas Supply | | | | |
| • | Supply gas<br>pressure, O2<br>and air | 29 to 94 PSI | 45 to 90 PSI | like Babylog 8000 |
| • | Gas delivery<br>system | microprocessor-<br>controlled valves | microprocessor-<br>controlled valve array<br>(2x10 valves) | like Babylog 8000 |
| Power supply | | | | |
| • | Mains voltage<br>range | 100, 120, 220 and<br>240 V, 50-60 Hz | 85 to 140 V;<br>47 to 63 Hz | like Babylog 8000 |
| • | Power<br>consumption | 50 W | 140 W | like Babylog 8000 |
| • | internal battery | yes | no | like Babylog 8000 |
| • | Operating time<br>with internal<br>battery | 30 min | NA | like Babylog 8000 |
| Special functions | | | | |
| • | Communications<br>interface | serial port RS232 +<br>analog output;<br>Master/Slave<br>connection | serial port RS232 +<br>analog outputs<br>(optional) | like Babylog 8000 |
| • | Medicament<br>nebulizer | yes | yes | yes |
{3}------------------------------------------------
## Dräger
and and the collection of the comments of the world
action maris as askais and the market for
・・・
:
.
{4}------------------------------------------------
### Dräger
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
i
{5}------------------------------------------------
# Dräger
a ... .. .. .. .. .. .. ..
ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ
Device performances compaired by Dr. Klaus Freudenstein / project manager
{6}------------------------------------------------
And Control Controller Comments of Children
#### Explanations of the Comparison
The ventilation modes of the Babylog 8000 plus that are not included in Babylog 8000 with the software 3.0 are substantial equivalent to ventilation modes of the Servo 300 ventilator of Siemens.
The additional monitoring features for the lung mechanics measurement like compliance, resistance and rate volume ratio may assist the user and are not essential for the safety and effectiveness concerning the intended use of the Babylog 8000 plus.
### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICES:
Dräger's Babylog 8000 plus is substantial equivalent to either the Dräger Babylog 8000 or to the Siemens Servo 300 ventilator.
The Babylog 8000 plus functions are also covered by the predicate devices and the intended use of the Babylog 8000 plus is covered by the predicate devices, too.
The Babylog 8000 plus fulfills at least the same technical standards as the predicate device of Dräger and has been tested according to these standards.
Therefore the device is as safe and effective as the predicate devices.
Suuuuuu
Oct. 8, 1997
Frank Clanzett Quality Management. Regulatory Affairs Dräger Medizintechnik GmbH, Lübeck, Germany
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing its wings, positioned to the right of a circular seal. The seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the circumference.
Public Health Service
JUN 17 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Harald Kneuer Dräger, Inc. Critical Care System 3136 Quarry Road Telford, PA 18969
K974176 Re: BABYLOG 8000 Plus Regulatory Class: II (two) Product Code: 73 CBK March 1, 1998 Dated: Received: March 19, 1998
Dear Mr. Kneuer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{8}------------------------------------------------
#### Page 2 - Mr. Harald Kneuer
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
The Babylog 8000 plus is a long term ventilator to be used in intensive care of premature and newborn babies and of infants up to 10 kg (22 lbs).
The device provides the following ventilation modes:
CMV Continuous Mandatory Ventilation Time-controlled, time-cycled, pressure-limited continuous flow ventilation.
- AIC Assist Control Ventilation Time-controlled, volume triggered, time-cycled, pressure-limited continuous flow ventilation that is synchronized with each spontaneous patient breath.
SIMV Synchronized Intermittent Mandatory Ventilation Time-controlled , volume triggered, flow cycled, pressure-limited continuous flow ventilation, synchronized with patient's spontaneous breathing at the set ventilation rate.
- Pressure Support Ventilation (available option) PSV Time-controlled , volume triggered, flow cycled, pressure-limited ventilation synchronized with each spontaneous patient breath.
- CAP Continuous Positive Airway Pressure
Spontaneous breathing with positive airway pressure
Ventilation mode extension:
- Volume Guarantee (available option), Volume controlled ventilation. VG The ventilator controls inspiratory pressure in order to deliver the preset tidal volume. May be combined with AJC, SIMV and PSV.
VIVE Variable Inspiratory. Variable Expiratory Flow Separate continuous flow during expiration in mandatory ventilation modes. ﺗﺮﺗ
Klaus Freudenstein
Dr. Klaus Freudenstein (project manager)
M. Puy
PESCRIPTION USE FOR
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.