INCARE PELVIC FLOOR THERAPY SYSTEM WITH DESKTOP COMPUTER
K974048 · Hollister, Inc. · KPI · Dec 23, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K974048
Device Name
INCARE PELVIC FLOOR THERAPY SYSTEM WITH DESKTOP COMPUTER
Applicant
Hollister, Inc.
Product Code
KPI · Gastroenterology, Urology
Decision Date
Dec 23, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The biofeedback components of the InCare Pelvic Floor Therapy System with Desktop Computer are intended to provide electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence. The electrical stimulation component of the InCare Pelvic Floor Therapy System with Desktop Computer provides stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.
Device Story
Office-based system for pelvic floor rehabilitation; used by physicians, nurses, and physiotherapists in clinical settings. Inputs include electromyographic (EMG) signals or pressure measurements from pelvic musculature via probes. System processes signals using Root Mean Squared (RMS) calculations; provides electrical stimulation (biphasic, symmetrical, balanced waveforms) to pelvic muscles. Output displayed on desktop computer for clinician monitoring and patient biofeedback. Device facilitates neuromuscular control restoration; treats urinary incontinence. Clinician controls stimulation parameters (frequency, pulse width, duty cycle) and session duration. Subject device adds 200 Hz stimulation frequency to predicate capabilities.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and bench testing (compliance with IEC 601-1, IEC 601-1-1, IEC 601-2-10, EN 60601-1-2/IEC 601-1-2, CISPR 11, and IEC 801 series).
Indicated for patients with urinary incontinence requiring rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control via electromyographic/pressure biofeedback or electrical stimulation.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
Predicate Devices
Hollister PRS9300 Pelvic Floor Therapy System (K961872)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Hollister logo. The logo consists of a geometric symbol resembling a stylized cross or asterisk, followed by the word "Hollister" in a serif typeface. The symbol is composed of intersecting bars, creating a symmetrical design with a central void. The text is positioned to the right of the symbol, with the letters evenly spaced and clearly legible.
DFC 2 3 1997
Hollister Incorporated 2000 Hollister Drive Libertyville, Illinois 60048-3781
P1073
# InCare Pelvic Floor Therapy System with Desktop Computer
### Safety and Effectiveness Summary
### 1. Submitter's name, Address and Contact Person
Submitter Hollister Incorporated 2000 Hollister Drive Libertyville, IL 60048
Contact Person Joseph S. Tokarz Manager, Regulatory Affairs (847)680-2849 Ph (847)918-3860 Fax -
Date Summary Prepared - October 21, 1997
#### 2. Name of Device:
InCare Pelvic Floor Therapy System with Desktop Computer
## 3. Name of Predicate Device(s)
Hollister PRS9300 Pelvic Floor Therapy System, K961872
## 4. Description of Device
The InCare Pelvic Floor Therapy System with Desktop Computer is an office based instrument that is intended to be used by physicians, nurses, nurse clinicians, and physiotherapists in a physicians office, clinic, or hospital for the purpose of providing electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence. The InCare Pelvic Floor Therapy System also provides electrical stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.
The premarket notification is being submitted to the Agency to include the electrical stimulation parameter of 200 Hz to the existing InCare Pelvic Floor Therapy System with Desktop Computer, K961872.
### 5. Statement of Intended Use
The biofeedback components of the InCare Pelvic Floor Therapy System with Desktop Computer are intended to provide electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence.
The electrical stimulation component of the InCare Pelvic Floor Therapy System with Desktop Computer provides stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.
{1}------------------------------------------------
K974048
P293
Image /page/1/Picture/1 description: The image shows a black geometric symbol. The symbol is composed of four thick, rectangular bars arranged in a cross-like formation. The bars intersect at their midpoints, creating a square-shaped void in the center of the symbol. The overall design is symmetrical and bold, with a small imperfection visible on the upper left corner of the symbol.
# Hollister.
# InCare Pelvic Floor Therapy System with Desktop Computer
# 6. Statement of Technological Characteristics of the Device
· The proposed device and the predicate device, K961872 are substantially equivalent based upon the following:
- 1. Intended Use of the two devices are identical.
2. The proposed device and the predicate device, K961872 utilizes the exact same instrumentation unit. The Instrumentation Unit controls features and functions that are "patient treatment actions."
3. The proposed device and the predicate device, K961872 utilizes the same software for the "data manipulation and presentation" activities associated with the personal computer and the "patient therapy" actions associated with the Instrumentation Unit.
4. The proposed device differs from the predicate device, K961872 in one aspect only. The proposed device has the additional stimulation parameter of 200 Hz.
| | Proposed Device | Predicate Device K961872 |
|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Stimulation Characteristics | | |
| Intended Use | Treatment of Urinary Incontinence | Treatment of Urinary Incontinence |
| Computer Type | Desktop Computer | Desktop Computer |
| Output (nominal) | 0-30 VDC | 0-30 VDC |
| Waveform | Square, Symmetrical, Balanced,<br>Biphasic | Square, Symmetrical, Balanced,<br>Biphasic |
| charge/pulse at 500Ω | 60 µC/phase; net charge/pulse = 0 | 60 µC/phase; net charge/pulse = 0 |
| Frequency | 12.5, 20, 50, 100, 200 Hz | 12.5, 20, 50, 100 Hz |
| Peak Pulse Intensity | 30 VDC | 30 VDC |
| Pulse width | 0.3, 1 ms | 0.3, 1 ms |
| Ramps | 20%, 40%, 60%, 80%, 100%<br>of “ON” time (no down ramping) | 20%, 40%, 60%, 80%, 100%<br>of “ON” time (no down ramping) |
| Duty Cycle | On (sec): 1-80 in 1 sec increments<br>Off (sec): 0 - 80 in 1 sec<br>increments | On (sec): 1-80 in 1 sec increments<br>Off (sec): 0 - 80 in 1 sec<br>increments |
| Session Duration (min) | 0-30, 1 minute increments | 0-30, 1 minute increments |
| Programmable Features | NONE BY PATIENT | NONE BY PATIENT |
| | By physician: Pulse width, Freq-<br>uency, Duty cycle, Session length | By physician: Pulse width, Freq-<br>uency, Duty cycle, Session length |
| Current Density<br>Conditions: Full output setting, 100Hz,<br>Ims pulse width at 500 ohms (nominal) | Probe 9595 - 0.003 amperes/cm²<br>Probe 9596 - 0.018 amperes/cm² | Probe 9595 - 0.003 amperes/cm²<br>Probe 9596 - 0.018 amperes/cm² |
· A chart showing the differences and similarities of the InCare Pelvic Floor Therapy System with Desktop Computer and the predicate device follows:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Hollister brand logo. On the left is a black geometric design that resembles a stylized cross or snowflake. To the right of the design is the word "Hollister" in a serif font, with a trademark symbol next to it.
K974048
f3093
# InCare Pelvic Floor Therapy System with Desktop Computer
| | Proposed Device | Predicate Device K961872 |
|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Power Density<br>Conditions: Full output setting, 100Hz,<br>1ms pulse width at 500 ohms (nominal) | Probe 9595 - 0.047 watts/cm²<br>Probe 9596 - 0.239 watts/cm² | Probe 9595 - 0.047 watts/cm²<br>Probe 9596 - 0.239 watts/cm² |
| EMI | **See Note below | **See Note below |
| Biofeedback Characteristics | | |
| Intended Use | Treatment of Urinary Incontinence | Treatment of Urinary Incontinence |
| Measurement Channels | Adjustable, 2 channels EMG, 2<br>channels pressure, or combination<br>EMG and pressure | Adjustable, 2 channels EMG, 2<br>channels pressure, or combination<br>EMG and pressure |
| EMG Sensitivity (microvolts) | 0-5, 0-10, 0-25, 0-50, 0-100, 0-<br>250, 0-500 | 0-5, 0-10, 0-25, 0-50, 0-100, 0-<br>250, 0-500 |
| EMG Bandwidth | 100-500 Hz | 100-500 Hz |
| EMG Signal Processing | Root Mean Squared (RMS) | Root Mean Squared (RMS) |
| EMG Detection | Bipolar | Bipolar |
| Pressure Sensitivity (cm-H2O) | 0-10, 0-25, 0-50, 0-100 | 0-10, 0-25, 0-50, 0-100 |
| Work Period (sec) | 1-80 in 1 second increments | 1-80 in 1 second increments |
| Rest Period (sec) | 0-80 in 1 second increments | 0-80 in 1 second increments |
| Session Duration (min) | 1-60 in 1 minute increments | 1-60 in 1 minute increments |
**NOTE
IEC 601-1 (except Amendment 2:1995, new subclause 56.3c) IEC 601-1-1 IEC 601-2-10 (except Paragraph 36) EN 60601-1-2/IEC 601-1-2 EN 55011/CISPR 11 EN60801-2/IEC 801-2, 3kv contact, 8kv air IEC 801-3, 3v/m, 1 khz IEC 801-4, 1kv IEC 801-5, 1kv
## 7. Conclusion
Based upon the information presented above it is concluded that the proposed InCare Pelvic Floor Therapy System with Desktop Computer is safe and effective for its intended use and is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a bird. The emblem is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The text is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 3 1997
Mr. Joseph S. Tokarz Manager, Regulatory Affairs Hollister Incorporated 2000 Hollister Drive Libertyville, IL 60048
Re: K974048
InCare Pelvic Floor Therapy System with Desktop Computer Dated: October 21, 1997 Received: October 24, 1997 Regulatory class: II 21 CFR §876.5320/Product code 78 KPI 21 CFR §884.1425/Product code 85 HIR
Dear Mr. Tokarz:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# InCare Pelvic Floor Therapy System with Desktop Computer
b.
Statement of Intended Use
| 510(k) Number (if Known): | |
|---------------------------|----------------------------------------------------------|
| Device Name: | InCare Pelvic Floor Therapy System with Desktop Computer |
## Indications For Use:
The biofeedback components of the InCare Pelvic Floor Therapy System with Desktop Computer are intended to provide electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence.
The electrical stimulation component of the InCare Pelvic Floor Therapy System with Desktop Computer provides stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.
-
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | V |
|----------------------|---|
| (Per 21 CFR 801.109) | |
Over-
Over-the-Counter-Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
| (Division Sign-Off) | |
|-------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, | |
| and Radiological Devices | |
| 510(k) Number | K974048 |
OR
Page 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.