K973960 · Victoreen, Inc. · LHN · Jan 14, 1998 · Radiology
Device Facts
Record ID
K973960
Device Name
MODEL 37-185
Applicant
Victoreen, Inc.
Product Code
LHN · Radiology
Decision Date
Jan 14, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The Model 37-185 provides a means of scanning the light field and radiation field generated by radiation therapy machines and presents this data for quality control purposes only. Sequential scans of the light field and the radiation field provides information as to whether the cross hairs in the light field are in the center of the radiation field (per TG-40), the coincidence of the light field to the radiation field, and beam flatness and symmetry. These data aid in the determination of the therapy machine's performance characteristics. The Model 37-185 is intended for use by persons responsible for the proper interpretation of its readings and observing the appropriate safety procedures in the presence of radiation. As a quality control tool the use of this product does not involve patient contact nor is it used to evaluate patient treatment planning. The patient is not in the room while this product is in use.
Device Story
Model 37-185 is a radiation therapy quality control system consisting of a readout module and a scanner module. The scanner, placed on a treatment couch, uses a stepper motor to move a light-detecting diode and a radiation-detecting diode in a 50 cm straight line to profile radiation and light fields. A fixed central radiation diode acts as a reference to nullify beam amplitude fluctuations. The readout module, located outside the treatment room, contains a microprocessor, LCD, and printer/computer interfaces. It processes data to determine beam flatness, symmetry, and field coincidence per TG-40 guidelines. Operated by qualified personnel, the device provides performance data for therapy machines. It does not contact patients and is not used for treatment planning; the patient is absent during operation.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
System comprises a scanner module with stepper motor, light-detecting diode, and radiation-detecting diode, connected via 25-meter shielded cable to a microprocessor-based readout module with LCD. Includes leveling mechanism for 90-degree rotation. Connectivity includes RS-232 and parallel printer interface.
Indications for Use
Indicated for use by qualified personnel to perform quality control measurements on radiation therapy machines, including light field/radiation field coincidence, cross-hair centering, and beam flatness/symmetry. Not for patient contact or treatment planning.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Beam Scan, Model 37-201 (K853671A)
Submission Summary (Full Text)
{0}------------------------------------------------
EXHIBIT A
# Victoreen, Inc.
K973960
Image /page/0/Picture/3 description: The image shows a logo with three horizontal bars. The top bar is black, the middle bar is white, and the bottom bar is white with the word "VICTOREEN" in black, bold, sans-serif font. The logo is simple and clean, with a focus on the text.
Premarket Notification [510k] Summary as required by section 807.92(c)
IAN I 4 1998
# Date Summary was prepared:
September 30, 1997
#### Submitter's Name
Victoreen, Inc. 6000 Cochran Road Cleveland, Ohio 44139-3395
#### Contact Person:
Linda S. Nash Director of Regulatory Affairs and Quality Assurance Phone: 440-248-9300 440-248-9301 Fax:
## Device Name:
Model 37-185
## Classification Name:
Medical charged - particle radiation therapy system
## Predicate Device:
Beam Scan, Model 37-201, 510(k) #K853671A
# Product Description:
The product consists of two main parts; the readout module, and the scanner (data acquisition) module. The readout module is located outside the treatment room while the scanner is positioned on the treatment couch in such a manner as to allow the detectors to transverse the radiation beam as determined by the operator. The readout and the scanner are connected via a single 25 meter (75 foot) shielded cable.
The readout contains a LCD (Liquid Crystal Display) and 8 softkeys providing the operator interface. The readout also includes the microprocessor system, memory, printer interface and computer interface. Connections are provided for the power supply, parallel printer, computer RS-232 interface and the scanner cable.
> 6000 Cochran Road Cleveland, Ohio 44139-3395 (216) 248-9300 FAX (216) 248-9301 DATA (216) 248-9043
{1}------------------------------------------------
The scanner module contains a stepper motor and a transport mechanism that moves a light detecting diode and a radiation detecting diode 50 cm, in a straight line to obtain a beam or light field profile during a measurement cycle. There is a fixed radiation detecting diode located in the center of the scanner which acts as a reference to null out any fluctuations in the amplitude of the radiation beam (as a function of time). The stepper motor controller and detector pre-amplifiers are also housed in the scanner assembly. The product includes a separate leveling mechanism to level the scanner module and rotate it 90 degrees.
#### Intended Use:
The Model 37-185 provides a means of scanning the light field and radiation (Function): field generated by radiation therapy machines and presents this data for quality control purposes only. Sequential scans of the light field and the radiation field provides information as to whether the cross hairs in the light field are in the center of the radiation field (per TG-40), the coincidence of the light field to the radiation field, and beam flatness and symmetry. These data aid in the determination of the therapy machine's performance characteristics. The Model 37-185 is intended for use by persons responsible for the proper interpretation of its readings and observing the appropriate safety procedures in the presence of radiation.
As a quality control tool the use of this product does not involve patient contact nor is it used to evaluate patient treatment planning. The patient is not in the room while this product is in use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, possibly representing people or services provided by the department.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Linda S.Nash Director of Regulatory Affairs & Quality Assurance Victoreen, Inc. 6000 Cochran Road Cleveland, OH 44139-3395
Dear Ms. Nash:
Re: K973960
Beam Scan Model 37-185 Dated: October 14, 1997 Received: October 16, 1997 Regulatory class: II 21 CFR 892.5050/Procode: 90 LHN 1481 / 1000
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the ise stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitry diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsmamain.html", ............................................................................................
Sincerely yours,
h. Tiau Yu.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
510(k) Number: K973960
Device Name: Model 37-185
Indications for Use:
Model 37-185 is used to provide a means of scanning the light field and radiation field generated by radiation therapy machines and presents this data for quality control purposes only. Sequential scans of the light field and the radiation field provides information as to whether the cross hairs in the light field are in the center of the radiation field (per TG-40), the coincidence of the light field to the radiation field, and beam flatness and symmetry. These data aid in the determination of the therapy machine's performance characteristics.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David C. Leyman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
Or
Over-The Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.