← Product Code [GCJ](/productcode/GCJ) · K973943

# INSTRUMENTS FOR ENDOSCOPIC SUBFASCIAL HARVESTING OF VEINS (ESHV), INSTRUMENTS FOR ENDOSCOPIC SUBFASCIAL DISCISION OF PER (K973943)

_Richard Wolf Medical Instruments Corp. · GCJ · Aug 27, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K973943

## Device Facts

- **Applicant:** Richard Wolf Medical Instruments Corp.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Aug 27, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The ESHV (Endoscopic Subfascial Harvesting of Veins) instrument set is used for minimally invasive preparation and harvesting of the veins for use in cardiac and vascular surgical procedures. The intervention takes place under sterile conditions. The Veins dissector (stripper) is used for the blunt preparation of tubular structures, specifically for the free-preparation of vessels, e.g. the vein saphena magma. The ESDP (Endoscopic Subfascial Discision of Perforating Veins) instrument set is used for minimally invasive visualization, diagnostic, and therapeutic procedures of subfascial veins in the lower extremities, e.g. the saphenous vein. The intervention takes place under sterile conditions.

## Device Story

Instruments for Endoscopic Subfascial Harvesting of Veins (ESHV) and Endoscopic Subfascial Discision of Perforating Veins (ESDP) consist of operating telescopes, endoscopic operating tubes, laser guides, hook knives, grasping forceps, scissors, and vein dissectors (strippers). Used in sterile surgical environments by physicians for minimally invasive vein preparation/harvesting for cardiac/vascular procedures or subfascial vein treatment in lower extremities. Auxiliary instruments (3.5-5.5 mm diameter) are inserted through the operation channel. Devices provide visualization and mechanical access to vessels; facilitate blunt preparation and dissection. Benefits include minimally invasive approach to vessel harvesting and subfascial vein procedures.

## Clinical Evidence

No clinical tests performed. Bench testing only.

## Technological Characteristics

Manual endoscopic surgical instruments including telescopes, tubes, laser guides, knives, forceps, scissors, and strippers. Auxiliary instruments 3.5-5.5 mm diameter. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Instruments for Endoscopic Subfascial Discision of Perforating Veins "ESDP", Model 8781 ([K964258](/device/K964258.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# AUG 27 1998

[[][[[[][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][][

전

:

ー

Image /page/0/Picture/1 description: The image shows a handwritten number sequence, "16973943". The numbers are written in a dark ink, and the background is white. The numbers are slightly smudged, but they are still legible. There are some other markings on the left side of the image.

# RICH ARD INSTRUMENTS CORPORATION

Image /page/0/Picture/3 description: The image shows a black and white square with a stylized symbol inside. The symbol appears to be a combination of lines and shapes, possibly representing a character or logo. The image quality is somewhat degraded, with some pixelation and blurring around the edges of the square and the symbol.

F

# 510(k) Summary of Safety and Effectiveness

| Submitter:                                                                 |                                                                                                                                  |                 |                                                    | Date of Preparation:<br>October 10, 1997            |  |  |
|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------|----------------------------------------------------|-----------------------------------------------------|--|--|
| Company / Institution name:<br>Richard Wolf Medical Instruments Corp.      |                                                                                                                                  |                 |                                                    | FDA establishment regulation number:<br>14 184 79   |  |  |
| Division name (if applicable):<br>N.A.                                     |                                                                                                                                  |                 |                                                    | Phone number (include area code):<br>(847) 913-1113 |  |  |
| Street address:                                                            | 353 Corporate Woods Parkway                                                                                                      |                 |                                                    | FAX number (include area code):<br>(847) 913-0924   |  |  |
| City:<br>Vernon Hills                                                      | State/Province:<br>Illinois                                                                                                      | Country:<br>USA |                                                    | ZIP/Postal Code:<br>60061                           |  |  |
| Contact name:<br>Mr. Robert L. Casarsa                                     |                                                                                                                                  |                 |                                                    |                                                     |  |  |
| Contact title:                                                             | Quality Assurance Manager                                                                                                        |                 |                                                    |                                                     |  |  |
| Product Information:<br>Trade name:                                        | Instruments for Endoscopic Subfascial Harvesting of Veins (ESHV) and Endoscopic Subfascial Discision of Perforating Veins (ESDP) |                 |                                                    | Model number:<br>8781, 8393, 8394 and others        |  |  |
| Common name:<br>Operating Tubes. Operating Telescope<br>Laser Guide. Knife |                                                                                                                                  |                 | Classification Name:<br>Endoscopes and Accessories |                                                     |  |  |
| Information on devices to which substantial equivalence is claimed:        |                                                                                                                                  |                 |                                                    |                                                     |  |  |
| 510(k) Number                                                              | Trade or proprietary or model name                                                                                               | Manufacturer    |                                                    |                                                     |  |  |
| 1 K964258                                                                  | 1 Instruments for Endoscopic Subfascial Discision of Perforating Veins "ESDP", Model 8781                                        | 1 Richard Wolf  |                                                    |                                                     |  |  |
| 2                                                                          | 2 See section 3, Equivalent Devices                                                                                              | 2               |                                                    |                                                     |  |  |
| 3                                                                          | 3                                                                                                                                | 3               |                                                    |                                                     |  |  |
| 4                                                                          | 4                                                                                                                                | 4               |                                                    |                                                     |  |  |

{1}------------------------------------------------

#### 1.0 Description

The Instruments for endoscopic subfascial Harvesting of Veins (ESHV) and Endoscopic Subfascial Discision of Perforating Beins (ESDP) consists of operating telescopes, endoscopic operating tubes, laser guide, hook knife, various monopolar and bipolar grasping forceps and scissors, and veins dissector (stripper).

#### 2.0 Intended Use

The ESHV (Endoscopic Subfascial Harvesting of Veins) instrument set is used for minimally invasive preparation and harvesting of the veins for use in cardiac and vascular surgical procedures. The intervention takes place under sterile conditions.

The Veins dissector (stripper) is used for the blunt preparation of tubular structures, specifically for the free-preparation of vessels, e.g. the vein saphena magma.

The ESDP (Endoscopic Subfascial Discision of Perforating Veins) instrument set is used for minimally invasive visualization, diagnostic, and therapeutic procedures of subfascial veins in the lower extremities, e.g. the saphenous vein. The intervention takes place under sterile conditions.

#### 3.0 Technological Characteristics

There are no new technological changes on the submitted devices compared to the existing devices. The EDHV endoscope, tubes, and instruments are longer than the ESDP.

#### 4.0 Substantial Equivalence

The submitted devices are substantially equivalent to existing pre-enactment and 510(k) devices sold by Richard Wolf and competitors.

#### 5.0 Performance Data

Performance data not generated.

- 6.0 Clinical Tests No clinical tests performed.
#### 7.0 Conclusions Drawn

These devices are designed and tested to guarantee the safety and effectiveness when used according to the instruction manual.

By: Robert J. Casassa

Robert L. Casarsa Quality Assurance Manager

Date: Oct 10, 97

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol featuring three human profiles connected by flowing lines, representing health and human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 27 1998

Mr. Robert L. Casarsa Manager of Quality Assurance Richard Wolf Medical Instruments Corp. 353 Corporate Woods Parkway Vernon Hills, IL 60064

Re : K973943 Instruments for Endoscopic Subfacial Harvesting of Veins and Endoscopic Subfacial Discision of Perforating Veins Regulatory Class: II (two) Product Code: 78 GCJ Dated: June 23, 1998 Received: June 24, 1998

Dear Mr. Casarsa:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this

{3}------------------------------------------------

response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2 Misbrandinq by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address_"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph. D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K973943 Instruments for Endoscopic Subfascial Harvesting of Veins Device Name: And Endoscopic Subfascial Discision of Perforating Veins

## Intended Use:

The ESHV (Endoscopic Subfascial Harvesting of Veins) instrument set is used for minimally invasive preparation and harvesting of the veins for use in cardiac and vascular surgical procedures. The intervention takes place under sterile conditions.

The Veins dissector (stripper) is used for the blunt preparation of tubular structures. specifically for the free-preparation of vessels, e.g. the vein saphena magma.

The ESDP (Endoscopic Subfascial Discision of Perforating Veins) instrument set is used for minimally invasive visualization, diagnostic, and therapeutic procedures of subfascial veins in the lower extremities, e.g. the saphenous vein. The intervention takes place under sterile conditions.

## Contraindications;

There are no known contraindications directly related to the product. The attending physician must consider the general condition of the patient when determining if the application is appropriate. Refer to the current technical literature for additional instructions.

### Combinations:

Auxiliary instruments from 3.5 mm to 5.5 mm diameter are used through the operation channel. Refer to the corresponding auxiliary instruction manuals.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation

Bera R. Kemper a

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number K973943

Prescription Use
Per 21 CFR 801.109

---

**Source:** [https://fda-staging.innolitics.com/device/K973943](https://fda-staging.innolitics.com/device/K973943)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
