K973932 · Beckman Instruments, Inc. · JIX · Dec 15, 1997 · Clinical Chemistry
Device Facts
Record ID
K973932
Device Name
BECKMAN CALIBRATOR 2 (CAL2)
Applicant
Beckman Instruments, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Dec 15, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
CAL 2 (Calibrator 2), when used in conjunction with Beckman alpha-antitypsin (AAT), ceruloplasmin (CER), properdin factor B (PFB), beta-2-microglobulin (B2M), alpha-macroglobulin (AMG) and antithrombin III (AT3) reagents, is intended for used on Array® 360, and IMMAGE™ Systems for the calibration of these reagents.
Device Story
Beckman Calibrator 2 (CAL2) is a human serum-derived, lyophilized matrix used to calibrate specific protein analyzers. It provides reference values for pre-programmed calibration curves on Array and IMMAGE systems. The device is used in clinical laboratory settings by trained technicians. By establishing calibration points for nephelometric assays, it enables accurate quantification of AAT, CER, PFB, B2M, AMG, and AT3 proteins in patient samples. This calibration process ensures the reliability of diagnostic test results, supporting clinical decision-making regarding the associated disease states.
Clinical Evidence
Bench testing only. The value assignment process for each analyte is correlated to a known standard via the anchor method. Performance data demonstrates that the calibrator yields acceptable assigned values for calibration on both Array and IMMAGE systems, supporting substantial equivalence to existing chemistry test systems.
Technological Characteristics
Lyophilized human serum matrix derived from defibrinated, processed human plasma. Functions as a calibrator for nephelometric protein analyzers. Compatible with Array, Array 360, and IMMAGE Immunochemistry Systems. No software or electronic components; purely a chemical/biological reference material.
Indications for Use
Indicated for use as a calibrator in test systems to establish points of reference for determining concentrations of AAT, CER, PFB, B2M, AMG, and AT3 in human specimens using Array, Array 360, and IMMAGE Immunochemistry Systems.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}------------------------------------------------
BECKMAN
KCN3932
# DFC 1 5 1997
Summary of Safety & Effectiveness Beckman Calibrator 2 (CAL2)
## Submitted By: 1.0
Annette Hellie Sr. Regulatory Specialist, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 993-8767 FAX: (714) 961-4457
## 2.0 Date Submitted:
October 14, 1997
## 3.0 Device Name(s):
## Proprietary Names 3.1
Beckman Calibrator 2 (CAL2)
## Classification Name 3.2
Calibrator (21 CFR § 862.1150)
#### 4.0 Predicate Device(s):
| Device | Predicate | Manufacturer | Docket Number |
|-------------------------|-------------------------|------------------------------|----------------------------------------------------------------------------------------------------|
| Beckman<br>Calibrator 2 | Beckman<br>Calibrator 2 | Beckman<br>Instruments, Inc. | K771603 (AAT)<br>K780913 (PFB)<br>K791339 (CER)<br>K791340 (AMG)<br>K901977 (AT3)<br>K940353 (B2M) |
## 5.0 Description:
Beckman Calibrator 2 is derived from fresh frozen human plasma that has been defibrinated and processed. Assay of CAL 2 provides a response value that is utilized for the adjustment of pre-programmed calibration curves from which AAT, CER, PFB, B2M, AMG, and AT3 concentration values are determined for test specimens.
{1}------------------------------------------------
#### 6.0 intended Use:
CAL 2 (Calibrator 2), when used in conjunction with Beckman alpha-antitypsin (AAT), ceruloplasmin (CER), properdin factor B (PFB), beta-2-microglobulin (B2M), alpha-macroglobulin (AMG) and antithrombin III (AT3) reagents, is intended for used on Array® 360, and IMMAGE™ Systems for the calibration of these reagents.
### 7.0 Comparison to Predicate(s):
The Beckman Calibrator 2 is a lyophilized human serum matrix identical to the current product. The only difference is the additional instrument platform on which it will be used. The existing Beckman CAL 2 is used with Beckman's Array® and Array® 360 Immunochemistry Systems. These systems are fully automated, specific protein analyzers, that measure by nephelometry. The new instrument platform, the IMMAGE™ Immunochemistry System, is also a fully automated, specific protein analyzer that measures by nephelometry.
#### 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. The value assignment process for each analyte is correlated to a known standard via the anchor method. The Beckman Calibrator 2 value assignment and verification processes yield acceptable calibrator assigned values for calibration on the Array® and IMMAGE™ Systems.
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized symbol resembling a person with outstretched arms, with three lines forming the body and arms. The symbol is enclosed within a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 5 1997
Annette Hellie Senior Regulatory Specialist Beckman Instruments, Inc. 200 S. Kraemer Boulevard, W-337 P.O. Box 8000 Brea, California 92822-8000
Re: K973932 Beckman Calibrator 2 Regulatory Class: II Product Code: JIX, JZG, CZP, DCF, DDB, DEB, JBQ Dated: October 14, 1997 Received: October 15, 1997
Dear Ms. Hellie:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity (sail 00), on. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
page page of
510(k) Number (if known):
Beckman Calibrator 2 (CAL2) Device Name:
Indications for Use:
CAL 2 (Calibrator 2), when used in conjunction with Beckman alpha,-antitrypsin (AAT), ceruloplasmin (CER), properdin factor B (PFB), beta-2-microglobulin (B2M), alphamacroglobulin (AMG) and antithrombin III (AT3) reagents, is intended for used on Array®, Array® 360, and IMMAGE™ Systems for the calibration of these reagents.
21 CFR 862.1150 Calibrator
(a) Identification. A calibrator if a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.
(b) Classification. Class II
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Image /page/4/Figure/9 description: The image shows a form titled "Concurrence of CDRH, Office of Device Evaluation (ODE)". The form has two options: "Prescription Use (per 21 CFR 801.109)" and "Over-the-Counter Use Optional Format 1-2-96". The "Prescription Use" option is selected. There is also a signature and a number "R973932" at the bottom of the form.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.