K973868 · Osteotech, Inc. · KWP · Jan 8, 1998 · Orthopedic
Device Facts
Record ID
K973868
Device Name
SSCS (SEGMENTAL SPINAL CORRECTION SYSTEM)
Applicant
Osteotech, Inc.
Product Code
KWP · Orthopedic
Decision Date
Jan 8, 1998
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Segmental Spinal Correction System (SSCS), including the SSCS Offset Hinged Screw, Offset Connector and Drop-on Crosslink Clamps, is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies in the thoraco-lumbo-sacral portion of the spine. It is intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the SSCS are dependent in part on the configuration of the assembled device as described below. When used as a thoracic/lumbar hook or sacral screw and hook system, the SSCS is intended for instrumentation of the spine at levels ranging from T1 to S2 and is indicated for: scoliosis, kyphosis, spinal fractures, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), tumors, stenosis, spondylolisthesis, pseudoarthrosis, previously failed attempts at spinal fusion. When used as a sacral/pedicle screw system, the SSCS is intended for patients: 1. Having severe spondylolisthesis (grade 3 or 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint 2. Who are receiving fusions using autogenous bone graft only. 3. Who are having the device fixed or attached to the lumbar and sacral spine. 4. Who are having the device removed after the development of a solid fusion mass. Although the levels of fusion may not go above the L5-S1 joint, the SSCS screws may be inserted into pedicles up to, but not above, L-3.
Device Story
SSCS is a posterior spinal fixation system; components include offset hinged screws, offset connectors, and drop-on crosslink clamps. System components are surgically implanted and assembled to create a spinal construct; consists of bone screws for spinal attachment, longitudinal rods for load transmission, and transverse connectors for stability. Used by surgeons in clinical settings to stabilize the spine during fusion. Output is a rigid construct that provides mechanical support; aids in correcting spinal deformities and pathologies. Benefits include stabilization of the spine to facilitate bone graft fusion. Mechanical performance is verified via bench testing.
Clinical Evidence
Bench testing only. Mechanical performance characteristics, including bending-compression static strength, stiffness, torsional stiffness, and fatigue strength, were evaluated per ASTM provisional standards for spinal implants. Results demonstrated performance comparable or superior to previously tested SSCS constructs and predicate devices.
Technological Characteristics
Posterior spinal fixation system; components include longitudinal rods, bone screws (hinged/offset), and transverse connectors. Materials are consistent with previously cleared SSCS components. Mechanical assembly via lateral connectors and crosslink clamps. Testing performed per ASTM provisional standards for spinal implants.
Indications for Use
Indicated for patients requiring posterior spinal fixation for scoliosis, kyphosis, spinal fractures, degenerative disc disease, tumors, stenosis, spondylolisthesis, pseudoarthrosis, or failed fusion (T1-S2). Pedicle screw use restricted to severe spondylolisthesis (Grade 3/4) at L5-S1, using autogenous bone graft, with screw insertion limited to L3-S1.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
VIII. 510(k) Summary of Safety and Effectiveness
JAN - 8 1998
1. Submission Applicant & Correspondent:
| Name: | Osteotech, Inc. |
|------------------|-------------------------------------|
| Address: | 51 James Way<br>Eatontown, NJ 07724 |
| Phone No.: | (732) 542-2800 |
| Contact Person: | Christopher Talbot |
| Date of Summary: | October 2, 1997 |
2. Name of Device:
| Trade/Proprietary/Model Name: | SSCS (Segmental Spinal<br>Correction System) - Addition of<br>Components: Offset Hinged Screw<br>and Connector; Drop-on Crosslink<br>Clamps |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Posterior Spinal Fixation Device |
| Classification Name: | Spinal Interlaminal Fixation<br>Orthosis; Spondylolisthesis<br>Spinal Fixation Device System |
3. Devices to Which New Device is Substantially Equivalent:
| Trade/Proprietary/Model Name |
|------------------------------|
| SSCS (K955173;K971289) |
| ISOLA |
| TSRH |
| CD (Cotrel-Dubousset) |
| USS |
Manufacturer Heinrich C. Ulrich Acromed Sofamor-Danek Sofamor-Danek Synthes
4. Device Description:
The SSCS (Segmental Spinal Correction System) is a spinal fixation system comprised of various types and sizes of components, including the SSCS Offset Hinged Screw, Offset Connector and Drop-on Crosslink Clamps, that are implanted via a posterior surgical
{1}------------------------------------------------
approach and assembled to create a spinal construct. The SSCS also includes instrumentation/accessories that aid in assembling and/or implanting the components. Like most other posterior spinal fixation systems, the SSCS is comprised basically of 1) bone screws for attachment of the device to the spine, 2) longitudinal rods that are attached to the bone screws either directly or indirectly by means of lateral connectors, and that transmit loads across the pathologic segments of the spine, and 3) optional transverse connecting elements that link the two longitudinal rods for added construct stability.
# 5. Intended Use/Indications
The Segmental Spinal Correction System (SSCS), including the SSCS Offset Hinged Screw, Offset Connector and Drop-on Crosslink Clamps, is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies in the thoraco-lumbo-sacral portion of the spine. It is intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the SSCS are dependent in part on the configuration of the assembled device as described below.
When used as a thoracic/lumbar hook or sacral screw and hook system, the SSCS is intended for instrumentation of the spine at levels ranging from T1 to S2 and is indicated for:
- scoliosis
- kyphosis
- spinal fractures
- degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).
- tumors
- stenosis
- spondylolisthesis
- pseudoarthrosis
- previously failed attempts at spinal fusion
When used as a sacral/pedicle screw system, the SSCS is intended for patients:
- 1. Having severe spondylolisthesis (grade 3 or 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint
- 2. Who are receiving fusions using autogenous bone graft only.
- 3. Who are having the device fixed or attached to the lumbar and sacral spine.
{2}------------------------------------------------
- 4. Who are having the device removed after the development of a solid fusion mass.
Although the levels of fusion may not go above the L5-S1 joint, the SSCS screws may be inserted into pedicles up to, but not above, L-3.
# 6. Technical Comparison
The SSCS, with the addition of the Offset Hinged Screw, Offset Connector and Drop-on Crosslink Clamps, is similar to the previously cleared SSCS that included Hinged Screws (K971289) and to other predicate devices in terms of its design, technical characteristics, and materials. These devices consist of several basic types and various sizes of implantable components including longitudinal rods, bone screws, and transverse connectors that are assembled to create a spinal construct.
The SSCS Offset Hinged Screws, like the SSCS Hinged Screws and the bone screws of other predicate spinal systems serve as anchoring elements to attach the device to the spine. In all of these devices, bilaterally placed rods, the longitudinal elements of the spinal system constructs, are attached to the heads of the implanted screws directly or indirectly by means of a lateral connector.
These new SSCS components share many of the same design characteristics (including the same interconnection mechanism for attachment of the components) and are comprised of the same materials as previously cleared SSCS components. In addition, the SSCS Offset Hinged Screw is similar to the SSCS Hinged Screw in that both types of screws have the screw head attached to the screw shaft by a hinge. However, the Offset Hinged Screw is designed for attachment to the Offset Connector which, in turn, attaches to the longitudinal rod whereas the Hinged Screw attaches directly to the longitudinal rod.
# 7. Performance Data
Mechanical testing of SSCS constructs incorporating the Offset Hinged Screw. Offset Connector and Drop-on Crosslink Clamps was performed in accordance with the ASTM Provisional Standard for testing spinal implant devices. The test results demonstrated that the mechanical performance characteristics (bending-compression static strength and stiffness, torsional stiffness and fatigue strength) of SSCS constructs with these new components are at least comparable to, if not better than, those of previously tested SSCS constructs and/or other predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a single snake entwined around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 8 1999
Mr. Christopher W. Talbot Manager of Regulatory Affairs Osteotech, Inc. 51 James Way Eatontown, New Jersey 07724
K973868 Re : SSCS Offset Hinged Screw, Offset Connector, Drop-on Crosslink Clamps, Hinged Screw and SSCS Rods to be used with the SSCS (Segmental Spinal Correction System) Regulatory Class : II Product Codes: MNH and KWP Dated: October 8, 1997 October 10, 1997 Received:
Dear Mr. Talbot:
We have reviewed your Section 510 (k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal, Food, Drug, and Cosmetic Act This decision is based on your device system being (Act) . found equivalent only to similar device systems labeled and intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid You may, therefore, market your device system fusion mass. subject to the general controls provisions of the Act and the limitations identified below.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment for indications other than severe spondylolisthesis, as described above, would cause the device system to be adulterated under 501(f) (1) of the Act.
{4}------------------------------------------------
### Page 2 - Mr. Christopher W. Talbot
This device system, when intended for pedicular screw fixation/attachment to the spine for indications other than severe spondylolisthesis, as described above, is a class III device under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:
- All labeling for this device, including the package 1. label, must state that there are labeling limitations. The package insert must prominently state that the device system using pedicle screws is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
- You may not label or in anyway promote this device system 2. for pedicular, screw fixation/attachment to the cervical, thoracic or lumbar vertebral column for intended uses other than severe spondylolisthesis, as described above. The package insert must include the following statements:
#### WARNINGS :
- . When used as a pedicle screw system, this device system is intended only for grade 3 or 4 spondylolisthesis at the fifth lumbar - first sacral (L5-S1) vertebral joint.
- The screws of this device system are not intended . for insertion into the pedicles to facilitate spinal fusions above the L5-S1 vertebral joint.
- . Benefit of spinal fusions utilizing any pedicle screw fixation system has not been adequately established in patients with stable spines.
- . Potential risks identified with the use of this device system, which may require additional surgery, include:
device component fracture, loss of fixation, non-union, fracture of the vertebra, neurological injury, and vascular or visceral injury.
{5}------------------------------------------------
See Warnings, Precautions, and Potential Adverse Events sections of the package insert for a complete list of potential risks.
- Any pedicular screw fixation/attachment for intended uses 3 . other than severe spondylolisthesis, as described by item 1, for this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment for intended uses other than severe spondylolisthesis, as described above, must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation.
- Any previous warning statements identified as part of 4 . previous 510(k) clearances or required by OC/Labeling and Promotion which stated that a component/system was not approved for screw fixation into the pedicles of the spine must be replaced by the warnings of items 1 and 2 above .
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under-sections 531 -----through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
FDA advises that the use of your device system with any other device components but those identified in this 510(k) would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system
{6}------------------------------------------------
Page 4 - Mr. Christopher W. Talbot
consisting of the subject device components and other device components, whether yours or those of other manufacturers, may also be required.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## I. Indications for Use Statement
510(k) Number (if known): K973868 Device Name: SSCS (Segmental Spinal Correction System)
Indications For Use:
The SSCS is intended for use as a posterior spinal fixation device, with specific indications as follows:
- 1) When used as a thoracic/lumbar hook or sacral screw and hook system, the SSCS is intended for instrumentation of the spine at levels ranging from T1 to S2 and is indicated for:
- scoliosis
- kyphosis
- spinal fractures
- degenerative disc disease, defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
- tumors
- stenosis
- spondylolisthesis
- pseudoarthrosis
- previously failed attempts at spinal fusion
- 2) When used as a sacral/pedicle screw system, the SSCS is intended for patients:
- a. Having severe spondvlolisthesis (grade 3 or 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint
- b. Who are receiving fusions using autogenous bone graft only.
- c. Who are having the device fixed or attached to the lumbar and sacral spine.
- d. Who are having the device removed after the development of a solid fusion mass.
Although the levels of fusion may not go above the L5-S1 joint, the SSCS screws may be inserted into pedicles up to, but not above, L-3.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
fcoàl
(Division Sign-Off) Division of General Restorative Devices
K973868 510(k) Nymber Over-The-Counter Use
Prescription Use X (Per 21 CFR 801.109)
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.