FLOSTAT VASCULAR REPORT GENERATOR (VRG)
Device Facts
| Record ID | K973857 |
|---|---|
| Device Name | FLOSTAT VASCULAR REPORT GENERATOR (VRG) |
| Applicant | Biomedix, Inc. |
| Product Code | JAF · Radiology |
| Decision Date | Dec 29, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1540 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device |
Indications for Use
The BioMedix, Inc. Flostat Vascular Report Generator is a device for the use to display the outputs from three strip chart recorder based peripheral vascular diagnostic instruments on a PC printer. In essence, this is changing one printer for another. This generic device can bu used to input a signal from any device into a is intended soley for use in conjunction with computer but Blood pressures, displayed peripheral diagnostic instruments. digitally, may be recorded with the original cuffs provided by the original manufacturer. No change in equipment or technique is warranted. The three target devices are the Imex Corp. 3000 Series, 301 The Life Sciences Corp./HealthWatch PVR Series Series, and All these incorporate analog or digital pen diagnostic devices. strip chart recorders to display the output waveforms. The sole use of the VRG is to display the waveforms on the PC printer. une lowest resolution dot matrix printer with 90x90 dots per inch(and certainly much higher resolution laser printers) has a much higher resolution than the relatively meager 100 Hz bandwidth that an analog pen recorder can provide. No alterations in the clinical instructions, indications, methodologies, or outcomes will occur. The VRG is simply connected to an analog output path on the target The VRG is then connected(via a high ESD rated RS-232 device. driver) to the computer. The waveforms are brought into the computer and the opportunity is provided for printing to a printer. The intended use of the VRG-Vascular Diagnostic Unit tandem is the the original intended use of the original vascular same រួន diagnostic unit(ie. peripheral vascular studies). The BMX Vascular Report Generator passes the output information of a peripheral vascular deviceto a computer and then onto a printer. All ourputs of a peripheral vascular device, throuhg the VRG, can displayed, including data that can be shown as a spectrum analysis. The Vascular Report Generator does not change the output of a peripheral vascular devicebut only allows the computer to display the informan: in various forms .
Device Story
Device acts as interface between peripheral vascular diagnostic instruments (strip chart recorders) and PC printer; inputs analog signals from vascular diagnostic devices; transforms signals via RS-232 connection to computer for digital display and printing; enables printing of waveforms and spectrum analysis data on standard PC printers; used in clinical settings where vascular studies are performed; operated by healthcare professionals; output allows clinicians to view diagnostic data on higher resolution media than traditional pen recorders; benefits patient by facilitating clearer documentation of peripheral vascular studies without altering original diagnostic methodology.
Clinical Evidence
No clinical data. Bench testing only; device functionality verified by comparing output resolution of PC printers (90x90 dpi) against analog pen recorder bandwidth (100 Hz).
Technological Characteristics
Interface device; connects to analog output of vascular diagnostic instruments; utilizes high ESD-rated RS-232 driver for computer connectivity; outputs to standard PC printers; no modification to diagnostic signal acquisition; standalone hardware/software interface.
Indications for Use
Indicated for use in peripheral vascular studies in conjunction with existing peripheral vascular diagnostic instruments (Imex Corp. 3000 Series, 301 Series, and Life Sciences Corp./HealthWatch PVR Series) to display and print diagnostic waveforms and data.
Regulatory Classification
Identification
A nonfetal ultrasonic monitor is a device that projects a continuous high-frequency sound wave into body tissue other than a fetus to determine frequency changes (doppler shift) in the reflected wave and is intended for use in the investigation of nonfetal blood flow and other nonfetal body tissues in motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Reference Devices
- Imex Corp. 3000 Series
- Imex Corp. 301 Series
- Life Sciences Corp./HealthWatch PVR Series