K973818 · Boston Scientific Corp · DWZ · Dec 16, 1997 · Cardiovascular
Device Facts
Record ID
K973818
Device Name
T-REX BIOPSY FORCEPS
Applicant
Boston Scientific Corp
Product Code
DWZ · Cardiovascular
Decision Date
Dec 16, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4075
Device Class
Class 2
Indications for Use
The T.REX™ biopsy forceps is used to obtain endomyocardial biopsy specimens from the right or left ventricle via percutaneous arterial or venous approach.
Device Story
Device is a percutaneous endomyocardial biopsy forceps used by clinicians to obtain heart tissue samples. Input is manual manipulation of a handle assembly; force is transmitted via a solid stainless steel pull wire to an end-effector jaw assembly. Actuator links translate longitudinal force into vertical jaw movement. Device features a locking mechanism to hold jaws open for specimen removal and a spring coil to maintain closed position. Used in clinical settings via percutaneous arterial or venous access. Output is a diagnostic tissue specimen. Benefits include improved performance and manufacturability over predicate via laser-welded joints, pre-curved body options, and ergonomic handle design.
Clinical Evidence
No clinical data. Safety and effectiveness established via bench testing, including comparative cutting ability, tensile strength of joints, and biocompatibility assays (cytotoxicity, pyrogenicity, hemocompatibility) conducted under GLP. Sterilization validation confirmed a 10^-6 SAL.
Technological Characteristics
Materials: surgical stainless steel. Components: handle assembly, body with solid stainless steel pull wire, end-effector jaw assembly. Features: laser-welded joints, pre-curved or flexible body, integral strain relief, handle locking mechanism, spring coil. Sterilization: 100% Ethylene Oxide (EtO).
Indications for Use
Indicated for obtaining endomyocardial biopsy specimens for diagnosis of diseased heart tissue or identification and monitoring of rejection factors in a transplanted heart.
Regulatory Classification
Identification
An endomyocardial biopsy device is a device used in a catheterization procedure to remove samples of tissue from the inner wall of the heart.
{0}------------------------------------------------
h9738818
# 510(k) for the T•REX™ Biopsy Forceps
## 510(k) SUMMARY
#### A. Introduction
This 510(k) is for the T-REX™ biopsy forceps previously manufactured as the Bycep® biopsy forceps by Mansfield, Inc., a subsidiary of Boston Scientific Corporation. Research, development and manufacturing of the Bycep forceps has been relocated to Boston Scientific Corporation Northwest Technology Center, Inc. Several design modifications have been made to improve the performance and manufacturablility of the device. The indications for use and operation of this device remain the same.
| Submitter: | Boston Scientific Corporation Northwest Technology Center, Inc.<br>17425 N.E. Union Hill Road<br>Redmond, WA 98052 | DEC 16 1997 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------|-------------|
| Contact: | Jocelyn Kersten<br>Phone: (425) 556-1667<br>Fax: (425) 558-1400 | |
| Preparation Date: | 10/06/97 | |
| Device Common Name: | Endomyocardial biopsy forceps | |
| Device Proprietary Name: | T•REX™ Biopsy Forceps | |
| Classification Name: | Device, biopsy, endomyocardial (74DWZ) (per 21 CFR 870.4875) | |
| Classification Panel: | Cardiovascular | |
| Manufacturing Facilities: | Heart Technology Manufacturing Center, Inc.<br>17425 N.E. Union Hill Road<br>Redmond, WA 98052 | |
#### B. Device Description
The T.REX™ forceps are designed to allow percutaneous access to the right or left ventricles of the heart in order to obtain diagnostic tissue samples. The forceps consist of three main components: a handle ergonomically designed for comfortable use, a body and surgical stainless steel cutting jaws.
## C. Intended Use
The T.REX™ biopsy forceps is used to obtain endomyocardial biopsy specimens from the right or left ventricle via percutaneous arterial or venous approach.
{1}------------------------------------------------
## Comparison to Predicate Device D.
The proposed T•REX™ forceps are substantially equivalent to the previously approved Bycep forceps. The indications for use remain the same, as well as the basic structure and operation of the device.
The following table summarizes the modifications.
| Component | Bycep® Forceps (K870186) | T-REX Forceps |
|--------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------|
| End-Effector and Jaw Assembly | | |
| Jaws | Machined to sharpness | Proprietary sharpening method |
| Actuator links | Soft or semi-hardened stainless steel | Hardened stainless steel |
| Actuator/pull wire connection | Brazed | Laser welded |
| Housing | Single unit | Laser welded housing halves |
| Housing to outer jacket connection | Soldered | Laser welded |
| Crank length/lever angle | 2.2 mm jaws, 0.0822"; 1.8 mm jaws, 0.0427" /20° | 2.2 mm jaws, 0.0774"; 1.8 mm jaws, 0.0541" /15° |
| Attachment of actuator/actuator links and jaws/housing | Rivets | Captured by laser welded housing halves |
| Body | | |
| Pull Wire | Braided stainless steel cable | Solid stainless steel wire |
| Shape | Flexible | Flexible and pre-curved |
| Liner | No liner | Liner between outer jacket and pull wire |
| Jacket sheath | None | Jacket sheath covering the outer jacket |
| Handle assembly | | |
| Handle transition to outer jacket | Solder joint with coil strain relief around the joint | Crimped joint with integral strain relief |
| Handle transition to handle body | Brass handle transition press fit into straight handle body | Stainless steel handle transition press fit into tapered handle body |
| Finger grip/Handle body | No locking feature | Locking feature for holding jaws open during specimen removal |
| Finger grip insert for attachment of pull wire | Brass/Sonic welding | Aluminum/Press fit |
| Handle spring | None | ½" diameter spring coil |
| Package | | |
| Configuration | Double barrier | Single barrier |
| Product holder | Tray | Coil or card |
•
3
{2}------------------------------------------------
# 510(k) for the T-Rex™ Biopsy Forceps
## End effector and Jaw Assembly 1.
In both the Bycep models and the T.REX™ models of the biopsy forceps, the pull wire, which translates movement of the handle into opening and closing the jaws, is attached to the jaws through the end effector assembly. The Bycep models utilize brazing, soldering and rivets to construct the necessary joints in the end effector assembly. The T.REXTM models utilize laser welding to construct these joints. The jaws of the Bycep forceps are sharpened with machining while the jaws of the T.REX™ models are sharpened using a proprietary process.
In both models, when the handle is manipulated, the force is transmitted down the pull wire to the actuator. The actuator links then translate the longitudinal force from the handle into vertical force on the jaws, which opens the jaws. When the handle is released, the jaws close. Gentle force on the thumb ring of the handle ensures the jaws remain closed tightly enough to retain a cut tissue sample.
#### 2. Bodv
-
The body of the forceps consists of a hollow outer jacket which contains the pull wire. The pull wire in the Bycep models is a braided stainless steel cable. The T.REXTM models contain a solid stainless steel pull wire. The solid wire allows forming by the manufacturer to produce the pre-curved models and allows the flexible models to be formed by the customer prior to use. The outer jackets of the T-REX™ models are lined and covered by a sheath. The Bycep forceps contain no liner and are not coated.
#### 3. Handle Assembly
The stainless steel handle transition is crimped onto the outer jacket of the T.REX™ models. The outer sheath is heat shrunk over the distal end of the handle transition to serve as an integral strain relief. The outer jacket of the Bycep models is soldered to the brass handle transition and utilizes a stainless steel coil strain relief. The handle transitions of both models are press fit into the handle bodies. The mating portion of the handle body is cylindrical in the Bycep models and reinforced by ribs in the T.REX™ model.
The handles of both the Bycep and T.REX models consist of a handle body, finger grip and thumb ring. Two fingers are placed into the finger grip and the thumb is placed in the thumb ring. The thumb is pulled back to open the jaws and released or pushed forward to close them. A spring coil within the handle of the T.REX™ models holds the jaws closed when the handle is in the neutral position. Also, a locking feature on the handle will secure the jaws in an open position to aid in removing the tissue sample from the jaws. The Bycep models do not contain the handle spring or the locking feature.
4
{3}------------------------------------------------
## 510(k) for the T-Rex™ Biopsy Forceps
In both models, the handle operates the jaws via the pull wire. The pull wire is attached to the handle at the finger grip insert. This insert allows the pull wire to be captured between an aluminum disc and the insert, and held in place by a set screw. The method of attachment is the same for both models, however, in the T-REX™ models, the finger grip insert is aluminum and is press fit into the finger grip, while in the Bycep® models, the finger grip insert is brass and is sonically welded into the finger grip.
## 4. Package
Both models of the forceps are provided sterile for single procedure use. The Bycep forceps are packaged in a tray contained in a double sterile barrier. The T.REX™ forceps pre-curved models are attached to a card and the flexible models are held in a plastic coil. Both the pre-curved and the flexible models are packaged in a single sterile barrier. Sterilization and packaging testing confirmed the proposed packaging will protect the product and maintain sterility.
### ડ. Performance Testing (In Vitro)
Comparative testing of the cutting abilities of the T.REX™ forceps and the Bycep forceps was performed. This testing confirmed the T.REX™ forceps are capable of performing the intended function of biopsy forceps, which is to obtain samples of myocardium.
### E. In Vitro Tests
Bench testing was done to show that models of the T.REX™ forceps built with the proposed design modifications continue to be safe and effective. Testing included the modified joints and the overall tensile strength of complete devices. Testing confirmed the T.REX™ forceps met its design specifications. Therefore, BSC Northwest concludes that the T.REXTM forceps are safe and effective.
## F. Biocompatibility
A biological evaluation was performed on the T.REX™ forceps to give a high degree of assurance that the final product is safe for human use. These tests included material characterization, cytotoxicity, pyrogenicity and hemocompatibility assays conducted under Good Laboratory Practices (GLP). The blood contacting portion of the device was evaluated as a whole composite following two ethylene oxide (EtO) sterilization cycles. The T.REXTM forceps were found to be compatible with biological systems.
## G. Sterilization
Following final packaging and labeling, finished devices are 100% EtO sterilized to achieve a sterility assurance level (SAL) of 10-6, or one nonsterile unit out of one million, Sterilization validation testing confirmed the ETO sterilization cycle is capable of delivering a minimum of 12 log reductions and confirmed a minimum Sterility Assurance Level (SAL) of 10 . The T.REX™ forceps are nonpyrogenic based on the Limulus Amebocyte Lysate (LAL) assay.
{4}------------------------------------------------
## H. Clinical Tests
Clinical studies were not conducted on the T.REX™ forceps. The aspects of the device that relate to patient safety are: 1) cutting ability, 2) tensile strength of the devices, 3) biocompatibility of the materials in the fluid contact path and 4) device sterility. All of these aspects of the T.REX™ forceps have been tested, as described above, and results show the device to be safe and effective.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle.
Public Health Service
Rockville MD 20857
Ms. Julie Wage Director, Requlatory Affairs Boston Scientific Corporation Northwest Technology Center, Inc. 17425 N.E. Union Hill Road Redmond, WA 98052-3376
DEC 1 6 1997
Re : K973818 T.REX™ Biopsy Forceps Requlatory Class: II (Two) Product Code: DWZ Dated: October 6, 1997 Received: October 7, 1997
Dear Ms. Wage:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 - Ms. Julie Wage
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
| 510(k) Number (if known): | K 1) 38 i 8 |
|---------------------------|-------------|
|---------------------------|-------------|
Device Name:
Indications For Use:
T.REX™ Biopsy Forceps Device:
The T-REX™ biopsy forceps are intended to obtain endomyocardial biopsy specimens. The specimens are taken for diagnosis of diseased heart tissue or for identification and monitoring of rejection factors in a transplanted heart.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices R9 138 510(k) Number Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) :
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.