K973810 · Dolphin Medical Technologies, Inc. · HWC · Jan 5, 1998 · Orthopedic
Device Facts
Record ID
K973810
Device Name
DRG LARGE SCREW SYSTEM
Applicant
Dolphin Medical Technologies, Inc.
Product Code
HWC · Orthopedic
Decision Date
Jan 5, 1998
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dolphin Medical Technologies DRG Large Screw™ System will be used on indications that are common with presently marketed devices. The indications for use of the DRG Large Screw™ System reconstruction of intra-anicular epiphyseal and of metaphyseal fractures. Specifically, intracapsular fractures of the hip, slipped capital femoral epiphyses, tibial plateau fractures, distal femoral condyle fractures, ankle, elbow and shoulder fractures where indicated.
Device Story
DRG Large Screw™ System consists of titanium alloy (6AL-4V ELI) bone screws used for fracture fixation and stabilization. Screws are available in 6.5 mm diameter with lengths ranging from 25-120 mm. Device is implanted by surgeons using standard manual orthopedic surgical instruments (stainless steel, ASTM F899-84). Operation involves manual insertion via cannulated hexhead screwdriver. System provides mechanical stabilization of bone fragments to facilitate healing. Used in clinical/surgical settings. Benefits include established clinical performance for fracture fixation based on over 75 years of orthopedic practice.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on established clinical performance of similar bone screw systems used for over 75 years and adherence to voluntary ASTM material and design standards.
Technological Characteristics
Material: Titanium alloy 6AL-4V ELI (ASTM F136-84). Instrumentation: Stainless steel (ASTM F899-84). Design: Cannulated bone screw, 6.5 mm diameter, 25-120 mm lengths. Principle: Mechanical fixation/stabilization. Sterilization: Non-sterile, provided in clean condition for user sterilization.
Indications for Use
Indicated for reconstruction of intra-articular epiphyseal and metaphyseal fractures, including intracapsular hip fractures, slipped capital femoral epiphyses, tibial plateau fractures, distal femoral condyle fractures, and ankle, elbow, and shoulder fractures. Contraindicated for pedicular screw fixation/attachment to the cervical, thoracic, or lumbar spine.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date January 5, 1998. The month is abbreviated to JAN. The numbers are written in a simple, sans-serif font. The image is a close-up of the date.
K973810
Confidential
#### FDA Notification of:
## Summary of Safety and Effectiveness Information Product: DRG Large Screw™ System
# of Safety and Effectiveness Information
### For Release Upon Request Only
### Regulatory Authority:
Safe Medical Devices Act of 1990, 21 CFR 807.92
### Company Name / Contact:
| Company: | DMT (DolphinMedical Technologies, Inc.)<br>5959 Cattlemen Lane<br>Sarasota, FL 34232 |
|------------------------------------|-----------------------------------------------------------------------------------------------------|
| Contact: | Regulatory Affairs Department<br>DMT<br>5959 Cattlemen Lane<br>Sarasota, FL 34232<br>(941) 342-0414 |
| Establishment Registration Number: | Applied for, not received |
| Classification Name: | Smooth or Threaded Bone Fixation<br>Fastener |
| Common Used Name: | Bone Screw |
| Trade Proprietary Name: | DRG Large ScrewTM System |
The FDA has classified similar products as a Class II device by the Orthopedic Device Section of the Surgical and Rehabilitation Devices Panel at Section 888-304. The product code generally refereed to is HWC ( Product Code: HWC ) , and DMT submits this application under this designation.
Section 5 - 1
115
{1}------------------------------------------------
### FDA Notification of:
### Summary of Safety and Effectiveness Information Product: DRG Large Screw™ System
### Performance Standards:
No performance standards applicable to the Bone Screw have been established by the FDA. However, the titanium alloy 6AL-4V ELI alloy used to manufacture the DMT screws meets the chemical and mechanical requirements in voluntary standards established by the American Society for Testing and Materials (ASTM F136-84).
#### Package and Labeling:
Package labeling has been developed to industry standards. Packaging is also standard commercially available type quality and is stored in a fashion which prevents damage to the container or package the device is in.
#### System Description:
The DMT DRG Large Screw™ System will be offered in Ti-GAL-4V ELL It will be available in common styles and assorted lengths for bone fraction and stabilization. Initially, a range of twenty-four screw lengths will be made available in 6.5 mm diameter (25-120 mm). Screw types are implantable using a standard (e.g. American Orthopedic) hexhead screwcriver, which is cannulated at center.
#### Indications for Use:
The DMT DRG Large Screw™ System will be used on indications that are common with presently marketed devices. The primary indications for use of the DRG Large Screw™ System are reconstruction of intra-articular epiphysed and of metaphysed fractures. Specifically, intracapeular fractures of the hip, slipped capital femoral epiphyses, tibial platesu fractures, distal femoral condyle frectures, ankle, elbow, and shoulder fractures where indicated. Fixations has been achieved in literally hundreds of thousands of cases with similar predicate devices. Ref. 1.2
Section 5 - 2
{2}------------------------------------------------
Confidential
# FDA Notification of:
儿妹
11:4
# Summary of Safety and Effectiveness Information Product: DRG Large Screw™ System
# Substantial Equivalent Devices:
This product is substantially equivalent in design, composition and function to other orthopedic screws manufactured and approved for market.
| Ace Medical Company: | K903810 |
|----------------------|---------|
| Alphatec Medical: | K921622 |
| Howmedica: | K931524 |
| Aesculap: | K940207 |
| Osteomed: | K924018 |
| Zimmer: | K792022 |
| A.O. Synthes | K792291 |
The DMT DRG Large Screw™ System meet the ASTM standards (ASTM B348-83, F136-84, F67-The for material and design for medical application. The bome serews are of the same thread 60) 10. Minutial and length as offered by Synthes, Zimmer, Johnson, Ace, Alphatec and many other orthopaedic companies. The minor and major dismeters as well as the head size are comparable.
{3}------------------------------------------------
#### FDA Notification of:
## Summary of Safety and II.flectiveness Information Product: DRG Large Screw™ System
#### Instrumentation:
DMT DRG Large Screw™ System instrumentation used for the preparation and insertion of the DRG Large Screws is considered to be general orthopsedic instrumentation. The system includes standard manual orthopedic surgical instruments of the appropriate size and type. All DRG Large Screw™ System instruments are manufactured from stainless steel meeting ASTM F899-84 standards.
#### Product Storilization:
DMT will supply all instruments and implants Non-Sterile. Non-Starile implants are packaged in "clean only" condition. The labeling of the implants and instruments clearly indicates their sterility status. The package insert contains a sterilization/ro-stailization guideline.
#### Summary:
Substantial Equivalence for the DRG Large Screw™ System may be found in comparison with devices from a number of manufactures. Bone Screw systems in general have been used for many years, and the clinical performance is well known and documented.
Another measure of the Safety and Effectiveness of a medical device is how it performs in Jong term use. The basic design concept of bone screws for use in the fixation and stabilization of fractures has had over 75 years of clinical evaluation. Uses, Indications, limitations and surgical techniques are well understood. Standardized manufacturing methods, design practices, material selections and testing techniques are known and represented within the guidelines of this submittal.
r
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a logo with a circular emblem on the left and a stylized graphic on the right. The emblem contains text around the perimeter, partially visible, and a solid black circle in the center. To the right, there are three curved lines stacked vertically, resembling a stylized representation of people or movement. The overall impression is of a formal or official seal, possibly representing an organization or department.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 5 1998
Mr. Mark G. Urbanski President & CEO Dolphin Medical Technologies, Inc. 5959 Cattlemen Lane Sarasota, Florida 34232
K973810 Re : DRG Large Screw System Trade Name: Regulatory Class: II Product Code: HWC September 26, 1997 Dated: Received: October 7,1997
Dear Mr. Urbanski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting a device for pedicular screw fixation/attachment would cause the device to be adulterated under 501(f)(1) of the Act. This device, if intended for use in pedicular screw fixation/attachment, would be found not substantially equivalent and would be a class III device under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:
- The package insert must prominently state that the device 1 . is intended for the specific use(s) described in the enclosure only; and
{5}------------------------------------------------
- You may not label or in any way promote this device for 2. pedicular screw attachment to, or fixation of the pearcular Berem asses lumbar vertebral column. If this derice is a screw with outer diameters of 3 mm - 10 mm and overall lengths of 10 mm - 75 mm inclusively, the package insent must include the following statement, This device is not approved for screw "WARNING: attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine." (pedicular screw fixation/attachment to the cervical, thoracic or lumbar vertebral column of this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conduct of the investigation.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS). for Medical ........ Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
{6}------------------------------------------------
If you desire specific advice for your device on our labeling
and and the first Book and additionally 809.10 for in If you desire specific advice tor your and additionally 809.10 for in
regulation (21 CFR Part 801 and additionally 809.10 for the regulation (21 CFR Part 801 and addresses the Office of
vitro diagnostic devices), please contact the Office of vitro diagnostic devices), prease contact une on questions on
Compliance at (301) 594-4659. Additionally, for questions on Compliance at (301) 594-4659. "Address," bour device, please contact
the promotion and advertising of your device, please not the promotion and advertising of your actise, also, please note
the Office of Compliate at (301) 594-4639. Also, please note the Office of Compliance ac (505) 330 by reference to the regulation entification" (21 CFR 807.97). Other general premarket notification" (21 cm our the Act may be information on your responsibilities ands - earers Assistance
obtained from the Division of Small Manufacturers (301) 443-6597 or 2 obtained from the Division of Smart Manufactor (301) 443-6597 or at
at its toll-free number (800) 638-2041 or (301) 443-6597 or at at its toll-free number (800) 638-2041 or (60%)
its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Confidential
January 2, 1998
Food and Drug Administration Center for Devices and Radiological Health Office of Evaluation Document Mail Center (HFZ-401) 1390 Piccard Drive Rockville, Maryland 20850
K973810 510(k) Application Product: DRG Large Screw™ System
510 (k) No.# :
Device Name: DRG Large Screw™ System
Indications For Use: As listed in the cover letter for this 510(k) application, are as follows:
The Dolphin Medical Technologies DRG Large Screw™ System will be used on indications that are common with presently marketed devices. The indications for use of the DRG Large Screw™ System reconstruction of intra-anicular epiphyseal and of metaphyseal fractures. Specifically, intracapsular fractures of the hip, slipped capital femoral epiphyses, tibial plateau fractures, distal femoral condyle fractures, ankle, elbow and shoulder fractures where indicated.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ..... ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(Per 21 CFR 801.109)
Over-The-Counter Use ... . . ........
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.