K973659 · Vasocor, Inc. · JOM · Dec 17, 1997 · Cardiovascular
Device Facts
Record ID
K973659
Device Name
VASOCOR PVR 100
Applicant
Vasocor, Inc.
Product Code
JOM · Cardiovascular
Decision Date
Dec 17, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2780
Device Class
Class 2
Indications for Use
The Vasocor PVR-100 (Pulse Volume Recorder) is a non-invasive medical device used in conjunction with other devices such as; continuous-wave Doppler ultrasound, treadmill testing, and ultrasonic imaging techniques for vascular studies of limbs and digits. Use of the PVR in this manner allows the physician to non-invasively diagnosis extremity arterial and venous disorders.
Device Story
Vasocor PVR-100 is a PC-based air plethysmography system for pulse volume recording (PVR). It uses standard blood pressure cuffs applied to limbs or digits; a pneumatic circuit injects air to ensure skin contact. A pressure transducer records pressure changes within the cuff caused by volume changes in the underlying extremity segment during cardiac cycles. The device performs internal calibration and manages patient data, test performance, and report printing. Used in clinical settings by physicians, it complements other vascular diagnostic tools like Doppler ultrasound or treadmill testing. The output consists of PVR tracings and reports, which assist clinicians in diagnosing arterial and venous disorders. The system benefits patients by providing a non-invasive method for vascular assessment.
Clinical Evidence
Clinical evaluation compared the Vasocor PVR-100 to the Life Sciences Pulse Volume Recorder in 6 subjects (3 normal, 3 with peripheral vascular disease). Cuffs were placed on the upper arm, thumb, calf, and ankle. Results demonstrated substantial equivalence in PVR tracing and contour analysis. Bench testing included a hemodynamic model of lower extremity arterial circulation, varying flowrate, compliance, and resistance to validate PVR amplitude and contour accuracy.
Technological Characteristics
PC-based system; pneumatic circuit for cuff inflation and internal calibration; pressure transducer for sensing volume-related pressure changes. Uses standard blood pressure cuffs. Software controls patient records, tracings, and pneumatic functions. Connectivity includes external printer. No specific material standards or software architecture class provided.
Indications for Use
Indicated for non-invasive diagnosis of extremity arterial and venous disorders in patients requiring vascular studies of limbs and digits. No specific contraindications noted.
Regulatory Classification
Identification
A hydraulic, pneumatic, or photoelectric plethysmograph is a device used to estimate blood flow in a region of the body using hydraulic, pneumatic, or photoelectric measurement techniques.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a sequence of digits, specifically the number 1973659. The digits are written in a handwritten style, with varying stroke thicknesses. The numbers are slightly tilted to the right.
DFC 17 1997
## APPENDIX E
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS FOR VASOCOR PVR
.
ﻟﻤﺴﺴﺴﺎ
{1}------------------------------------------------
K973659
Submitter Vasocor 4001 N.W. 97th Avenue, Suite 101 Miami. Florida 33178 Ph (305) 592-7260 Fax (305)-592-7937
## Date summary was prepared: September 20, 1995
Name(s) of the device Vasocor PVR100
Identification of predicate device(s) Life Sciences Pulse Volume Recorder (preamendment and K832679)
### Description of the device
The Vasocor PVR-100 provides Pulse Volume Recording (PVR) made possible through air plethysmography. The PVR-100 device, which is based on a PC platform, includes various standard blood pressure cuff sizes that allow use of the PVR measurement on various limb and digit extremities, a pneumatic circuit that allows for internal calibration as well as inflation cuff pressure, and a user interface that includes various menus and screens for inputting patient demographic data, performing tests, displaying and entering results, and printing reports. The PVR tracing is accomplished by attaching a standard blood pressure cuff on various areas of the extremities. Once atmospheric air is injected into the cuff to provide skin contact, a pressure transducer records the pressure change in the PVR cuff secondary to change in volume of the extremity segment under the cuff during repeated cardiac cycles. PVR measurement is accomplished by calculating the volume change in the extremity segment under evaluation, where the PVR tracing shows the change in pressure of the same segment during the cardiac cycle.
#### Intended Use
The Vasocor PVR-100 (Pulse Volume Recorder) is a non-invasive medical device used in conjunction with other devices such as; continuous-wave Doppler ultrasound, treadmill testing, and ultrasonic imaging techniques for vascular studies of limbs and digits. Use of the PVR in this manner allows the physician to non-invasively diagnosis extremity arterial and venous disorders.
{2}------------------------------------------------
| Table of Technological Characteristics | | | |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Vasocor PVR | Life Science PVR IV | Parks Medical<br>Mini Lab VI |
| 510(k) numbers | new submission | K832679 | unknown |
| Indications for use | venous and arterial<br>studies of limbs | same | same |
| Contraindications | none | same | same |
| Warnings / Precautions | prescription device | same | same |
| General Design | computer based,<br>provides PVR<br>tracings, includes<br>pnematic circuit,<br>standard blood<br>pressure cuffs and<br>tubing | mechanical design, manual<br>calibration | same as Life Sciences<br>PVR IV |
| Features | Internal calibration,<br>computer controlled ,<br>external printer, save<br>patient data, print<br>patient reports | built in strip chart printer,<br>cannot print reports or save<br>patient data | comes with either a<br>built in strip chart<br>recorder or a PC<br>computer for record<br>management and<br>printing, or both |
| Gauges & Sensors | pressure transducer | pressure transducer | pressure transducer |
| Software | software control of<br>patient records,<br>tracings, and<br>pneumatics | none | software control of<br>patient records and<br>tracings |
| Anatomical Site(s) | extremities and digits | same | same |
| Biocompatibility | blood pressure cuffs<br>are standard marketed<br>products are the same<br>ones used by the Life<br>Sciences PVR IV | same | similar |
)
)
E-2
{3}------------------------------------------------
### Non clinical testing:
ﻤﺴﺴﺴ
A hemodynamic model of the arterial circulation in the lower extremity was developed specifically for the Vasocor PVR. The major feature of this model is that it produces Pressure and Volume contours that mimic the normal human circulation in lower extremities. By changing various parameters the operator can control hemodynamic variables. Flowrate, system compliance, and resistance to control pule pressure and heart rate were varied in the in-vitro comparison testing. The major comparative parameter is the PVR amplitude and contour. Amplitudes were measured in millimeters to the nearest 0.5mm. Contour analysis were made by visual analysis (method used clinically).
#### Clinical testing:
Six subjects were evaluated using the new Vasocor PVR-100 subject to this 510(k) and the predicate device, the Life Sciences Pulse Volume Recorder. Three(3) normal subjects with no evidence of peripheral vascular disease and three(3) subjects with known peripheral vascular disease were selected for this comparison. Standard blood pressure cuffs were connected to the subjects. The cuffs were placed on upper arm, thumb, calf, and ankle for comparisons in both the normal subjects and subjects with known peripheral vascular disease.
#### Conclusion:
The Vasocor PVR is substatntially equivalent to the Life Sciences Pulse Volume Recorder
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
DEC 17 1997
Jeff Raines, M.D., Ph.D. Vasocor, Inc. 4001 NW 97th Avenue, Suite 101 Miami, FL 33178
Re : K973659 Vasocor PVR 100 Trade Name: Regulatory Class: II Product Code: 74 JOM Dated: September 5, 1997 Received: September 25, 1997
Dear Dr. Raines:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
Image /page/4/Picture/9 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle, with its wings spread.
{5}------------------------------------------------
#### Page 2 - Jeff Raines, M.D., Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callanan, Ph. D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
:
{6}------------------------------------------------
510(k) Number None assigned as of this time
Device Name Vasocor PVR-100
### Indications for Use
The Vasocor PVR-100 (Pulse Volume Recorder) is a non-invasive medical device used in conjunction with other devices such as; continuous-wave Doppler ultrasound, treadmill testing, and ultrasonic imaging techniques for vascular studies of limbs and digits. Use of the PVR in this manner allows the physician to non-invasively diagnosis extremity arterial and venous disorders.
=============================================================================================================================================================================
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109) ロ Over-the Counter Use
(Division Sign-09 12/6/97
Division of Cardiova
and Neurological Devices
510(k) Number K973659
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.