SEEMOR - DISPLAY & PROCESSING PROGRAM EXECUTING ON PERSONAL COMPUTER SYSTEM
Applicant
Areeda Assoc., Ltd.
Product Code
LLZ · Radiology
Decision Date
Dec 12, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The SeeMor™ software program should be used for the display and image manipulation of multimodality diagnostic medical images. The intended use of this program was to provide the physician with a display program which would allow for the remote viewing and interpretation of medical images. This program serves merely as a display program to aid in the diagnostic interpretation of a patients' study. It does not provide any diagnostic interpretive output other than the display of the images. Therefore, this program has no direct adverse effect on health since it is only providing a means of displaying the medical images for the physician. The final responsibility for interpretation of the study lies with the physician.
Device Story
SeeMor™ is a software program for personal computers (IBM PC or Macintosh) designed for viewing and manipulating diagnostic medical images. It accepts medical image data as input and provides tools for image processing, including clipping, window/level adjustment, magnification, pan, relate, add, delete, next, cine, lock, select, view, flip, color table adjustment, and orthogonal view reconstruction. The device is intended for use by physicians in clinical settings to facilitate remote viewing and interpretation of patient studies. It does not provide automated diagnostic interpretation; the physician retains final responsibility for clinical decision-making based on the displayed images. The software aims to improve diagnostic workflow by providing a non-proprietary platform for image review.
Clinical Evidence
In-house trial validation conducted on 25 patients to establish effectiveness. No clinical prospective or retrospective studies reported. Bench testing included design, coding, debugging, and validation.
Technological Characteristics
Software-based medical imaging display program. Executes on non-proprietary hardware (IBM PC or Macintosh). Provides image manipulation tools (clipping, window/level, magnification, pan, cine, orthogonal reconstruction).
Indications for Use
Indicated for the display and manipulation of multimodality diagnostic medical images to aid physicians in remote viewing and interpretation of patient studies.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# ITEM I
K973653 ×973653
# 510(k) SUMMARY
DEC 1 2 1997
## Safety and Effectiveness
#### 1. Medical Device Establishment
Areeda Associates Ltd 516 N. Curson Avenue Los Angeles, CA 90036 Voice: (213) 653-5515 (213) 653-5514 FAX: e-mail: joe@areeda.com Contact person: Joseph Areeda or Kenneth Van Train Date Summary Prepared: September 24, 1997
#### 2. Medical Device:
SeeMor™ - Medical Imaging Display & Processing program executing on personal computer systems.
### 3. Medical Device Equivalence:
MedVision™ Imaging Software developed by Evergreen Technologies, Inc., Ref. 510(k) #: K924178.
#### 4. Device Description:
This medical device (SeeMor™) is a display program for viewing diagnostic medical images. The program provides the capabilities of manipulating the images being displayed with the following command options: clipping, window/level adjustment, magnification, pan, relate, add, delete, next, cine, lock, select, view, flip vertical/horizontal, set color table, orthogonal view reconstruction, cascade, tile, and reset.
### 5. Intended Use and Potential Adverse Effect on Health:
The intended use of this program was to provide the physician with a display program which would allow for the remote viewing and interpretation of
{1}------------------------------------------------
medical images. This program serves merely as a display program to aid in the diagnostic interpretation of a patients' study. It does not provide any diagnostic interpretive output other than the display of the images. Therefore, this program has no direct adverse effect on health since it is only providing a means of displaying the medical images for the physician. The final responsibility for interpretation of the study lies with the physician.
### 6. Marketing History:
There have been multiple medical device display programs marketed in the past which perform similar functions to those performed by SeeMor™. These programs are all used for the sole purpose of displaying medical images for the diagnostic interpretation by a physician. The SeeMor™ program provides a program which executes on non-propriety hardware (IBM PC or Macintosh) and we believe is substantially equivalent to the MedVision Viewer developed by Evergreen Medical Technologies K924178. To our knowledge there have been no safety problems with the MedVision Viewer medical display program which has been in the marketplace for over four years. In addition, we are not aware of any problems on any of the proprietary medical display programs which have been in the marketplace for over 20 years.
### 7. Conclusions:
The safety of this program has been determined through the various stages of software development which included the initial design, coding, debugging, testing, and in-house validation. The effectiveness of the program has been established in an in-house trial validation which included an evaluation of 25 patients. We contend that the method employed for the development and the final in-house trial validation results of this medical display software program (SeeMor™) have proven its safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three human figures in profile, with flowing lines suggesting movement or connection.
Food and Druq Administration 9200 Corporate Boulevard Rockville MD 20850
DEC | 2 |997
Joseph Areeda President Areeda Associates, Ltd. 516 N. Curson Ave. Los Angeles, CA 90036-1814 Re:
K973653 SeeMor™ (Image Processing System) Dated: September 24, 1997 Received: September 25, 1997 Regulatory class: Unclassified Procode: 90 LLZ
Dear Mr. Areeda:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liau Yu
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal. Ear. Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
510(k) Number (if known): ___K973653
Device Name: _________________________________________________________________________________________________________________________________________________________________ Personal Computers
Indications For Use:
The SeeMor™ software program should be used for the display and image manipulation of multimodality diagnostic medical images.
(PLEASE DO NOT WRITE BELOW THIS UNE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ 1 (Per 21 CER 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Willian Y
(Division Sign-Off) (Division Sign-Off)
Division of Denductive, Abdominal, ENT, and Rachelogical Devices
510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.