This notification is for an in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.
Device Story
The Vitalab Flexor (also known as Selectra 2 or Viva) is a random-access, automated clinical micro-chemistry analyzer. It processes patient samples to measure analytes for clinical chemistry, drugs of abuse, and therapeutic drug monitoring. The system operates in a clinical laboratory setting, managed by laboratory technicians or clinicians. It supports multiple analytical modes: endpoint, kinetic with linearity check, bichromatic endpoint, two-point kinetic, and non-linear calibration. The device utilizes an optional ISE accessory for sodium, potassium, and chloride determination. Output consists of quantitative analyte measurements, which clinicians use to assess patient health, monitor drug levels, or screen for substance use. The system provides automated, high-throughput testing (up to 180 tests/hour) to assist in diagnostic decision-making.
Clinical Evidence
Bench testing only. Evaluation performed on 38 representative methods including clinical chemistry analytes (e.g., AST, ALT, Glucose, Creatinine), drugs of abuse (e.g., Benzodiazepine, Cocaine, Opiates), and therapeutic drugs (e.g., Cyclosporin, Carbamazepine). Data assessed precision, accuracy, linearity, and drift. Results demonstrated positive correlation and agreement with commercially marketed systems.
Technological Characteristics
Automated random-access micro-chemistry analyzer. Analytical modes: endpoint, kinetic, bichromatic endpoint, two-point kinetic, non-linear calibration. Throughput: up to 180 tests/hour. Optional ISE accessory for Na, K, Cl. No specific materials or software architecture standards provided.
Indications for Use
Indicated for clinical laboratory use to measure various analytes in clinical chemistry, drugs of abuse monitoring, and therapeutic drug monitoring. No specific patient population age or gender restrictions are stated.
Regulatory Classification
Identification
A micro chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. The distinguishing characteristic of the device is that it requires only micro volume samples obtainable from pediatric patients. This device is intended for use in conjunction with certain materials to measure a variety of analytes.
Predicate Devices
Roche Cobas Mira
Reference Devices
AVL Scientific Corp. ISE unit
Submission Summary (Full Text)
{0}------------------------------------------------
K9736028
# VITAL MAR 1 3 1998
USA: One Gateway Center, Suite 415 New
nsselaenvers 4 Spankers
September 19, 1997
## Premarket Notification [510(k)] Summary
## for the VITALAB FLEXOR Clinical Laboratory Analysis and Reagent System
#### Device Name:
Vitalab Flexor (In-vitro diagnostic system). The Vitalab Flexor is also tradenamed the Vitalab Selectra 2 and the Vitalab Viva. References in this document submission may use either of the two names interchangeably.
### Common Name:
Automated chemistry analyzer, micro, for clinical use.
#### Classification Name:
Analyzer, chemistry, micro, for clinical use - has been classified as Class I devices, as per 21 CFR 862.2170. This device is intended for clinical use in conjunction with certain materials to measure a variety of analytes.
#### Establishment Registration:
Vital Scientific NV, a company established under the laws of the Netherlands, with headquarters at van Rensselaerweg 4, Spankeren, 6956 AV, The Netherlands and with a USA branch office located at One Gateway Center, Suite 415, Newton, MA 02158. Vital Scientific is a registered as a medical device manufacturer under files number 8030478. All instruments are manufactured in the Netherlands. Contact person: Israel M. Stein MD.
#### Performance Standards:
There are no performance standards for these devices.
{1}------------------------------------------------
# Labeling:
This device is labeled in accordance with 21 CFR 809.10.
# Summary description of device:
A random access, automated Clinical Micro-Chemistry Analyzer with a throughput of up to 180 test per hour which is intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry. monitoring drugs of abuse and in therapeutic drug monitoring.
# Substantial Equivalence:
The Vitalab Flexor System is substantially equivalent to the Roche Cobas Mira presently in distribution in the United States. The Vitalab Flexor System has the same technological characteristics and intended use as these systems. These systems are automated micro-chemistry analyzers intended for clinical use in conjunction with certain materials to measure a variety of analytes (21 CFR 862.2170).
We are including data on twenty-nine (29) representative Clinical Chemistry analytes manufactured by Trace Scientific Pty. Ltd., seven (7) representative analytes for Drugs of Abuse manufactured by Behring Diagnostics and two (2) representative analytes for therapeutic drug monitoring, manufactured by Behring Diagnostics. All of these analytes are presently 510(k) approved under separate submissions by their respective companies. These analytes were chosen because thev are representative of the reaction types to be processed on the analyzer.
Thirty-eight (38) representative methods were selected for evaluation including Acid Phosphatase, Alkaline Phosphatase, Albumin, ALT, Amylase, AST, Bicarbonate, Direct Bilirubin, Total Bilirubin, Calcium, Chloride, Creatinine Kinase, Creatinine, Gamma GT, Glucose, Inorganic Phosphate, Iron, LDH, Magnesium, Total Protein, Triglycerides, Uric Acid, Benzodiazepine, Cocaine, Opiate, Amphetamine, Methadone, THC, Cyclosporin, and Carbamazepine. These include the following analytical modes: endpoint, kinetic with linearity check, bichromatic endpoint, two-point kinetic, and non-linear calibration curve.
The data for precision, accuracy, linearity and drift demonstrate general agreement with levels that are comparable to other systems presently being
{2}------------------------------------------------
marketed commercially in the United States. The data demonstrate positive correlation and substantial equivalence.
The optional Vitalab ISE accessory for the determination of Sodium, Potassium, and Chloride is manufactured by AVL Scientific Corp. This unit is essentially equivalent to other ISE units marketed by AVL in the United States.
Vital Scientific NV, submits that the Vitalab Flexor System has clinically acceptable performance comparable to similar devices currently in commercial distribution. These data provide scientific evidence of equivalency for Vitalab Flexor System. The data indicates a positive correlation, and therefore, is substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three intertwined strands and a single wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 1 3 1998
Israel M. Stein, M.D. Managing Director Vital Scientific One Gateway Center, Suite 415 Newton, Massachusetts 02158
K973628 Re : VITALAB FLEXOR Analyzer Regulatory Class: II Product Code: JJF, CDN, CEM, CGA, DJC, DKZ, JGS, JXO, KLT, LAR, JRE Dated: January 30, 1998 February 2, 1998 Received:
Dear Dr. Stein:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K973628 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
VITALAB FLEXOR Device Name:
Indications For Use:
- 17
This notification is for an in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 1973628
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.